Image quality and subject experience of quiet T1-weighted 7-T brain imaging using a silent gradient coil
OBJECTIVES: Acoustic noise in magnetic resonance imaging (MRI) negatively impacts patients. We assessed a silent gradient coil switched at 20 kHz combined with a T(1)-weighted magnetisation prepared rapid gradient-echo (MPRAGE) sequence at 7 T. METHODS: Five healthy subjects (21–29 years; three fema...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Springer Vienna
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9428090/ https://www.ncbi.nlm.nih.gov/pubmed/36042139 http://dx.doi.org/10.1186/s41747-022-00293-x |
_version_ | 1784779037418192896 |
---|---|
author | Jacobs, Sarah M. Versteeg, Edwin van der Kolk, Anja G. Visser, Leonie N. C. Oliveira, Ícaro A. F. van Maren, Emiel Klomp, Dennis W. J. Siero, Jeroen C. W. |
author_facet | Jacobs, Sarah M. Versteeg, Edwin van der Kolk, Anja G. Visser, Leonie N. C. Oliveira, Ícaro A. F. van Maren, Emiel Klomp, Dennis W. J. Siero, Jeroen C. W. |
author_sort | Jacobs, Sarah M. |
collection | PubMed |
description | OBJECTIVES: Acoustic noise in magnetic resonance imaging (MRI) negatively impacts patients. We assessed a silent gradient coil switched at 20 kHz combined with a T(1)-weighted magnetisation prepared rapid gradient-echo (MPRAGE) sequence at 7 T. METHODS: Five healthy subjects (21–29 years; three females) without previous 7-T MRI experience underwent both a quiet MPRAGE (Q-MPRAGE) and conventional MPRAGE (C-MPRAGE) sequence twice. Image quality was assessed quantitatively, and qualitatively by two neuroradiologists. Sound level was measured objectively and rated subjectively on a 0 to 10 scale by all subjects immediately following each sequence and after the whole examination (delayed). All subjects also reported comfort level, overall experience and willingness to undergo the sequence again. RESULTS: Compared to C-MPRAGE, Q-MPRAGE showed higher signal-to-noise ratio (10%; p = 0.012) and lower contrast-to-noise ratio (20%; p < 0.001) as well as acceptable to good image quality. Q-MPRAGE produced 27 dB lower sound level (76 versus 103 dB). Subjects reported lower sound level for Q-MPRAGE both immediate (4.4 ± 1.4 versus 6.4 ± 1.3; p = 0.007) and delayed (4.6 ± 1.4 versus 6.3 ± 1.3; p = 0.005), while they rated comfort level (7.4 ± 1.0 versus 6.1 ± 1.7; p = 0.016) and overall experience (7.6 ± 1.0 versus 6.0 ± 0.9; p = 0.005) higher. Willingness to undergo the sequence again was also higher, however not significantly (8.1 ± 1.0 versus 7.2 ± 1.3; p = 0.066). CONCLUSION: Q-MPRAGE using a silent gradient coil reduced sound level by 27 dB compared to C-MPRAGE at 7 T while featuring acceptable-to-good image quality and a quieter and more pleasant subject experience. |
format | Online Article Text |
id | pubmed-9428090 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | Springer Vienna |
record_format | MEDLINE/PubMed |
spelling | pubmed-94280902022-09-01 Image quality and subject experience of quiet T1-weighted 7-T brain imaging using a silent gradient coil Jacobs, Sarah M. Versteeg, Edwin van der Kolk, Anja G. Visser, Leonie N. C. Oliveira, Ícaro A. F. van Maren, Emiel Klomp, Dennis W. J. Siero, Jeroen C. W. Eur Radiol Exp Original Article OBJECTIVES: Acoustic noise in magnetic resonance imaging (MRI) negatively impacts patients. We assessed a silent gradient coil switched at 20 kHz combined with a T(1)-weighted magnetisation prepared rapid gradient-echo (MPRAGE) sequence at 7 T. METHODS: Five healthy subjects (21–29 years; three females) without previous 7-T MRI experience underwent both a quiet MPRAGE (Q-MPRAGE) and conventional MPRAGE (C-MPRAGE) sequence twice. Image quality was assessed quantitatively, and qualitatively by two neuroradiologists. Sound level was measured objectively and rated subjectively on a 0 to 10 scale by all subjects immediately following each sequence and after the whole examination (delayed). All subjects also reported comfort level, overall experience and willingness to undergo the sequence again. RESULTS: Compared to C-MPRAGE, Q-MPRAGE showed higher signal-to-noise ratio (10%; p = 0.012) and lower contrast-to-noise ratio (20%; p < 0.001) as well as acceptable to good image quality. Q-MPRAGE produced 27 dB lower sound level (76 versus 103 dB). Subjects reported lower sound level for Q-MPRAGE both immediate (4.4 ± 1.4 versus 6.4 ± 1.3; p = 0.007) and delayed (4.6 ± 1.4 versus 6.3 ± 1.3; p = 0.005), while they rated comfort level (7.4 ± 1.0 versus 6.1 ± 1.7; p = 0.016) and overall experience (7.6 ± 1.0 versus 6.0 ± 0.9; p = 0.005) higher. Willingness to undergo the sequence again was also higher, however not significantly (8.1 ± 1.0 versus 7.2 ± 1.3; p = 0.066). CONCLUSION: Q-MPRAGE using a silent gradient coil reduced sound level by 27 dB compared to C-MPRAGE at 7 T while featuring acceptable-to-good image quality and a quieter and more pleasant subject experience. Springer Vienna 2022-08-31 /pmc/articles/PMC9428090/ /pubmed/36042139 http://dx.doi.org/10.1186/s41747-022-00293-x Text en © The Author(s) under exclusive licence to European Society of Radiology 2022 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) . |
spellingShingle | Original Article Jacobs, Sarah M. Versteeg, Edwin van der Kolk, Anja G. Visser, Leonie N. C. Oliveira, Ícaro A. F. van Maren, Emiel Klomp, Dennis W. J. Siero, Jeroen C. W. Image quality and subject experience of quiet T1-weighted 7-T brain imaging using a silent gradient coil |
title | Image quality and subject experience of quiet T1-weighted 7-T brain imaging using a silent gradient coil |
title_full | Image quality and subject experience of quiet T1-weighted 7-T brain imaging using a silent gradient coil |
title_fullStr | Image quality and subject experience of quiet T1-weighted 7-T brain imaging using a silent gradient coil |
title_full_unstemmed | Image quality and subject experience of quiet T1-weighted 7-T brain imaging using a silent gradient coil |
title_short | Image quality and subject experience of quiet T1-weighted 7-T brain imaging using a silent gradient coil |
title_sort | image quality and subject experience of quiet t1-weighted 7-t brain imaging using a silent gradient coil |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9428090/ https://www.ncbi.nlm.nih.gov/pubmed/36042139 http://dx.doi.org/10.1186/s41747-022-00293-x |
work_keys_str_mv | AT jacobssarahm imagequalityandsubjectexperienceofquiett1weighted7tbrainimagingusingasilentgradientcoil AT versteegedwin imagequalityandsubjectexperienceofquiett1weighted7tbrainimagingusingasilentgradientcoil AT vanderkolkanjag imagequalityandsubjectexperienceofquiett1weighted7tbrainimagingusingasilentgradientcoil AT visserleonienc imagequalityandsubjectexperienceofquiett1weighted7tbrainimagingusingasilentgradientcoil AT oliveiraicaroaf imagequalityandsubjectexperienceofquiett1weighted7tbrainimagingusingasilentgradientcoil AT vanmarenemiel imagequalityandsubjectexperienceofquiett1weighted7tbrainimagingusingasilentgradientcoil AT klompdenniswj imagequalityandsubjectexperienceofquiett1weighted7tbrainimagingusingasilentgradientcoil AT sierojeroencw imagequalityandsubjectexperienceofquiett1weighted7tbrainimagingusingasilentgradientcoil |