Comparison of Ring Pulse Oximetry Using Reflective Photoplethysmography and PSG in the Detection of OSA in Chinese Adults: A Pilot Study
OBJECTIVE: A novel ring-worn oximeter (Circul) uses reflective photoplethysmography and automated signal processing to calculate oxygen desaturations. We evaluated the ability of Circul to detect obstructive sleep apnea in Chinese adults. METHODS: We recruited 207 Chinese Han subjects: 70% males, me...
Autores principales: | , , , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Dove
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9394522/ https://www.ncbi.nlm.nih.gov/pubmed/36003191 http://dx.doi.org/10.2147/NSS.S367400 |
Sumario: | OBJECTIVE: A novel ring-worn oximeter (Circul) uses reflective photoplethysmography and automated signal processing to calculate oxygen desaturations. We evaluated the ability of Circul to detect obstructive sleep apnea in Chinese adults. METHODS: We recruited 207 Chinese Han subjects: 70% males, mean age 48.2±14.7 years, mean BMI 27.6±4.8 kg/m2 and mean AHI 28.6±25.2 events/h. All participants underwent simultaneous polysomnography (PSG) and Circul testing in a sleep laboratory. Oxygen desaturation index (ODI), mean oxygen saturation (MSpO(2)), cumulative time at SpO2<90% (CT90), cumulative percentage of sleep time spent with SpO2<90% (CT90/TST) were derived and compared for the Circul and the PSG. RESULTS: The ODI was 25.3±24.5 events/h using PSG and 22.2±24.5 events/h using Circul (P<0.0001), with an intraclass correlation coefficient (ICC) of 0.884. CT90 and CT90/TST between the two methods were not different; the MSpO(2) level calculated by PSG was slightly lower than Circul, 95.0% (93.0–96.0%) vs 95.3% (93.9–96.6%), P<0.0001. Circul-ODI had a good correlation (r=0.91, p<0.0001) and close agreement with PSG-AHI (Bland-Altman analysis: Mean Difference 6.4, 95% CI −14.8 to 27.5 events/h). Using a threshold of AHI ≥5 events/h, the Circul had 87% sensitivity, 83% specificity, 5.09 positive likelihood ratio (LR+), 86% accuracy, and 0.929 area under the curve (AUC). CONCLUSION: Circul ring pulse oximetry can detect OSA with reasonable reliability. The Circul system is a reliable and comfortable choice for OSA assessment. |
---|