Is a weekly qualitative picket fence test sufficient? A proposed alternate EPID‐based weekly MLC QA program
PURPOSE: Well‐designed routine multileaf collimator (MLC) quality assurance (QA) is important to assure external‐beam radiation treatment delivery accuracy. This study evaluates the clinical necessity of a comprehensive weekly (C‐Weekly) MLC QA program compared to the American Association of Physics...
Autores principales: | , , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9359020/ https://www.ncbi.nlm.nih.gov/pubmed/35856943 http://dx.doi.org/10.1002/acm2.13699 |
Sumario: | PURPOSE: Well‐designed routine multileaf collimator (MLC) quality assurance (QA) is important to assure external‐beam radiation treatment delivery accuracy. This study evaluates the clinical necessity of a comprehensive weekly (C‐Weekly) MLC QA program compared to the American Association of Physics in Medicinerecommended weekly picket fence test (PF‐Weekly), based on our seven‐year experience with weekly MLC QA. METHODS: The C‐Weekly MLC QA program used in this study includes 5 tests to analyze: (1) absolute MLC leaf position; (2) interdigitation MLC leaf position; (3) picket fence MLC leaf positions at static gantry angle; (4) minimum leaf‐gap setting; and (5) volumetric‐modulated arc therapy delivery. A total of 20,226 QA images from 16,855 tests (3,371 tests × 5) for 11 linacs at 5 photon clinical sites from May 2014 to June 2021 were analyzed. Failure mode and effects analysis was performed with 5 failure modes related to the 5 tests. For each failure mode, a risk probability number (RPN) was calculated for a C‐Weekly and a PF‐Weekly MLC QA program. The probability of occurrence was evaluated from statistical analyses of the C‐Weekly MLC QA. RESULTS: The total number of failures for these 16,855 tests was 143 (0.9%): 39 (27.3%) for absolute MLC leaf position, 13 (9.1%) for interdigitation position, 9 (6.3%) for static gantry picket fence, 2 (1.4%) for minimum leaf‐gap setting, and 80 (55.9%) for VMAT delivery. RPN scores for PF‐Weekly MLC QA ranged from 60 to 192 and from 48 to 96 for C‐Weekly MLC QA. CONCLUSION: RPNs for the 5 failure modes of MLC QA tests were quantitatively determined and analyzed. A comprehensive weekly MLC QA is imperative to lower the RPNs of the 5 failure modes to the desired level (<125); those from the PF‐Weekly MLC QA program were found to be higher (>125). This supports the clinical necessity for comprehensive weekly MLC QA. |
---|