Assessment of the New Acceleromyograph TOF 3D Compared with the Established TOF Watch SX: Bland-Altman Analysis of the Precision and Limits of Agreement between Both Devices—A Randomized Clinical Comparison

The new acceleromyograph TOF 3D was compared with the established TOF Watch SX in patients undergoing elective laparoscopic gynecological surgery. Neuromuscular transmission was assessed by simultaneous recording with both devices. Measurements were performed simultaneously at the left and the right...

Descripción completa

Detalles Bibliográficos
Autores principales: Soltesz, Stefan, Thomas, Jan, Anapolski, Michael, Noé, Guenter Karl
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9369732/
https://www.ncbi.nlm.nih.gov/pubmed/35956198
http://dx.doi.org/10.3390/jcm11154583
_version_ 1784766560960774144
author Soltesz, Stefan
Thomas, Jan
Anapolski, Michael
Noé, Guenter Karl
author_facet Soltesz, Stefan
Thomas, Jan
Anapolski, Michael
Noé, Guenter Karl
author_sort Soltesz, Stefan
collection PubMed
description The new acceleromyograph TOF 3D was compared with the established TOF Watch SX in patients undergoing elective laparoscopic gynecological surgery. Neuromuscular transmission was assessed by simultaneous recording with both devices. Measurements were performed simultaneously at the left and the right M. adductor pollicis (Group A, 25 patients), or the M. corrugator supercilii (Group CS, 25 patients). The repeatability, time course, and limits of agreement (Bland-Altman) were compared. The primary endpoint was the 90% train-of-four recovery time (TOFR 0.9). Other endpoints included onset time of block, maximum T1 depression, time to 25% T1 recovery, the recovery time course of T1 response, and TOF ratio, respectively. In group CS, the repeatability coefficient of the TOF 3D was lower (4.66 (1.6)) than of the TOF Watch SX (6.02 (1.9); p = 0.026). In group A, the onset of the block was faster when measured by the TOF 3D (98.7 (30) s vs. 112.2 (36) s (mean (SD)); p = 0.032). In group A, time to recovery to a TOFR of 90% was measured earlier by the TOF 3D (bias −0.71 min, limits of agreement from −8.94 to +7.51 min). The TOF 3D provides adequate information with high precision and sensitivity. It is suitable even for measurement sites with small muscle contractions such as the M. corrugator supercilii.
format Online
Article
Text
id pubmed-9369732
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-93697322022-08-12 Assessment of the New Acceleromyograph TOF 3D Compared with the Established TOF Watch SX: Bland-Altman Analysis of the Precision and Limits of Agreement between Both Devices—A Randomized Clinical Comparison Soltesz, Stefan Thomas, Jan Anapolski, Michael Noé, Guenter Karl J Clin Med Article The new acceleromyograph TOF 3D was compared with the established TOF Watch SX in patients undergoing elective laparoscopic gynecological surgery. Neuromuscular transmission was assessed by simultaneous recording with both devices. Measurements were performed simultaneously at the left and the right M. adductor pollicis (Group A, 25 patients), or the M. corrugator supercilii (Group CS, 25 patients). The repeatability, time course, and limits of agreement (Bland-Altman) were compared. The primary endpoint was the 90% train-of-four recovery time (TOFR 0.9). Other endpoints included onset time of block, maximum T1 depression, time to 25% T1 recovery, the recovery time course of T1 response, and TOF ratio, respectively. In group CS, the repeatability coefficient of the TOF 3D was lower (4.66 (1.6)) than of the TOF Watch SX (6.02 (1.9); p = 0.026). In group A, the onset of the block was faster when measured by the TOF 3D (98.7 (30) s vs. 112.2 (36) s (mean (SD)); p = 0.032). In group A, time to recovery to a TOFR of 90% was measured earlier by the TOF 3D (bias −0.71 min, limits of agreement from −8.94 to +7.51 min). The TOF 3D provides adequate information with high precision and sensitivity. It is suitable even for measurement sites with small muscle contractions such as the M. corrugator supercilii. MDPI 2022-08-05 /pmc/articles/PMC9369732/ /pubmed/35956198 http://dx.doi.org/10.3390/jcm11154583 Text en © 2022 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Soltesz, Stefan
Thomas, Jan
Anapolski, Michael
Noé, Guenter Karl
Assessment of the New Acceleromyograph TOF 3D Compared with the Established TOF Watch SX: Bland-Altman Analysis of the Precision and Limits of Agreement between Both Devices—A Randomized Clinical Comparison
title Assessment of the New Acceleromyograph TOF 3D Compared with the Established TOF Watch SX: Bland-Altman Analysis of the Precision and Limits of Agreement between Both Devices—A Randomized Clinical Comparison
title_full Assessment of the New Acceleromyograph TOF 3D Compared with the Established TOF Watch SX: Bland-Altman Analysis of the Precision and Limits of Agreement between Both Devices—A Randomized Clinical Comparison
title_fullStr Assessment of the New Acceleromyograph TOF 3D Compared with the Established TOF Watch SX: Bland-Altman Analysis of the Precision and Limits of Agreement between Both Devices—A Randomized Clinical Comparison
title_full_unstemmed Assessment of the New Acceleromyograph TOF 3D Compared with the Established TOF Watch SX: Bland-Altman Analysis of the Precision and Limits of Agreement between Both Devices—A Randomized Clinical Comparison
title_short Assessment of the New Acceleromyograph TOF 3D Compared with the Established TOF Watch SX: Bland-Altman Analysis of the Precision and Limits of Agreement between Both Devices—A Randomized Clinical Comparison
title_sort assessment of the new acceleromyograph tof 3d compared with the established tof watch sx: bland-altman analysis of the precision and limits of agreement between both devices—a randomized clinical comparison
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9369732/
https://www.ncbi.nlm.nih.gov/pubmed/35956198
http://dx.doi.org/10.3390/jcm11154583
work_keys_str_mv AT solteszstefan assessmentofthenewacceleromyographtof3dcomparedwiththeestablishedtofwatchsxblandaltmananalysisoftheprecisionandlimitsofagreementbetweenbothdevicesarandomizedclinicalcomparison
AT thomasjan assessmentofthenewacceleromyographtof3dcomparedwiththeestablishedtofwatchsxblandaltmananalysisoftheprecisionandlimitsofagreementbetweenbothdevicesarandomizedclinicalcomparison
AT anapolskimichael assessmentofthenewacceleromyographtof3dcomparedwiththeestablishedtofwatchsxblandaltmananalysisoftheprecisionandlimitsofagreementbetweenbothdevicesarandomizedclinicalcomparison
AT noeguenterkarl assessmentofthenewacceleromyographtof3dcomparedwiththeestablishedtofwatchsxblandaltmananalysisoftheprecisionandlimitsofagreementbetweenbothdevicesarandomizedclinicalcomparison