Impact of systematic MLC positional uncertainties on the quality of single‐isocenter multi‐target VMAT‐SRS treatment plans

PURPOSE: To study the impact of systematic MLC leaf positional uncertainties (stemming from mechanical inaccuracies or sub‐optimal MLC modeling) on the quality of intracranial single‐isocenter multi‐target VMAT‐SRS treatment plans. An estimation of appropriate tolerance levels is attempted. METHODS:...

Descripción completa

Detalles Bibliográficos
Autores principales: Prentou, Georgia, Pappas, Eleftherios P, Prentou, Eleni, Yakoumakis, Nikolaos, Paraskevopoulou, Chryssa, Koutsouveli, Efi, Pantelis, Evaggelos, Papagiannis, Panagiotis, Karaiskos, Pantelis
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9359048/
https://www.ncbi.nlm.nih.gov/pubmed/35733367
http://dx.doi.org/10.1002/acm2.13708
_version_ 1784764057770786816
author Prentou, Georgia
Pappas, Eleftherios P
Prentou, Eleni
Yakoumakis, Nikolaos
Paraskevopoulou, Chryssa
Koutsouveli, Efi
Pantelis, Evaggelos
Papagiannis, Panagiotis
Karaiskos, Pantelis
author_facet Prentou, Georgia
Pappas, Eleftherios P
Prentou, Eleni
Yakoumakis, Nikolaos
Paraskevopoulou, Chryssa
Koutsouveli, Efi
Pantelis, Evaggelos
Papagiannis, Panagiotis
Karaiskos, Pantelis
author_sort Prentou, Georgia
collection PubMed
description PURPOSE: To study the impact of systematic MLC leaf positional uncertainties (stemming from mechanical inaccuracies or sub‐optimal MLC modeling) on the quality of intracranial single‐isocenter multi‐target VMAT‐SRS treatment plans. An estimation of appropriate tolerance levels is attempted. METHODS: Five patients, with three to four metastases and at least one target lying in close proximity to organs‐at‐risk (OARs) were included in this study. A single‐isocenter multi‐arc VMAT plan per patient was prepared, which served as the reference for dosimetric impact evaluation. A range of leaf offsets was introduced (±0.03 mm up to ±0.30 mm defined at the MLC plane) to both leaf banks, by varying the leaf offset MLC modeling parameter in Monaco for all the prepared plans, in order to simulate projected leaf offsets of ±0.09 mm up to ±0.94 mm at the isocenter plane, respectively. For all offsets simulated and cases studied, dose distributions were re‐calculated and compared with the corresponding reference ones. An experimental dosimetric procedure using the SRS mapCHECK diode array was also performed to support the simulation study results and investigate its suitability to detect small systematic leaf positional errors. RESULTS: Projected leaf offsets of ±0.09 mm were well‐tolerated with respect to both target dosimetry and OAR‐sparing. A linear relationship was found between D (95%) percentage change and projected leaf offset (slope: 12%/mm). Impact of projected offset on target dosimetry was strongly associated with target volume. In two cases, plans that could be considered potentially clinically unacceptable (i.e., clinical dose constraint violation) were obtained even for projected offsets as small as 0.19 mm. The performed experimental dosimetry check can detect potential small systematic leaf errors. CONCLUSIONS: Plan quality indices and dose–volume metrics are very sensitive to systematic sub‐millimeter leaf positional inaccuracies, projected at the isocenter plane. Acceptable and tolerance levels in systematic MLC uncertainties need to be tailored to VMAT‐SRS spatial and dosimetric accuracy requirements.
format Online
Article
Text
id pubmed-9359048
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-93590482022-08-10 Impact of systematic MLC positional uncertainties on the quality of single‐isocenter multi‐target VMAT‐SRS treatment plans Prentou, Georgia Pappas, Eleftherios P Prentou, Eleni Yakoumakis, Nikolaos Paraskevopoulou, Chryssa Koutsouveli, Efi Pantelis, Evaggelos Papagiannis, Panagiotis Karaiskos, Pantelis J Appl Clin Med Phys Radiation Oncology Physics PURPOSE: To study the impact of systematic MLC leaf positional uncertainties (stemming from mechanical inaccuracies or sub‐optimal MLC modeling) on the quality of intracranial single‐isocenter multi‐target VMAT‐SRS treatment plans. An estimation of appropriate tolerance levels is attempted. METHODS: Five patients, with three to four metastases and at least one target lying in close proximity to organs‐at‐risk (OARs) were included in this study. A single‐isocenter multi‐arc VMAT plan per patient was prepared, which served as the reference for dosimetric impact evaluation. A range of leaf offsets was introduced (±0.03 mm up to ±0.30 mm defined at the MLC plane) to both leaf banks, by varying the leaf offset MLC modeling parameter in Monaco for all the prepared plans, in order to simulate projected leaf offsets of ±0.09 mm up to ±0.94 mm at the isocenter plane, respectively. For all offsets simulated and cases studied, dose distributions were re‐calculated and compared with the corresponding reference ones. An experimental dosimetric procedure using the SRS mapCHECK diode array was also performed to support the simulation study results and investigate its suitability to detect small systematic leaf positional errors. RESULTS: Projected leaf offsets of ±0.09 mm were well‐tolerated with respect to both target dosimetry and OAR‐sparing. A linear relationship was found between D (95%) percentage change and projected leaf offset (slope: 12%/mm). Impact of projected offset on target dosimetry was strongly associated with target volume. In two cases, plans that could be considered potentially clinically unacceptable (i.e., clinical dose constraint violation) were obtained even for projected offsets as small as 0.19 mm. The performed experimental dosimetry check can detect potential small systematic leaf errors. CONCLUSIONS: Plan quality indices and dose–volume metrics are very sensitive to systematic sub‐millimeter leaf positional inaccuracies, projected at the isocenter plane. Acceptable and tolerance levels in systematic MLC uncertainties need to be tailored to VMAT‐SRS spatial and dosimetric accuracy requirements. John Wiley and Sons Inc. 2022-06-22 /pmc/articles/PMC9359048/ /pubmed/35733367 http://dx.doi.org/10.1002/acm2.13708 Text en © 2022 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, LLC on behalf of The American Association of Physicists in Medicine. https://creativecommons.org/licenses/by/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
Prentou, Georgia
Pappas, Eleftherios P
Prentou, Eleni
Yakoumakis, Nikolaos
Paraskevopoulou, Chryssa
Koutsouveli, Efi
Pantelis, Evaggelos
Papagiannis, Panagiotis
Karaiskos, Pantelis
Impact of systematic MLC positional uncertainties on the quality of single‐isocenter multi‐target VMAT‐SRS treatment plans
title Impact of systematic MLC positional uncertainties on the quality of single‐isocenter multi‐target VMAT‐SRS treatment plans
title_full Impact of systematic MLC positional uncertainties on the quality of single‐isocenter multi‐target VMAT‐SRS treatment plans
title_fullStr Impact of systematic MLC positional uncertainties on the quality of single‐isocenter multi‐target VMAT‐SRS treatment plans
title_full_unstemmed Impact of systematic MLC positional uncertainties on the quality of single‐isocenter multi‐target VMAT‐SRS treatment plans
title_short Impact of systematic MLC positional uncertainties on the quality of single‐isocenter multi‐target VMAT‐SRS treatment plans
title_sort impact of systematic mlc positional uncertainties on the quality of single‐isocenter multi‐target vmat‐srs treatment plans
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9359048/
https://www.ncbi.nlm.nih.gov/pubmed/35733367
http://dx.doi.org/10.1002/acm2.13708
work_keys_str_mv AT prentougeorgia impactofsystematicmlcpositionaluncertaintiesonthequalityofsingleisocentermultitargetvmatsrstreatmentplans
AT pappaseleftheriosp impactofsystematicmlcpositionaluncertaintiesonthequalityofsingleisocentermultitargetvmatsrstreatmentplans
AT prentoueleni impactofsystematicmlcpositionaluncertaintiesonthequalityofsingleisocentermultitargetvmatsrstreatmentplans
AT yakoumakisnikolaos impactofsystematicmlcpositionaluncertaintiesonthequalityofsingleisocentermultitargetvmatsrstreatmentplans
AT paraskevopoulouchryssa impactofsystematicmlcpositionaluncertaintiesonthequalityofsingleisocentermultitargetvmatsrstreatmentplans
AT koutsouveliefi impactofsystematicmlcpositionaluncertaintiesonthequalityofsingleisocentermultitargetvmatsrstreatmentplans
AT pantelisevaggelos impactofsystematicmlcpositionaluncertaintiesonthequalityofsingleisocentermultitargetvmatsrstreatmentplans
AT papagiannispanagiotis impactofsystematicmlcpositionaluncertaintiesonthequalityofsingleisocentermultitargetvmatsrstreatmentplans
AT karaiskospantelis impactofsystematicmlcpositionaluncertaintiesonthequalityofsingleisocentermultitargetvmatsrstreatmentplans