Cargando…
Use of roller pump in venovenous extracorporeal membrane oxygenation as an emergency rescue procedure
A 41 years old male patient, having acute respiratory distress syndrome (ARDS) due to viral pneumonia, was put on venovenous (VV) extracorporeal membrane oxygenation (ECMO). After 3 h of initiation, the ECMO pump malfunctioned. The patient was initially managed on a hand crank for 5 h. However, as a...