Cargando…

Utility of a percutaneous mechanical thrombectomy device in retrieval of an iatrogenic intravascular foreign body

Intravascular foreign bodies can result from endovascular procedures and from other medical implants and devices. A wide variety of techniques and devices have been described for the retrieval of such intravascular foreign bodies in reported studies. In the present report, we have described the case...

Descripción completa

Detalles Bibliográficos
Autor principal: Saleem, Taimur
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9424359/
https://www.ncbi.nlm.nih.gov/pubmed/36052214
http://dx.doi.org/10.1016/j.jvscit.2022.07.008
Descripción
Sumario:Intravascular foreign bodies can result from endovascular procedures and from other medical implants and devices. A wide variety of techniques and devices have been described for the retrieval of such intravascular foreign bodies in reported studies. In the present report, we have described the case of a patient with a symptomatic left innominate vein deep vein thrombosis who also had a retained catheter fragment from a fractured tunneled infusion catheter in the left innominate vein. Using the Inari ClotTriever system (Inari Medical, Irvine, CA), we were able to, not only restore venous outflow by treating the acute deep vein thrombosis, but also retrieve the fragments of the fractured catheter.