Cargando…

Strategies for the Accurate Measurement of the Resonance Frequency in QCM-D Systems via Low-Cost Digital Techniques

In this paper, an FPGA (Field Programmable Gate Array)-based digital architecture for the measurement of quartz crystal microbalance (QCM) oscillating frequency of transient responses, i.e., in QCM-D (QCM and Dissipation) applications, is presented. The measurement system is conceived for operations...

Descripción completa

Detalles Bibliográficos
Autores principales: Addabbo, Tommaso, Fort, Ada, Landi, Elia, Moretti, Riccardo, Mugnaini, Marco, Vignoli, Valerio
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9371052/
https://www.ncbi.nlm.nih.gov/pubmed/35957285
http://dx.doi.org/10.3390/s22155728
_version_ 1784767014529662976
author Addabbo, Tommaso
Fort, Ada
Landi, Elia
Moretti, Riccardo
Mugnaini, Marco
Vignoli, Valerio
author_facet Addabbo, Tommaso
Fort, Ada
Landi, Elia
Moretti, Riccardo
Mugnaini, Marco
Vignoli, Valerio
author_sort Addabbo, Tommaso
collection PubMed
description In this paper, an FPGA (Field Programmable Gate Array)-based digital architecture for the measurement of quartz crystal microbalance (QCM) oscillating frequency of transient responses, i.e., in QCM-D (QCM and Dissipation) applications, is presented. The measurement system is conceived for operations in liquid, with short QCM transient responses due to the large mechanical load. The proposed solution allows for avoiding the complex processing systems typically required by the QCM-D techniques and grants frequency resolutions better than 1 ppm. The core of the architecture is a reciprocal digital frequency meter, combined with the preprocessing of the QCM signal through mixing operations, such as a step-down of the input frequency and reducing the measurement error. The measurement error is further reduced through averaging. Different strategies are proposed to implement the proposed measurement solution, comprising an all-digital circuit and mixed analog/digital ones. The performance of the proposed architectures is theoretically derived, compared, and analyzed by means of experimental data obtained considering 10 MHz QCMs and 200 μs long transient responses. A frequency resolution of about 240 ppb, which corresponds to a Sauerbrey mass resolution of 8 ng/cm(2), is obtained for the all-digital solution, whereas for the mixed solution the resolution halves to 120 ppb, with a measurement time of about one second over 100 repetitions.
format Online
Article
Text
id pubmed-9371052
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-93710522022-08-12 Strategies for the Accurate Measurement of the Resonance Frequency in QCM-D Systems via Low-Cost Digital Techniques Addabbo, Tommaso Fort, Ada Landi, Elia Moretti, Riccardo Mugnaini, Marco Vignoli, Valerio Sensors (Basel) Article In this paper, an FPGA (Field Programmable Gate Array)-based digital architecture for the measurement of quartz crystal microbalance (QCM) oscillating frequency of transient responses, i.e., in QCM-D (QCM and Dissipation) applications, is presented. The measurement system is conceived for operations in liquid, with short QCM transient responses due to the large mechanical load. The proposed solution allows for avoiding the complex processing systems typically required by the QCM-D techniques and grants frequency resolutions better than 1 ppm. The core of the architecture is a reciprocal digital frequency meter, combined with the preprocessing of the QCM signal through mixing operations, such as a step-down of the input frequency and reducing the measurement error. The measurement error is further reduced through averaging. Different strategies are proposed to implement the proposed measurement solution, comprising an all-digital circuit and mixed analog/digital ones. The performance of the proposed architectures is theoretically derived, compared, and analyzed by means of experimental data obtained considering 10 MHz QCMs and 200 μs long transient responses. A frequency resolution of about 240 ppb, which corresponds to a Sauerbrey mass resolution of 8 ng/cm(2), is obtained for the all-digital solution, whereas for the mixed solution the resolution halves to 120 ppb, with a measurement time of about one second over 100 repetitions. MDPI 2022-07-31 /pmc/articles/PMC9371052/ /pubmed/35957285 http://dx.doi.org/10.3390/s22155728 Text en © 2022 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Addabbo, Tommaso
Fort, Ada
Landi, Elia
Moretti, Riccardo
Mugnaini, Marco
Vignoli, Valerio
Strategies for the Accurate Measurement of the Resonance Frequency in QCM-D Systems via Low-Cost Digital Techniques
title Strategies for the Accurate Measurement of the Resonance Frequency in QCM-D Systems via Low-Cost Digital Techniques
title_full Strategies for the Accurate Measurement of the Resonance Frequency in QCM-D Systems via Low-Cost Digital Techniques
title_fullStr Strategies for the Accurate Measurement of the Resonance Frequency in QCM-D Systems via Low-Cost Digital Techniques
title_full_unstemmed Strategies for the Accurate Measurement of the Resonance Frequency in QCM-D Systems via Low-Cost Digital Techniques
title_short Strategies for the Accurate Measurement of the Resonance Frequency in QCM-D Systems via Low-Cost Digital Techniques
title_sort strategies for the accurate measurement of the resonance frequency in qcm-d systems via low-cost digital techniques
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9371052/
https://www.ncbi.nlm.nih.gov/pubmed/35957285
http://dx.doi.org/10.3390/s22155728
work_keys_str_mv AT addabbotommaso strategiesfortheaccuratemeasurementoftheresonancefrequencyinqcmdsystemsvialowcostdigitaltechniques
AT fortada strategiesfortheaccuratemeasurementoftheresonancefrequencyinqcmdsystemsvialowcostdigitaltechniques
AT landielia strategiesfortheaccuratemeasurementoftheresonancefrequencyinqcmdsystemsvialowcostdigitaltechniques
AT morettiriccardo strategiesfortheaccuratemeasurementoftheresonancefrequencyinqcmdsystemsvialowcostdigitaltechniques
AT mugnainimarco strategiesfortheaccuratemeasurementoftheresonancefrequencyinqcmdsystemsvialowcostdigitaltechniques
AT vignolivalerio strategiesfortheaccuratemeasurementoftheresonancefrequencyinqcmdsystemsvialowcostdigitaltechniques