Sensor-Augmented Insulin Pump with Predictive Low-Glucose Suspend (PLGS): Determining Optimal Settings of Pump and Sensor in a Multicenter Cohort of Patients with Type 1 Diabetes

INTRODUCTION: The use of predictive low-glucose suspend (PLGS) sensor-augmented pumps has been shown to lead to a significant reduction in hypoglycemic episodes in patients with type 1 diabetes (T1D), but their effects on hyperglycemia exposure are heterogeneous. The aim of this study was to determi...

Descripción completa

Detalles Bibliográficos
Autores principales: Joubert, Michael, Briant, Anaïs R., Kessler, Laurence, Fall-Mostaine, Fatéma, Dubois, Severine, Guerci, Bruno, Schoumacker-Ley, Laurène, Reznik, Yves, Parienti, Jean-Jacques
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Healthcare 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9399327/
https://www.ncbi.nlm.nih.gov/pubmed/35913656
http://dx.doi.org/10.1007/s13300-022-01302-3
_version_ 1784772494781054976
author Joubert, Michael
Briant, Anaïs R.
Kessler, Laurence
Fall-Mostaine, Fatéma
Dubois, Severine
Guerci, Bruno
Schoumacker-Ley, Laurène
Reznik, Yves
Parienti, Jean-Jacques
author_facet Joubert, Michael
Briant, Anaïs R.
Kessler, Laurence
Fall-Mostaine, Fatéma
Dubois, Severine
Guerci, Bruno
Schoumacker-Ley, Laurène
Reznik, Yves
Parienti, Jean-Jacques
author_sort Joubert, Michael
collection PubMed
description INTRODUCTION: The use of predictive low-glucose suspend (PLGS) sensor-augmented pumps has been shown to lead to a significant reduction in hypoglycemic episodes in patients with type 1 diabetes (T1D), but their effects on hyperglycemia exposure are heterogeneous. The aim of this study was to determine the settings of the Medtronic 640G system to obtain the optimal balance between occurrence of both hypoglycemia and hyperglycemia. METHODS: The hypo- and hyperglycemia area under the curve (AUC), as well as system settings [hypoglycemic threshold, mean insulin total daily dose (TDD), mean basal insulin percentage, and mean daily duration of PLGS] were collected between 2 and 12 times during 1 year in patients from four university hospital centers. Univariate/multivariate analyses and receiver operating characteristics (ROC) curves were performed to determine factors associated with hyper- and hypoglycemia AUC. RESULTS: A total of 864 observations were analyzed from 110 patients with T1D. Two preselected settings predictive of low hyperglycemia AUC were a basal insulin percentage < 52.0% [sensitivity (Se) = 0.66 and specificity (Sp) = 0.53] and a PLGS duration > 157.5 min/day (Se = 0.47 and Sp = 0.73). The preselected setting predictive of a low hypoglycemia AUC was a PLGS duration ≤ 174.4 min (Se = 0.83 and Sp = 0.51). Between-visit variation of PLGS and TDD was positively correlated (r = 0.61; p < 0.0001). CONCLUSION: The most important Medtronic 640G setting was the mean daily PLGS duration, where a value between 157.5 and 174.4 min/day was associated with the best reduction in both hypo- and hyperglycemia AUC. In this study, we showed that PLGS duration could be indirectly modified through total daily insulin dose adaptation. Trial Registration: This study is registered in clinicaltrials.gov (NCT 03047486).
format Online
Article
Text
id pubmed-9399327
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Springer Healthcare
record_format MEDLINE/PubMed
spelling pubmed-93993272022-08-25 Sensor-Augmented Insulin Pump with Predictive Low-Glucose Suspend (PLGS): Determining Optimal Settings of Pump and Sensor in a Multicenter Cohort of Patients with Type 1 Diabetes Joubert, Michael Briant, Anaïs R. Kessler, Laurence Fall-Mostaine, Fatéma Dubois, Severine Guerci, Bruno Schoumacker-Ley, Laurène Reznik, Yves Parienti, Jean-Jacques Diabetes Ther Original Research INTRODUCTION: The use of predictive low-glucose suspend (PLGS) sensor-augmented pumps has been shown to lead to a significant reduction in hypoglycemic episodes in patients with type 1 diabetes (T1D), but their effects on hyperglycemia exposure are heterogeneous. The aim of this study was to determine the settings of the Medtronic 640G system to obtain the optimal balance between occurrence of both hypoglycemia and hyperglycemia. METHODS: The hypo- and hyperglycemia area under the curve (AUC), as well as system settings [hypoglycemic threshold, mean insulin total daily dose (TDD), mean basal insulin percentage, and mean daily duration of PLGS] were collected between 2 and 12 times during 1 year in patients from four university hospital centers. Univariate/multivariate analyses and receiver operating characteristics (ROC) curves were performed to determine factors associated with hyper- and hypoglycemia AUC. RESULTS: A total of 864 observations were analyzed from 110 patients with T1D. Two preselected settings predictive of low hyperglycemia AUC were a basal insulin percentage < 52.0% [sensitivity (Se) = 0.66 and specificity (Sp) = 0.53] and a PLGS duration > 157.5 min/day (Se = 0.47 and Sp = 0.73). The preselected setting predictive of a low hypoglycemia AUC was a PLGS duration ≤ 174.4 min (Se = 0.83 and Sp = 0.51). Between-visit variation of PLGS and TDD was positively correlated (r = 0.61; p < 0.0001). CONCLUSION: The most important Medtronic 640G setting was the mean daily PLGS duration, where a value between 157.5 and 174.4 min/day was associated with the best reduction in both hypo- and hyperglycemia AUC. In this study, we showed that PLGS duration could be indirectly modified through total daily insulin dose adaptation. Trial Registration: This study is registered in clinicaltrials.gov (NCT 03047486). Springer Healthcare 2022-08-01 2022-09 /pmc/articles/PMC9399327/ /pubmed/35913656 http://dx.doi.org/10.1007/s13300-022-01302-3 Text en © The Author(s) 2022 https://creativecommons.org/licenses/by-nc/4.0/Open AccessThis article is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License, which permits any non-commercial use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by-nc/4.0/ (https://creativecommons.org/licenses/by-nc/4.0/) .
spellingShingle Original Research
Joubert, Michael
Briant, Anaïs R.
Kessler, Laurence
Fall-Mostaine, Fatéma
Dubois, Severine
Guerci, Bruno
Schoumacker-Ley, Laurène
Reznik, Yves
Parienti, Jean-Jacques
Sensor-Augmented Insulin Pump with Predictive Low-Glucose Suspend (PLGS): Determining Optimal Settings of Pump and Sensor in a Multicenter Cohort of Patients with Type 1 Diabetes
title Sensor-Augmented Insulin Pump with Predictive Low-Glucose Suspend (PLGS): Determining Optimal Settings of Pump and Sensor in a Multicenter Cohort of Patients with Type 1 Diabetes
title_full Sensor-Augmented Insulin Pump with Predictive Low-Glucose Suspend (PLGS): Determining Optimal Settings of Pump and Sensor in a Multicenter Cohort of Patients with Type 1 Diabetes
title_fullStr Sensor-Augmented Insulin Pump with Predictive Low-Glucose Suspend (PLGS): Determining Optimal Settings of Pump and Sensor in a Multicenter Cohort of Patients with Type 1 Diabetes
title_full_unstemmed Sensor-Augmented Insulin Pump with Predictive Low-Glucose Suspend (PLGS): Determining Optimal Settings of Pump and Sensor in a Multicenter Cohort of Patients with Type 1 Diabetes
title_short Sensor-Augmented Insulin Pump with Predictive Low-Glucose Suspend (PLGS): Determining Optimal Settings of Pump and Sensor in a Multicenter Cohort of Patients with Type 1 Diabetes
title_sort sensor-augmented insulin pump with predictive low-glucose suspend (plgs): determining optimal settings of pump and sensor in a multicenter cohort of patients with type 1 diabetes
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9399327/
https://www.ncbi.nlm.nih.gov/pubmed/35913656
http://dx.doi.org/10.1007/s13300-022-01302-3
work_keys_str_mv AT joubertmichael sensoraugmentedinsulinpumpwithpredictivelowglucosesuspendplgsdeterminingoptimalsettingsofpumpandsensorinamulticentercohortofpatientswithtype1diabetes
AT briantanaisr sensoraugmentedinsulinpumpwithpredictivelowglucosesuspendplgsdeterminingoptimalsettingsofpumpandsensorinamulticentercohortofpatientswithtype1diabetes
AT kesslerlaurence sensoraugmentedinsulinpumpwithpredictivelowglucosesuspendplgsdeterminingoptimalsettingsofpumpandsensorinamulticentercohortofpatientswithtype1diabetes
AT fallmostainefatema sensoraugmentedinsulinpumpwithpredictivelowglucosesuspendplgsdeterminingoptimalsettingsofpumpandsensorinamulticentercohortofpatientswithtype1diabetes
AT duboisseverine sensoraugmentedinsulinpumpwithpredictivelowglucosesuspendplgsdeterminingoptimalsettingsofpumpandsensorinamulticentercohortofpatientswithtype1diabetes
AT guercibruno sensoraugmentedinsulinpumpwithpredictivelowglucosesuspendplgsdeterminingoptimalsettingsofpumpandsensorinamulticentercohortofpatientswithtype1diabetes
AT schoumackerleylaurene sensoraugmentedinsulinpumpwithpredictivelowglucosesuspendplgsdeterminingoptimalsettingsofpumpandsensorinamulticentercohortofpatientswithtype1diabetes
AT reznikyves sensoraugmentedinsulinpumpwithpredictivelowglucosesuspendplgsdeterminingoptimalsettingsofpumpandsensorinamulticentercohortofpatientswithtype1diabetes
AT parientijeanjacques sensoraugmentedinsulinpumpwithpredictivelowglucosesuspendplgsdeterminingoptimalsettingsofpumpandsensorinamulticentercohortofpatientswithtype1diabetes