Cargando…
An evaluation of the use of EBT-XD film for SRS/SBRT commissioning of a 1.5 Tesla MR-Linac system
PURPOSE: The goal of this study was to evaluate the use of EBT-XD film for SRS/SBRT commissioning in a 1.5T hybrid MR-Linac (MRL). METHOD: The output factors (S(t)), from 1x1, 2x2, 3x3 cm(2), were measured with film in solid water. The results were compared with (1) the measurements by a PTW diamond...
Autores principales: | Lim, Seng Boh, Tyagi, Neelam, Subashi, Ergys, Liang, Jiayi, Chan, Maria |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9396448/ https://www.ncbi.nlm.nih.gov/pubmed/35189431 http://dx.doi.org/10.1016/j.ejmp.2022.02.012 |