Performance evaluation between two automated biochemical analyzer systems: Roche Cobas 8000 and Mindray BS2000M

BACKGROUND: The values of biomarkers play a central role in routine clinical decision-making. Whereas the performance of different automated chemical analyzers remains unclear. To determine the performance of different platforms, we compared the consistency and accuracy between Roche Cobas 8000 and...

Descripción completa

Detalles Bibliográficos
Autores principales: Chen, Mingxing, Qin, Simeng, Yang, Sitao, Chen, Huaping, Lu, Liuyi, Qin, Xue
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Society of Medical Biochemists of Serbia, Belgrade 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9375536/
https://www.ncbi.nlm.nih.gov/pubmed/36042904
http://dx.doi.org/10.5937/jomb0-34328
_version_ 1784767985461755904
author Chen, Mingxing
Qin, Simeng
Yang, Sitao
Chen, Huaping
Lu, Liuyi
Qin, Xue
author_facet Chen, Mingxing
Qin, Simeng
Yang, Sitao
Chen, Huaping
Lu, Liuyi
Qin, Xue
author_sort Chen, Mingxing
collection PubMed
description BACKGROUND: The values of biomarkers play a central role in routine clinical decision-making. Whereas the performance of different automated chemical analyzers remains unclear. To determine the performance of different platforms, we compared the consistency and accuracy between Roche Cobas 8000 and Mindray BS2000M. METHODS: A total of 1869 remaining serum samples were collected. CK, LDH-1, RBP, Cys-C, IgA, IgM, and IgG were assessed using paired t-test, Passing-Bablok regression analysis, and Bland-Altman analysis according to CLSI EP5-A3. RESULTS: There were significant differences in the average bias of all items between the two machines (P<0.001). Because the 95% confidence interval of intercept A included 0, CK, LDH-1, Cys-C and IgG did not show systematic error in Passing-Bablok regression analysis. The confidence interval of 95% of the slope B in IgM contained 1, and there was no difference in the two measurements in IgM. Except for IgA, the r values and correlation coefficient of all items were higher than 0.91, which showed that the correlation and consistency were good. The Bland-Altman analysis showed that two instruments had more than 95% of the points apart from CK, LDH-1, and IgA. CONCLUSIONS: It can be considered that the two instruments have good correlation and consistency in CK, LDH-1, RBP, Cys-C, IgM, and IgG, and the two instruments are interchangeable and can replace each other.
format Online
Article
Text
id pubmed-9375536
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Society of Medical Biochemists of Serbia, Belgrade
record_format MEDLINE/PubMed
spelling pubmed-93755362022-08-29 Performance evaluation between two automated biochemical analyzer systems: Roche Cobas 8000 and Mindray BS2000M Chen, Mingxing Qin, Simeng Yang, Sitao Chen, Huaping Lu, Liuyi Qin, Xue J Med Biochem Original Paper BACKGROUND: The values of biomarkers play a central role in routine clinical decision-making. Whereas the performance of different automated chemical analyzers remains unclear. To determine the performance of different platforms, we compared the consistency and accuracy between Roche Cobas 8000 and Mindray BS2000M. METHODS: A total of 1869 remaining serum samples were collected. CK, LDH-1, RBP, Cys-C, IgA, IgM, and IgG were assessed using paired t-test, Passing-Bablok regression analysis, and Bland-Altman analysis according to CLSI EP5-A3. RESULTS: There were significant differences in the average bias of all items between the two machines (P<0.001). Because the 95% confidence interval of intercept A included 0, CK, LDH-1, Cys-C and IgG did not show systematic error in Passing-Bablok regression analysis. The confidence interval of 95% of the slope B in IgM contained 1, and there was no difference in the two measurements in IgM. Except for IgA, the r values and correlation coefficient of all items were higher than 0.91, which showed that the correlation and consistency were good. The Bland-Altman analysis showed that two instruments had more than 95% of the points apart from CK, LDH-1, and IgA. CONCLUSIONS: It can be considered that the two instruments have good correlation and consistency in CK, LDH-1, RBP, Cys-C, IgM, and IgG, and the two instruments are interchangeable and can replace each other. Society of Medical Biochemists of Serbia, Belgrade 2022-07-29 2022-07-29 /pmc/articles/PMC9375536/ /pubmed/36042904 http://dx.doi.org/10.5937/jomb0-34328 Text en 2022 Mingxing Chen, Simeng Qin, Sitao Yang, Huaping Chen, Liuyi Lu, Xue Qin, published by CEON/CEES https://creativecommons.org/licenses/by/4.0/This work is licensed under the Creative Commons Attribution 4.0 License.
spellingShingle Original Paper
Chen, Mingxing
Qin, Simeng
Yang, Sitao
Chen, Huaping
Lu, Liuyi
Qin, Xue
Performance evaluation between two automated biochemical analyzer systems: Roche Cobas 8000 and Mindray BS2000M
title Performance evaluation between two automated biochemical analyzer systems: Roche Cobas 8000 and Mindray BS2000M
title_full Performance evaluation between two automated biochemical analyzer systems: Roche Cobas 8000 and Mindray BS2000M
title_fullStr Performance evaluation between two automated biochemical analyzer systems: Roche Cobas 8000 and Mindray BS2000M
title_full_unstemmed Performance evaluation between two automated biochemical analyzer systems: Roche Cobas 8000 and Mindray BS2000M
title_short Performance evaluation between two automated biochemical analyzer systems: Roche Cobas 8000 and Mindray BS2000M
title_sort performance evaluation between two automated biochemical analyzer systems: roche cobas 8000 and mindray bs2000m
topic Original Paper
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9375536/
https://www.ncbi.nlm.nih.gov/pubmed/36042904
http://dx.doi.org/10.5937/jomb0-34328
work_keys_str_mv AT chenmingxing performanceevaluationbetweentwoautomatedbiochemicalanalyzersystemsrochecobas8000andmindraybs2000m
AT qinsimeng performanceevaluationbetweentwoautomatedbiochemicalanalyzersystemsrochecobas8000andmindraybs2000m
AT yangsitao performanceevaluationbetweentwoautomatedbiochemicalanalyzersystemsrochecobas8000andmindraybs2000m
AT chenhuaping performanceevaluationbetweentwoautomatedbiochemicalanalyzersystemsrochecobas8000andmindraybs2000m
AT luliuyi performanceevaluationbetweentwoautomatedbiochemicalanalyzersystemsrochecobas8000andmindraybs2000m
AT qinxue performanceevaluationbetweentwoautomatedbiochemicalanalyzersystemsrochecobas8000andmindraybs2000m