Comparison of the performance of Aptima HIV-1 Quant Dx Assay with Abbott RealTime HIV Assay for viral load monitoring using plasma and Dried Blood Spots collected in Kenya

INTRODUCTION: HIV-1 viral Load (VL) testing is recommended for the monitoring of antiretroviral treatment. Dried Blood Spots (DBS) are an effective sample type in resource limited settings, where safe phlebotomy and reliable shipping are hard to guarantee. In HIV high burden countries, high throughp...

Descripción completa

Detalles Bibliográficos
Autores principales: Mwau, Matilu, Schaffer, Sven, Kimani, Humphrey, Kasiano, Purity, Ogolla, Francis, Ajema, Elizabeth, Adoyo, Scriven, Nyairo, Ednah, Saleri, Norah, Nair, Sangeetha Vijaysri
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9394820/
https://www.ncbi.nlm.nih.gov/pubmed/35994447
http://dx.doi.org/10.1371/journal.pone.0269838
Descripción
Sumario:INTRODUCTION: HIV-1 viral Load (VL) testing is recommended for the monitoring of antiretroviral treatment. Dried Blood Spots (DBS) are an effective sample type in resource limited settings, where safe phlebotomy and reliable shipping are hard to guarantee. In HIV high burden countries, high throughput assays can improve access to testing services. The Hologic Aptima HIV-1 Quant Dx Assay (Aptima Assay) is a high throughput assay that runs on the CE-IVD approved Panther platform. The objectives of this study were to assess the performance characteristics of Aptima for VL monitoring using plasma and venous DBS specimens and to determine the stability of HIV-1 RNA in DBS. MATERIALS AND METHODS: This was a cross-sectional study of 2227 HIV infected adults visiting health facilities in Nairobi and Busia, Kenya. Each provided a venous blood sample; plasma was prepared from 1312 samples while paired DBS samples and plasma were prepared from the remaining 915 samples. The agreement between the Aptima assay and the Abbott RealTime HIV-1 Assay (Abbott RT) was analysed by comparing the HIV-1 VL in both assays at the medical decision point of 1000 copies/mL. To assess stability of HIV-1 RNA in DBS, VL in DBS spotted on day 0 were compared with that from the same DBS card after 21 days of storage at room temperature. RESULTS: Overall, 436 plasma samples had quantifiable results in both Aptima and Abbott RT. The agreement between the two assays at 1000 copies/mL was 97.48% with a Pearson’s correlation coefficient (r) of 0.9589 and gave a mean bias of 0.33 log copies/mL on Bland-Altman analysis. For fresh DBS, the agreement in both assays was 94.64% at 1000 copies/mL, with an r of 0.8692 and a mean bias of 0.35 log copies/mL. The overall agreement between DBS tested in Aptima on day 0 versus day 21 was 95.71%, with a mean bias of -0.154. CONCLUSION: The Aptima HIV-1 Quant Dx assay is an accurate test for VL monitoring of HIV-1 using DBS and plasma sample types in Kenya.