The impacts of contact force, power and application time on ablation effect indicated by serial measurements of impedance drop in both conventional and high-power short-duration ablation settings of atrial fibrillation
BACKGROUND: This study aimed to clarify the interrelationship and additive effects of contact force (CF), power and application time in both conventional and high-power short-duration (HPSD) settings. METHODS: Among 38 patients with paroxysmal atrial fibrillation who underwent first-time pulmonary v...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Springer US
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9399015/ https://www.ncbi.nlm.nih.gov/pubmed/33891228 http://dx.doi.org/10.1007/s10840-021-00990-4 |
_version_ | 1784772428325453824 |
---|---|
author | Shi, Li-Bin Wang, Yu-Chuan Chu, Song-Yun De Bortoli, Alessandro Schuster, Peter Solheim, Eivind Chen, Jian |
author_facet | Shi, Li-Bin Wang, Yu-Chuan Chu, Song-Yun De Bortoli, Alessandro Schuster, Peter Solheim, Eivind Chen, Jian |
author_sort | Shi, Li-Bin |
collection | PubMed |
description | BACKGROUND: This study aimed to clarify the interrelationship and additive effects of contact force (CF), power and application time in both conventional and high-power short-duration (HPSD) settings. METHODS: Among 38 patients with paroxysmal atrial fibrillation who underwent first-time pulmonary vein isolation, 787 ablation points were collected at the beginning of the procedure at separate sites. Energy was applied for 60 s under power outputs of 25, 30 or 35 W (conventional group), or 10 s when using 50 W (HPSD group). An impedance drop (ID) of 10 Ω was regarded as a marker of adequate lesion formation. RESULTS: ID ≥ 10 Ω could not be achieved with CF < 5 g under any power setting. With CF ≥ 5 g, ID could be enhanced by increasing power output or prolonging ablation time. ID for 30 and 35 W was greater than for 25 W (p < 0.05). Ablation with 35 W resulted in greater ID than with 30 W only when CF of 10–20 g was applied for 20–40 s (p < 0.05). Under the same power output, ID increased with CF level at different time points. The higher the CF, the shorter the time needed to reach ID of 10 Ω and maximal ID. ID correlated well with ablation index under each power, except for lower ID values at 25 W. ID with 50 W for 10 s was equivalent to that with 25 W for 40 s, but lower than that with 30 W for 40 s or 35 W for 30 s. CONCLUSIONS: CF of at least 5 g is required for adequate ablation effect. With CF ≥ 5g, CF, power output, and ablation time can compensate for each other. Time to reach maximal ablation effect can be shortened by increasing CF or power. The effect of HPSD ablation with 50 W for 10 s is equivalent to conventional ablation with 25 W for 40 s and 30–35 W for 20–30 s in terms of ID. |
format | Online Article Text |
id | pubmed-9399015 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | Springer US |
record_format | MEDLINE/PubMed |
spelling | pubmed-93990152022-08-25 The impacts of contact force, power and application time on ablation effect indicated by serial measurements of impedance drop in both conventional and high-power short-duration ablation settings of atrial fibrillation Shi, Li-Bin Wang, Yu-Chuan Chu, Song-Yun De Bortoli, Alessandro Schuster, Peter Solheim, Eivind Chen, Jian J Interv Card Electrophysiol Article BACKGROUND: This study aimed to clarify the interrelationship and additive effects of contact force (CF), power and application time in both conventional and high-power short-duration (HPSD) settings. METHODS: Among 38 patients with paroxysmal atrial fibrillation who underwent first-time pulmonary vein isolation, 787 ablation points were collected at the beginning of the procedure at separate sites. Energy was applied for 60 s under power outputs of 25, 30 or 35 W (conventional group), or 10 s when using 50 W (HPSD group). An impedance drop (ID) of 10 Ω was regarded as a marker of adequate lesion formation. RESULTS: ID ≥ 10 Ω could not be achieved with CF < 5 g under any power setting. With CF ≥ 5 g, ID could be enhanced by increasing power output or prolonging ablation time. ID for 30 and 35 W was greater than for 25 W (p < 0.05). Ablation with 35 W resulted in greater ID than with 30 W only when CF of 10–20 g was applied for 20–40 s (p < 0.05). Under the same power output, ID increased with CF level at different time points. The higher the CF, the shorter the time needed to reach ID of 10 Ω and maximal ID. ID correlated well with ablation index under each power, except for lower ID values at 25 W. ID with 50 W for 10 s was equivalent to that with 25 W for 40 s, but lower than that with 30 W for 40 s or 35 W for 30 s. CONCLUSIONS: CF of at least 5 g is required for adequate ablation effect. With CF ≥ 5g, CF, power output, and ablation time can compensate for each other. Time to reach maximal ablation effect can be shortened by increasing CF or power. The effect of HPSD ablation with 50 W for 10 s is equivalent to conventional ablation with 25 W for 40 s and 30–35 W for 20–30 s in terms of ID. Springer US 2021-04-23 2022 /pmc/articles/PMC9399015/ /pubmed/33891228 http://dx.doi.org/10.1007/s10840-021-00990-4 Text en © The Author(s) 2021 https://creativecommons.org/licenses/by/4.0/Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) . |
spellingShingle | Article Shi, Li-Bin Wang, Yu-Chuan Chu, Song-Yun De Bortoli, Alessandro Schuster, Peter Solheim, Eivind Chen, Jian The impacts of contact force, power and application time on ablation effect indicated by serial measurements of impedance drop in both conventional and high-power short-duration ablation settings of atrial fibrillation |
title | The impacts of contact force, power and application time on ablation effect indicated by serial measurements of impedance drop in both conventional and high-power short-duration ablation settings of atrial fibrillation |
title_full | The impacts of contact force, power and application time on ablation effect indicated by serial measurements of impedance drop in both conventional and high-power short-duration ablation settings of atrial fibrillation |
title_fullStr | The impacts of contact force, power and application time on ablation effect indicated by serial measurements of impedance drop in both conventional and high-power short-duration ablation settings of atrial fibrillation |
title_full_unstemmed | The impacts of contact force, power and application time on ablation effect indicated by serial measurements of impedance drop in both conventional and high-power short-duration ablation settings of atrial fibrillation |
title_short | The impacts of contact force, power and application time on ablation effect indicated by serial measurements of impedance drop in both conventional and high-power short-duration ablation settings of atrial fibrillation |
title_sort | impacts of contact force, power and application time on ablation effect indicated by serial measurements of impedance drop in both conventional and high-power short-duration ablation settings of atrial fibrillation |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9399015/ https://www.ncbi.nlm.nih.gov/pubmed/33891228 http://dx.doi.org/10.1007/s10840-021-00990-4 |
work_keys_str_mv | AT shilibin theimpactsofcontactforcepowerandapplicationtimeonablationeffectindicatedbyserialmeasurementsofimpedancedropinbothconventionalandhighpowershortdurationablationsettingsofatrialfibrillation AT wangyuchuan theimpactsofcontactforcepowerandapplicationtimeonablationeffectindicatedbyserialmeasurementsofimpedancedropinbothconventionalandhighpowershortdurationablationsettingsofatrialfibrillation AT chusongyun theimpactsofcontactforcepowerandapplicationtimeonablationeffectindicatedbyserialmeasurementsofimpedancedropinbothconventionalandhighpowershortdurationablationsettingsofatrialfibrillation AT debortolialessandro theimpactsofcontactforcepowerandapplicationtimeonablationeffectindicatedbyserialmeasurementsofimpedancedropinbothconventionalandhighpowershortdurationablationsettingsofatrialfibrillation AT schusterpeter theimpactsofcontactforcepowerandapplicationtimeonablationeffectindicatedbyserialmeasurementsofimpedancedropinbothconventionalandhighpowershortdurationablationsettingsofatrialfibrillation AT solheimeivind theimpactsofcontactforcepowerandapplicationtimeonablationeffectindicatedbyserialmeasurementsofimpedancedropinbothconventionalandhighpowershortdurationablationsettingsofatrialfibrillation AT chenjian theimpactsofcontactforcepowerandapplicationtimeonablationeffectindicatedbyserialmeasurementsofimpedancedropinbothconventionalandhighpowershortdurationablationsettingsofatrialfibrillation AT shilibin impactsofcontactforcepowerandapplicationtimeonablationeffectindicatedbyserialmeasurementsofimpedancedropinbothconventionalandhighpowershortdurationablationsettingsofatrialfibrillation AT wangyuchuan impactsofcontactforcepowerandapplicationtimeonablationeffectindicatedbyserialmeasurementsofimpedancedropinbothconventionalandhighpowershortdurationablationsettingsofatrialfibrillation AT chusongyun impactsofcontactforcepowerandapplicationtimeonablationeffectindicatedbyserialmeasurementsofimpedancedropinbothconventionalandhighpowershortdurationablationsettingsofatrialfibrillation AT debortolialessandro impactsofcontactforcepowerandapplicationtimeonablationeffectindicatedbyserialmeasurementsofimpedancedropinbothconventionalandhighpowershortdurationablationsettingsofatrialfibrillation AT schusterpeter impactsofcontactforcepowerandapplicationtimeonablationeffectindicatedbyserialmeasurementsofimpedancedropinbothconventionalandhighpowershortdurationablationsettingsofatrialfibrillation AT solheimeivind impactsofcontactforcepowerandapplicationtimeonablationeffectindicatedbyserialmeasurementsofimpedancedropinbothconventionalandhighpowershortdurationablationsettingsofatrialfibrillation AT chenjian impactsofcontactforcepowerandapplicationtimeonablationeffectindicatedbyserialmeasurementsofimpedancedropinbothconventionalandhighpowershortdurationablationsettingsofatrialfibrillation |