Cargando…

Evaluation of a novel optical smartphone blood pressure application: a method comparison study against invasive arterial blood pressure monitoring in intensive care unit patients

BACKGROUND: Arterial hypertension is a worldwide public health problem. While it is currently diagnosed and monitored non-invasively using the oscillometric method, having the ability to measure blood pressure (BP) using a smartphone application could provide more widespread access to hypertension s...

Descripción completa

Detalles Bibliográficos
Autores principales: Desebbe, Olivier, Anas, Chbabou, Alexander, Brenton, Kouz, Karim, Knebel, Jean-Francois, Schoettker, Patrick, Creteur, Jacques, Vincent, Jean-Louis, Joosten, Alexandre
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9376576/
https://www.ncbi.nlm.nih.gov/pubmed/35971072
http://dx.doi.org/10.1186/s12871-022-01797-0
_version_ 1784768174873378816
author Desebbe, Olivier
Anas, Chbabou
Alexander, Brenton
Kouz, Karim
Knebel, Jean-Francois
Schoettker, Patrick
Creteur, Jacques
Vincent, Jean-Louis
Joosten, Alexandre
author_facet Desebbe, Olivier
Anas, Chbabou
Alexander, Brenton
Kouz, Karim
Knebel, Jean-Francois
Schoettker, Patrick
Creteur, Jacques
Vincent, Jean-Louis
Joosten, Alexandre
author_sort Desebbe, Olivier
collection PubMed
description BACKGROUND: Arterial hypertension is a worldwide public health problem. While it is currently diagnosed and monitored non-invasively using the oscillometric method, having the ability to measure blood pressure (BP) using a smartphone application could provide more widespread access to hypertension screening and monitoring. In this observational study in intensive care unit patients, we compared blood pressure values obtained using a new optical smartphone application (OptiBP™; test method) with arterial BP values obtained using a radial artery catheter (reference method) in order to help validate the technology. METHODS: We simultaneously measured three BP values every hour for five consecutive hours on two consecutive days using both the smartphone and arterial methods. Bland–Altman and error grid analyses were used for agreement analysis between both approaches. The performance of the smartphone application was investigated using the Association for the Advancement of Medical Instrumentation (AAMI) and the International Organization for Standardization (ISO) definitions, which require the bias ± SD between two technologies to be below 5 ± 8 mmHg. RESULTS: Among the 30 recruited patients, 22 patients had adequate OptiBP™ values and were thus analyzed. In the other 8 patients, no BP could be measured due to inadequate signals. The Bland–Altman analysis revealed a mean of the differences ± SD between both methods of 0.9 ± 7 mmHg for mean arterial pressure (MAP), 0.2 ± 14 mmHg for systolic arterial pressure (SAP), and 1.1 ± 6 mmHg for diastolic arterial pressure (DAP). Error grid analysis demonstrated that the proportions of measurement pairs in risk zones A to E were 88.8% (no risk), 10% (low risk), 1% (moderate risk), 0% (significant risk), and 0% (dangerous risk) for MAP and 88.4%, 8.6%, 3%, 0%, 0%, respectively, for SAP. CONCLUSIONS: This method comparison study revealed good agreement between BP values obtained using the OptiBP™ and those done invasively. The OptiBP™ fulfills the AAMI/ISO universal standards for MAP and DAP (but not SAP). Error grid showed that the most measurements (≥ 97%) were in risk zones A and B. TRIAL REGISTRATION: ClinicalTrials.gov registration: NCT04728477
format Online
Article
Text
id pubmed-9376576
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-93765762022-08-15 Evaluation of a novel optical smartphone blood pressure application: a method comparison study against invasive arterial blood pressure monitoring in intensive care unit patients Desebbe, Olivier Anas, Chbabou Alexander, Brenton Kouz, Karim Knebel, Jean-Francois Schoettker, Patrick Creteur, Jacques Vincent, Jean-Louis Joosten, Alexandre BMC Anesthesiol Research BACKGROUND: Arterial hypertension is a worldwide public health problem. While it is currently diagnosed and monitored non-invasively using the oscillometric method, having the ability to measure blood pressure (BP) using a smartphone application could provide more widespread access to hypertension screening and monitoring. In this observational study in intensive care unit patients, we compared blood pressure values obtained using a new optical smartphone application (OptiBP™; test method) with arterial BP values obtained using a radial artery catheter (reference method) in order to help validate the technology. METHODS: We simultaneously measured three BP values every hour for five consecutive hours on two consecutive days using both the smartphone and arterial methods. Bland–Altman and error grid analyses were used for agreement analysis between both approaches. The performance of the smartphone application was investigated using the Association for the Advancement of Medical Instrumentation (AAMI) and the International Organization for Standardization (ISO) definitions, which require the bias ± SD between two technologies to be below 5 ± 8 mmHg. RESULTS: Among the 30 recruited patients, 22 patients had adequate OptiBP™ values and were thus analyzed. In the other 8 patients, no BP could be measured due to inadequate signals. The Bland–Altman analysis revealed a mean of the differences ± SD between both methods of 0.9 ± 7 mmHg for mean arterial pressure (MAP), 0.2 ± 14 mmHg for systolic arterial pressure (SAP), and 1.1 ± 6 mmHg for diastolic arterial pressure (DAP). Error grid analysis demonstrated that the proportions of measurement pairs in risk zones A to E were 88.8% (no risk), 10% (low risk), 1% (moderate risk), 0% (significant risk), and 0% (dangerous risk) for MAP and 88.4%, 8.6%, 3%, 0%, 0%, respectively, for SAP. CONCLUSIONS: This method comparison study revealed good agreement between BP values obtained using the OptiBP™ and those done invasively. The OptiBP™ fulfills the AAMI/ISO universal standards for MAP and DAP (but not SAP). Error grid showed that the most measurements (≥ 97%) were in risk zones A and B. TRIAL REGISTRATION: ClinicalTrials.gov registration: NCT04728477 BioMed Central 2022-08-15 /pmc/articles/PMC9376576/ /pubmed/35971072 http://dx.doi.org/10.1186/s12871-022-01797-0 Text en © The Author(s) 2022 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) . The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/ (https://creativecommons.org/publicdomain/zero/1.0/) ) applies to the data made available in this article, unless otherwise stated in a credit line to the data.
spellingShingle Research
Desebbe, Olivier
Anas, Chbabou
Alexander, Brenton
Kouz, Karim
Knebel, Jean-Francois
Schoettker, Patrick
Creteur, Jacques
Vincent, Jean-Louis
Joosten, Alexandre
Evaluation of a novel optical smartphone blood pressure application: a method comparison study against invasive arterial blood pressure monitoring in intensive care unit patients
title Evaluation of a novel optical smartphone blood pressure application: a method comparison study against invasive arterial blood pressure monitoring in intensive care unit patients
title_full Evaluation of a novel optical smartphone blood pressure application: a method comparison study against invasive arterial blood pressure monitoring in intensive care unit patients
title_fullStr Evaluation of a novel optical smartphone blood pressure application: a method comparison study against invasive arterial blood pressure monitoring in intensive care unit patients
title_full_unstemmed Evaluation of a novel optical smartphone blood pressure application: a method comparison study against invasive arterial blood pressure monitoring in intensive care unit patients
title_short Evaluation of a novel optical smartphone blood pressure application: a method comparison study against invasive arterial blood pressure monitoring in intensive care unit patients
title_sort evaluation of a novel optical smartphone blood pressure application: a method comparison study against invasive arterial blood pressure monitoring in intensive care unit patients
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9376576/
https://www.ncbi.nlm.nih.gov/pubmed/35971072
http://dx.doi.org/10.1186/s12871-022-01797-0
work_keys_str_mv AT desebbeolivier evaluationofanovelopticalsmartphonebloodpressureapplicationamethodcomparisonstudyagainstinvasivearterialbloodpressuremonitoringinintensivecareunitpatients
AT anaschbabou evaluationofanovelopticalsmartphonebloodpressureapplicationamethodcomparisonstudyagainstinvasivearterialbloodpressuremonitoringinintensivecareunitpatients
AT alexanderbrenton evaluationofanovelopticalsmartphonebloodpressureapplicationamethodcomparisonstudyagainstinvasivearterialbloodpressuremonitoringinintensivecareunitpatients
AT kouzkarim evaluationofanovelopticalsmartphonebloodpressureapplicationamethodcomparisonstudyagainstinvasivearterialbloodpressuremonitoringinintensivecareunitpatients
AT knebeljeanfrancois evaluationofanovelopticalsmartphonebloodpressureapplicationamethodcomparisonstudyagainstinvasivearterialbloodpressuremonitoringinintensivecareunitpatients
AT schoettkerpatrick evaluationofanovelopticalsmartphonebloodpressureapplicationamethodcomparisonstudyagainstinvasivearterialbloodpressuremonitoringinintensivecareunitpatients
AT creteurjacques evaluationofanovelopticalsmartphonebloodpressureapplicationamethodcomparisonstudyagainstinvasivearterialbloodpressuremonitoringinintensivecareunitpatients
AT vincentjeanlouis evaluationofanovelopticalsmartphonebloodpressureapplicationamethodcomparisonstudyagainstinvasivearterialbloodpressuremonitoringinintensivecareunitpatients
AT joostenalexandre evaluationofanovelopticalsmartphonebloodpressureapplicationamethodcomparisonstudyagainstinvasivearterialbloodpressuremonitoringinintensivecareunitpatients