Evaluation of Reconstruction Algorithms to Validate the NEMA Phantom Results in Clinical Scenario – A Comparative Study Using Time-of-Flight versus Non-Time-of-Flight Positron Emission Tomography Imaging
OBJECTIVES: The objective is to standardize the reconstruction parameters for the time-of-flight (TOF) versus non-TOF positron-emission tomography/computed tomography (PET/CT) imaging data and validation of the same in a clinical setting. METHODS: The four spheres (10.0/13.0/17.0/22.0 mm) of the PET...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Wolters Kluwer - Medknow
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9380809/ https://www.ncbi.nlm.nih.gov/pubmed/35982821 http://dx.doi.org/10.4103/ijnm.ijnm_137_21 |
Sumario: | OBJECTIVES: The objective is to standardize the reconstruction parameters for the time-of-flight (TOF) versus non-TOF positron-emission tomography/computed tomography (PET/CT) imaging data and validation of the same in a clinical setting. METHODS: The four spheres (10.0/13.0/17.0/22.0 mm) of the PET phantom (NEMA IQ Nu 2-2001) were filled with four times higher activity of [(18)F]-NaF than the background (5.3kBq/mL). Imaging (image matrix – 128 × 128 × 47, 2 min, 3D model) was done using two different (TOF/non-TOF) PET scanners. Phantom data were reconstructed in TOF and non-TOF modes for lutetium–yttrium oxyorthosilicate and non-TOF mode for bismuth germanate-based PET scanners. The reconstructed data (by varying iteration and subsets) that provided the best image contrast and signal-to-noise ratio (SNR) were evaluated. The whole-body [(18)F]-fludeoxyglucose (FDG) PET/CT scans (7–8 frames; 2.0 min/frame) in 16 lymphoma patients were acquired at 60 min after injecting the radioactivity (370.0–444.0 MBq of [(18)F]-FDG. The clinical PET/CT data were reconstructed using phantom-derived reconstruction parameters and evaluated for image contrast and SNR of the detected lesions. RESULTS: TOF reconstruction at second iteration provided significantly (P ≤ 0.02) higher SNR (20.7) and contrast (contrast recovery coefficient/background variability = 3.21) for the smallest hot lesions (10.0 mm) in the phantom than the non-TOF system. Similarly, in patient data analysis for the selected FDG avid lesions, the SNR values were significantly (P = 0.02) higher (13.3 ± 6.49) in TOF than (11 ± 6.48) in non-TOF system. Further, the small (≤10.0 mm) lesions were seen more distinctly in TOF system. CONCLUSION: It is thus observed that TOF reconstruction converged faster than the non-TOF, and the applicability of the same may impact the image quality and interpretation in the clinical PET data. The validation of the phantom-based experimental reconstruction parameters to clinical PET imaging data is highly warranted. |
---|