Evaluation of the precision of navigation-assisted endoscopy according to the navigation tool setup and the type of endoscopes

OBJECT: Preoperative image-based neuronavigation-assisted endoscopy during intracranial procedures is gaining great interest. This study aimed to analyze the precision of navigation-assisted endoscopy according to the navigation setup, the type of optic and its working angulation. METHODS: A custom-...

Descripción completa

Detalles Bibliográficos
Autores principales: Chavaz, Lara, Davidovic, Alioucha, Meling, Torstein R., Momjian, Shahan, Schaller, Karl, Bijlenga, Philippe, Haemmerli, Julien
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Vienna 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9427865/
https://www.ncbi.nlm.nih.gov/pubmed/35764694
http://dx.doi.org/10.1007/s00701-022-05276-w
Descripción
Sumario:OBJECT: Preoperative image-based neuronavigation-assisted endoscopy during intracranial procedures is gaining great interest. This study aimed to analyze the precision of navigation-assisted endoscopy according to the navigation setup, the type of optic and its working angulation. METHODS: A custom-made box with four screws was referenced. The navigation-assisted endoscope was aligned on the screws (targets). The precision on the navigation screen was defined as the virtual distance-to-target between the tip of the endoscope and the center of the screws. Three modifiers were assessed: (1) the distance D between the box and the reference array (CLOSE 13 cm – MIDDLE 30 cm – FAR 53 cm), (2) the distance between the tip of the endoscope and the navigation array on the endoscope (close 5 cm – middle 10 cm – far 20 cm), (3) the working angulation of the endoscope (0°-endoscope and 30°-endoscope angled at 90° and 45° with the box). RESULTS: The median precision was 1.3 mm (Q1: 1.1; Q3: 1.7) with the best setting CLOSE/close. The best setting in surgical condition (CLOSE/far) showed a distance-to-target of 2.3 mm (Q1: 1.9; Q3: 2.5). The distance D was correlated to the precision (Spearman rho = 0.82), but not the distance d (Spearman rho = 0.04). The type of optic and its angulation with the box were also correlated to the precision (Spearman rho =  − 0.37). The best setting was the use of a 30°-endoscope angled at 45° (1.4 mm (Q1: 1.0; Q3: 1.9)). CONCLUSION: Navigated-assisted endoscopy is feasible and offers a good precision. The navigation setup should be optimized, reducing the risk of inadvertent perifocal damage.