Cargando…
Estimation of the respiratory rate from ballistocardiograms using the Hilbert transform
BACKGROUND: Measuring the respiratory rate is usually associated with discomfort for the patient due to contact sensors or a high time demand for healthcare personnel manually counting it. METHODS: In this paper, two methods for the continuous extraction of the respiratory rate from unobtrusive ball...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9354426/ https://www.ncbi.nlm.nih.gov/pubmed/35927665 http://dx.doi.org/10.1186/s12938-022-01024-4 |
Sumario: | BACKGROUND: Measuring the respiratory rate is usually associated with discomfort for the patient due to contact sensors or a high time demand for healthcare personnel manually counting it. METHODS: In this paper, two methods for the continuous extraction of the respiratory rate from unobtrusive ballistocardiography signals are introduced. The Hilbert transform is used to generate an amplitude-invariant phase signal in-line with the respiratory rate. The respiratory rate can then be estimated, first, by using a simple peak detection, and second, by differentiation. RESULTS: By analysis of a sleep laboratory data set consisting of nine records of healthy individuals lasting more than 63 h and including more than 59,000 breaths, a mean absolute error of as low as 0.7 BPM for both methods was achieved. CONCLUSION: The results encourage further assessment for hospitalised patients and for home-care applications especially with patients suffering from diseases of the respiratory system like COPD or sleep apnoea. |
---|