Initial Specimen Diversion Device Utilization Mitigates Blood Culture Contamination Across Regional Community Hospital and Acute Care Facility

A West Virginia regional community hospital incorporated an initial specimen diversion device (ISDD) into conventional blood culture protocol with the objective to bring the hospital-wide blood culture contamination (BCC) rate from a 3.06% preintervention rate to a target performance level below 1%....

Descripción completa

Detalles Bibliográficos
Autor principal: Povroznik, Mark D.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Lippincott Williams & Wilkins 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9426727/
https://www.ncbi.nlm.nih.gov/pubmed/35353719
http://dx.doi.org/10.1097/JMQ.0000000000000055
Descripción
Sumario:A West Virginia regional community hospital incorporated an initial specimen diversion device (ISDD) into conventional blood culture protocol with the objective to bring the hospital-wide blood culture contamination (BCC) rate from a 3.06% preintervention rate to a target performance level below 1%. Emergency department staff, laboratory phlebotomists, and nursing staff on acute-critical care floors were trained on ISDD (Steripath Gen2, Magnolia Medical Technologies, Inc., Seattle, WA) operating procedure and utilized the device for blood culture sample collection with adult patients from September 2020 through April 2021. Of 5642 blood culture sets collected hospital-wide, 4631 were collected with the ISDD, whereas the remaining sets were collected via the conventional method. The ISDD BCC rate of 0.78% differed from the conventional method BCC rate of 4.06% observed during the intervention period (chi-squared test P < 0.00001). The ISDD group attained a sub-1% BCC rate to satisfy the intervention objective.