Performance evaluation between two automated biochemical analyzer systems: Roche Cobas 8000 and Mindray BS2000M
BACKGROUND: The values of biomarkers play a central role in routine clinical decision-making. Whereas the performance of different automated chemical analyzers remains unclear. To determine the performance of different platforms, we compared the consistency and accuracy between Roche Cobas 8000 and...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Society of Medical Biochemists of Serbia, Belgrade
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9375536/ https://www.ncbi.nlm.nih.gov/pubmed/36042904 http://dx.doi.org/10.5937/jomb0-34328 |
_version_ | 1784767985461755904 |
---|---|
author | Chen, Mingxing Qin, Simeng Yang, Sitao Chen, Huaping Lu, Liuyi Qin, Xue |
author_facet | Chen, Mingxing Qin, Simeng Yang, Sitao Chen, Huaping Lu, Liuyi Qin, Xue |
author_sort | Chen, Mingxing |
collection | PubMed |
description | BACKGROUND: The values of biomarkers play a central role in routine clinical decision-making. Whereas the performance of different automated chemical analyzers remains unclear. To determine the performance of different platforms, we compared the consistency and accuracy between Roche Cobas 8000 and Mindray BS2000M. METHODS: A total of 1869 remaining serum samples were collected. CK, LDH-1, RBP, Cys-C, IgA, IgM, and IgG were assessed using paired t-test, Passing-Bablok regression analysis, and Bland-Altman analysis according to CLSI EP5-A3. RESULTS: There were significant differences in the average bias of all items between the two machines (P<0.001). Because the 95% confidence interval of intercept A included 0, CK, LDH-1, Cys-C and IgG did not show systematic error in Passing-Bablok regression analysis. The confidence interval of 95% of the slope B in IgM contained 1, and there was no difference in the two measurements in IgM. Except for IgA, the r values and correlation coefficient of all items were higher than 0.91, which showed that the correlation and consistency were good. The Bland-Altman analysis showed that two instruments had more than 95% of the points apart from CK, LDH-1, and IgA. CONCLUSIONS: It can be considered that the two instruments have good correlation and consistency in CK, LDH-1, RBP, Cys-C, IgM, and IgG, and the two instruments are interchangeable and can replace each other. |
format | Online Article Text |
id | pubmed-9375536 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | Society of Medical Biochemists of Serbia, Belgrade |
record_format | MEDLINE/PubMed |
spelling | pubmed-93755362022-08-29 Performance evaluation between two automated biochemical analyzer systems: Roche Cobas 8000 and Mindray BS2000M Chen, Mingxing Qin, Simeng Yang, Sitao Chen, Huaping Lu, Liuyi Qin, Xue J Med Biochem Original Paper BACKGROUND: The values of biomarkers play a central role in routine clinical decision-making. Whereas the performance of different automated chemical analyzers remains unclear. To determine the performance of different platforms, we compared the consistency and accuracy between Roche Cobas 8000 and Mindray BS2000M. METHODS: A total of 1869 remaining serum samples were collected. CK, LDH-1, RBP, Cys-C, IgA, IgM, and IgG were assessed using paired t-test, Passing-Bablok regression analysis, and Bland-Altman analysis according to CLSI EP5-A3. RESULTS: There were significant differences in the average bias of all items between the two machines (P<0.001). Because the 95% confidence interval of intercept A included 0, CK, LDH-1, Cys-C and IgG did not show systematic error in Passing-Bablok regression analysis. The confidence interval of 95% of the slope B in IgM contained 1, and there was no difference in the two measurements in IgM. Except for IgA, the r values and correlation coefficient of all items were higher than 0.91, which showed that the correlation and consistency were good. The Bland-Altman analysis showed that two instruments had more than 95% of the points apart from CK, LDH-1, and IgA. CONCLUSIONS: It can be considered that the two instruments have good correlation and consistency in CK, LDH-1, RBP, Cys-C, IgM, and IgG, and the two instruments are interchangeable and can replace each other. Society of Medical Biochemists of Serbia, Belgrade 2022-07-29 2022-07-29 /pmc/articles/PMC9375536/ /pubmed/36042904 http://dx.doi.org/10.5937/jomb0-34328 Text en 2022 Mingxing Chen, Simeng Qin, Sitao Yang, Huaping Chen, Liuyi Lu, Xue Qin, published by CEON/CEES https://creativecommons.org/licenses/by/4.0/This work is licensed under the Creative Commons Attribution 4.0 License. |
spellingShingle | Original Paper Chen, Mingxing Qin, Simeng Yang, Sitao Chen, Huaping Lu, Liuyi Qin, Xue Performance evaluation between two automated biochemical analyzer systems: Roche Cobas 8000 and Mindray BS2000M |
title | Performance evaluation between two automated biochemical analyzer systems: Roche Cobas 8000 and Mindray BS2000M |
title_full | Performance evaluation between two automated biochemical analyzer systems: Roche Cobas 8000 and Mindray BS2000M |
title_fullStr | Performance evaluation between two automated biochemical analyzer systems: Roche Cobas 8000 and Mindray BS2000M |
title_full_unstemmed | Performance evaluation between two automated biochemical analyzer systems: Roche Cobas 8000 and Mindray BS2000M |
title_short | Performance evaluation between two automated biochemical analyzer systems: Roche Cobas 8000 and Mindray BS2000M |
title_sort | performance evaluation between two automated biochemical analyzer systems: roche cobas 8000 and mindray bs2000m |
topic | Original Paper |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9375536/ https://www.ncbi.nlm.nih.gov/pubmed/36042904 http://dx.doi.org/10.5937/jomb0-34328 |
work_keys_str_mv | AT chenmingxing performanceevaluationbetweentwoautomatedbiochemicalanalyzersystemsrochecobas8000andmindraybs2000m AT qinsimeng performanceevaluationbetweentwoautomatedbiochemicalanalyzersystemsrochecobas8000andmindraybs2000m AT yangsitao performanceevaluationbetweentwoautomatedbiochemicalanalyzersystemsrochecobas8000andmindraybs2000m AT chenhuaping performanceevaluationbetweentwoautomatedbiochemicalanalyzersystemsrochecobas8000andmindraybs2000m AT luliuyi performanceevaluationbetweentwoautomatedbiochemicalanalyzersystemsrochecobas8000andmindraybs2000m AT qinxue performanceevaluationbetweentwoautomatedbiochemicalanalyzersystemsrochecobas8000andmindraybs2000m |