Analysis of Range and Use of a Hybrid Hydrogen Peroxide System for Biosafety Level 3 and Animal Biosafety Level 3 Agriculture Laboratory Decontamination

INTRODUCTION: The applications of fumigation and the challenges that high-containment facilities face in achieving effective large volume decontamination are well understood. The Biosecurity Research Institute at Kansas State University sought to evaluate a novel system within their biosafety level...

Descripción completa

Detalles Bibliográficos
Autores principales: Henneman, John R., McQuade, Elizabeth A., Sullivan, Rachael R., Downard, Jen, Thackrah, Ashley, Hislop, Meaghan
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Mary Ann Liebert, Inc., publishers 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9402250/
https://www.ncbi.nlm.nih.gov/pubmed/36032318
http://dx.doi.org/10.1089/apb.2021.0012
_version_ 1784773132586844160
author Henneman, John R.
McQuade, Elizabeth A.
Sullivan, Rachael R.
Downard, Jen
Thackrah, Ashley
Hislop, Meaghan
author_facet Henneman, John R.
McQuade, Elizabeth A.
Sullivan, Rachael R.
Downard, Jen
Thackrah, Ashley
Hislop, Meaghan
author_sort Henneman, John R.
collection PubMed
description INTRODUCTION: The applications of fumigation and the challenges that high-containment facilities face in achieving effective large volume decontamination are well understood. The Biosecurity Research Institute at Kansas State University sought to evaluate a novel system within their biosafety level 3 (BSL-3) and animal biosafety level 3 agriculture (ABSL-3Ag) facility. METHODS: The system chosen for this study is the CURIS(®) Hybrid Hydrogen Peroxide(TM) (HHP(TM)) system, comprising a mobile 36-pound (16 kg) device delivering a proprietary 7% hydrogen peroxide (H(2)O(2)) solution. To examine the system's efficacy in multiple laboratory settings, two BSL-3 laboratories (2,281 [65 m(3)] and 4,668 ft(3) [132 m(3)]) with dropped ceiling interstitial spaces and an ABSL-3Ag necropsy suite (44,212 ft(3) [1,252 m(3)]) with 21-foot (6.4 m) ceilings were selected. Biological indicators (BIs) of Geobacillus stearothermophilus (1.7 × 10(6) organisms) on steel spore carriers and H(2)O(2) chemical indicators (CIs) were used to provide validation. RESULTS: After cycle optimization, the smaller laboratory had a total of 60 BIs over two treatments that demonstrated a greater than 6-log reduction of bacterial spores. The larger laboratory (192 BIs) and the necropsy suite (206 BIs) had no BIs positive for spore growth when incubated at 60°C for 24 h per manufacturer's specifications. CONCLUSION: Overall successful results through multiple components of this study demonstrate that the HHP device, paired with the pulsed 7% H(2)O(2) solution, achieved efficacy regardless of variables in laboratory size and layout. Perceived challenges such as 21-ft (6.4 m) ceiling heights, active equipment, and difficult to access ceiling interstitial spaces proved unfounded. Given the successful sterilization of all challenged BIs, the HHP system presents a useful alternative for high level decontamination within BSL-3 and ABSL-3Ag facilities.
format Online
Article
Text
id pubmed-9402250
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Mary Ann Liebert, Inc., publishers
record_format MEDLINE/PubMed
spelling pubmed-94022502022-08-25 Analysis of Range and Use of a Hybrid Hydrogen Peroxide System for Biosafety Level 3 and Animal Biosafety Level 3 Agriculture Laboratory Decontamination Henneman, John R. McQuade, Elizabeth A. Sullivan, Rachael R. Downard, Jen Thackrah, Ashley Hislop, Meaghan Appl Biosaf Original Articles INTRODUCTION: The applications of fumigation and the challenges that high-containment facilities face in achieving effective large volume decontamination are well understood. The Biosecurity Research Institute at Kansas State University sought to evaluate a novel system within their biosafety level 3 (BSL-3) and animal biosafety level 3 agriculture (ABSL-3Ag) facility. METHODS: The system chosen for this study is the CURIS(®) Hybrid Hydrogen Peroxide(TM) (HHP(TM)) system, comprising a mobile 36-pound (16 kg) device delivering a proprietary 7% hydrogen peroxide (H(2)O(2)) solution. To examine the system's efficacy in multiple laboratory settings, two BSL-3 laboratories (2,281 [65 m(3)] and 4,668 ft(3) [132 m(3)]) with dropped ceiling interstitial spaces and an ABSL-3Ag necropsy suite (44,212 ft(3) [1,252 m(3)]) with 21-foot (6.4 m) ceilings were selected. Biological indicators (BIs) of Geobacillus stearothermophilus (1.7 × 10(6) organisms) on steel spore carriers and H(2)O(2) chemical indicators (CIs) were used to provide validation. RESULTS: After cycle optimization, the smaller laboratory had a total of 60 BIs over two treatments that demonstrated a greater than 6-log reduction of bacterial spores. The larger laboratory (192 BIs) and the necropsy suite (206 BIs) had no BIs positive for spore growth when incubated at 60°C for 24 h per manufacturer's specifications. CONCLUSION: Overall successful results through multiple components of this study demonstrate that the HHP device, paired with the pulsed 7% H(2)O(2) solution, achieved efficacy regardless of variables in laboratory size and layout. Perceived challenges such as 21-ft (6.4 m) ceiling heights, active equipment, and difficult to access ceiling interstitial spaces proved unfounded. Given the successful sterilization of all challenged BIs, the HHP system presents a useful alternative for high level decontamination within BSL-3 and ABSL-3Ag facilities. Mary Ann Liebert, Inc., publishers 2022-03-01 2022-03-15 /pmc/articles/PMC9402250/ /pubmed/36032318 http://dx.doi.org/10.1089/apb.2021.0012 Text en © John R. Henneman et al. 2022; Published by Mary Ann Liebert, Inc. https://creativecommons.org/licenses/by-nc/4.0/This Open Access article is distributed under the terms of the Creative Commons Attribution Noncommercial License [CC-BY-NC] (http://creativecommons.org/licenses/by-nc/4.0/ (https://creativecommons.org/licenses/by-nc/4.0/) ) which permits any noncommercial use, distribution, and reproduction in any medium, provided the original author(s) and the source are cited.
spellingShingle Original Articles
Henneman, John R.
McQuade, Elizabeth A.
Sullivan, Rachael R.
Downard, Jen
Thackrah, Ashley
Hislop, Meaghan
Analysis of Range and Use of a Hybrid Hydrogen Peroxide System for Biosafety Level 3 and Animal Biosafety Level 3 Agriculture Laboratory Decontamination
title Analysis of Range and Use of a Hybrid Hydrogen Peroxide System for Biosafety Level 3 and Animal Biosafety Level 3 Agriculture Laboratory Decontamination
title_full Analysis of Range and Use of a Hybrid Hydrogen Peroxide System for Biosafety Level 3 and Animal Biosafety Level 3 Agriculture Laboratory Decontamination
title_fullStr Analysis of Range and Use of a Hybrid Hydrogen Peroxide System for Biosafety Level 3 and Animal Biosafety Level 3 Agriculture Laboratory Decontamination
title_full_unstemmed Analysis of Range and Use of a Hybrid Hydrogen Peroxide System for Biosafety Level 3 and Animal Biosafety Level 3 Agriculture Laboratory Decontamination
title_short Analysis of Range and Use of a Hybrid Hydrogen Peroxide System for Biosafety Level 3 and Animal Biosafety Level 3 Agriculture Laboratory Decontamination
title_sort analysis of range and use of a hybrid hydrogen peroxide system for biosafety level 3 and animal biosafety level 3 agriculture laboratory decontamination
topic Original Articles
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9402250/
https://www.ncbi.nlm.nih.gov/pubmed/36032318
http://dx.doi.org/10.1089/apb.2021.0012
work_keys_str_mv AT hennemanjohnr analysisofrangeanduseofahybridhydrogenperoxidesystemforbiosafetylevel3andanimalbiosafetylevel3agriculturelaboratorydecontamination
AT mcquadeelizabetha analysisofrangeanduseofahybridhydrogenperoxidesystemforbiosafetylevel3andanimalbiosafetylevel3agriculturelaboratorydecontamination
AT sullivanrachaelr analysisofrangeanduseofahybridhydrogenperoxidesystemforbiosafetylevel3andanimalbiosafetylevel3agriculturelaboratorydecontamination
AT downardjen analysisofrangeanduseofahybridhydrogenperoxidesystemforbiosafetylevel3andanimalbiosafetylevel3agriculturelaboratorydecontamination
AT thackrahashley analysisofrangeanduseofahybridhydrogenperoxidesystemforbiosafetylevel3andanimalbiosafetylevel3agriculturelaboratorydecontamination
AT hislopmeaghan analysisofrangeanduseofahybridhydrogenperoxidesystemforbiosafetylevel3andanimalbiosafetylevel3agriculturelaboratorydecontamination