Twiddler's syndrome: Between mechanical recoil and behavioral troubles
A 72-year-old male with severe dilated cardiomyopathy, benefited of implantable cardiac defibrillator implementation. Device control shows high impedance. On X-ray, electrodes were completely twisted in the generator pocket, they were replaced and the generator was fixed to pectoralis-major fascia....
Autores principales: | , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Elsevier
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9372525/ https://www.ncbi.nlm.nih.gov/pubmed/35965939 http://dx.doi.org/10.1016/j.radcr.2022.07.035 |
Sumario: | A 72-year-old male with severe dilated cardiomyopathy, benefited of implantable cardiac defibrillator implementation. Device control shows high impedance. On X-ray, electrodes were completely twisted in the generator pocket, they were replaced and the generator was fixed to pectoralis-major fascia. Nurses report patient abnormal movements, scratching implantation area. This was identified as probably the trigger of the complication. |
---|