Cargando…

Mirau-based line-field confocal optical coherence tomography for three-dimensional high-resolution skin imaging

SIGNIFICANCE: Line-field confocal optical coherence tomography (LC-OCT) is a recently introduced high-resolution imaging modality based on a combination of low-coherence optical interferometry and reflectance confocal optical microscopy with line illumination and line detection. Capable of producing...

Descripción completa

Detalles Bibliográficos
Autores principales: Xue, Weikai, Ogien, Jonas, Bulkin, Pavel, Coutrot, Anne-Lise, Dubois, Arnaud
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Society of Photo-Optical Instrumentation Engineers 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9374567/
https://www.ncbi.nlm.nih.gov/pubmed/35962466
http://dx.doi.org/10.1117/1.JBO.27.8.086002
Descripción
Sumario:SIGNIFICANCE: Line-field confocal optical coherence tomography (LC-OCT) is a recently introduced high-resolution imaging modality based on a combination of low-coherence optical interferometry and reflectance confocal optical microscopy with line illumination and line detection. Capable of producing three-dimensional (3D) images of the skin with cellular resolution, in vivo, LC-OCT has been mainly applied in dermatology and dermo-cosmetology. The LC-OCT devices capable of acquiring 3D images reported so far are based on a Linnik interferometer using two identical microscope objectives. In this configuration, LC-OCT cannot be designed to be a very compact and light device, and the image acquisition speed is limited. AIM: The objective of this work was to develop a more compact and lighter LC-OCT device that is capable of acquiring images faster without significant degradation of the resolution and with optimized detection sensitivity. APPROACH: We developed an LC-OCT device based on a Mirau interferometer using a single objective. Dynamic adjustment of the camera frequency during the depth scan is implemented, using a faster camera and a more powerful light source. The reflectivity of the beam-splitter in the Mirau interferometer was optimized to maximize the detection sensitivity. A galvanometer scanner was incorporated into the device for scanning the illumination line laterally. A stack of adjacent B-scans, constituting a 3D image, can thus be acquired. RESULTS: The device is able to acquire and display B-scans at 17 fps. 3D images with a quasi-isotropic resolution of [Formula: see text] (1.3, 1.9, and [Formula: see text] in the [Formula: see text] , and [Formula: see text] directions, respectively) over a field of [Formula: see text] ([Formula: see text]) can be obtained. 3D imaging of human skin at cellular resolution, in vivo, is reported. CONCLUSIONS: The acquisition rate of the B-scans, at 17 fps, is unprecedented in LC-OCT. Compared with the conventional LC-OCT devices based on a Linnik interferometer, the reported Mirau-based LC-OCT device can acquire B-scans [Formula: see text] times faster. With potential advantages in terms of compactness and weight, a Mirau-based device could easily be integrated into a smaller and lighter handheld probe for use by dermatologists in their daily medical practice.