Evaluation of the Aptima HBV Quant Assay Compared to Abbott RealTime M2000 HBV Quant Assay and Procleix Ultrio Plus dHBV Assay in Plasma Samples
Analytical performance of hepatitis B virus (HBV) DNA quantitative assay is critical for screening infection and initiating and monitoring antiviral treatment. In this study, the limit of detection (LoD) and linearity of Aptima HBV Quant assay were evaluated, and analytical performance was compared...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
American Society for Microbiology
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9431630/ https://www.ncbi.nlm.nih.gov/pubmed/35880868 http://dx.doi.org/10.1128/spectrum.01761-22 |
_version_ | 1784780108643434496 |
---|---|
author | Yan, Ying Chang, Le Ji, Huimin Han, Yang Wang, Lunan |
author_facet | Yan, Ying Chang, Le Ji, Huimin Han, Yang Wang, Lunan |
author_sort | Yan, Ying |
collection | PubMed |
description | Analytical performance of hepatitis B virus (HBV) DNA quantitative assay is critical for screening infection and initiating and monitoring antiviral treatment. In this study, the limit of detection (LoD) and linearity of Aptima HBV Quant assay were evaluated, and analytical performance was compared with that of the Abbott RealTime M2000 HBV Quant assay and the Procleix Ultrio Plus dHBV assay in plasma samples. The LoDs for genotypes B, C, and D plasma samples were 2.139 (1.531, 4.520), 3.120 (2.140, 7.373), and 3.330 (2.589, 4.907) IU/mL, respectively. The R(2) value fitted by linear regression of serially diluted samples less than 2,000 IU/mL was above 0.9. There was no difference in positive rate between Aptima and Abbott or between Aptima and Procleix. Quantitative results of Aptima and Abbott showed good correlation with an r of >0.9 using Spearman analysis, while the quantitative results of Aptima were slightly lower than those of Abbott. Usual mutations in the HBV S region had no impact on Aptima assay. This study showed that Aptima is a dual-targeted transcription-mediated amplification (TMA) assay suitable for HBV DNA detection in clinical practice, with quantitative performance comparable to that of the Abbott RealTime M2000 HBV Quant assay and qualitative performance comparable to that of the Procleix Ultrio Plus dHBV assay. IMPORTANCE The Aptima HBV Quant assay (Hologic Inc., San Diego, CA, USA) is a dual-target real-time transcription-mediated amplification (RT-TMA) assay. This study aims to evaluate whether this assay is suitable for HBV DNA detection. As a result, the assay showed high sensitivity with LoDs below 3.5 IU/mL. The amplification efficiency of Aptima for samples below 2,000 IU/mL is adequate for clinical practice, with an R(2) of >0.9 fitted by linear regression. Usual mutations in the HBV S region did not affect the performance of Aptima. Moreover, its performance was comparable to the widely used Abbott RealTime M2000 HBV Quant assay for detecting HBV DNA in plasma specimens. Although not indicated for use as a diagnostic or blood screening assay, the Aptima HBV Quant assay demonstrated comparable qualitative performance to the Procleix Ultrio Plus dHBV system. |
format | Online Article Text |
id | pubmed-9431630 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | American Society for Microbiology |
record_format | MEDLINE/PubMed |
spelling | pubmed-94316302022-09-01 Evaluation of the Aptima HBV Quant Assay Compared to Abbott RealTime M2000 HBV Quant Assay and Procleix Ultrio Plus dHBV Assay in Plasma Samples Yan, Ying Chang, Le Ji, Huimin Han, Yang Wang, Lunan Microbiol Spectr Research Article Analytical performance of hepatitis B virus (HBV) DNA quantitative assay is critical for screening infection and initiating and monitoring antiviral treatment. In this study, the limit of detection (LoD) and linearity of Aptima HBV Quant assay were evaluated, and analytical performance was compared with that of the Abbott RealTime M2000 HBV Quant assay and the Procleix Ultrio Plus dHBV assay in plasma samples. The LoDs for genotypes B, C, and D plasma samples were 2.139 (1.531, 4.520), 3.120 (2.140, 7.373), and 3.330 (2.589, 4.907) IU/mL, respectively. The R(2) value fitted by linear regression of serially diluted samples less than 2,000 IU/mL was above 0.9. There was no difference in positive rate between Aptima and Abbott or between Aptima and Procleix. Quantitative results of Aptima and Abbott showed good correlation with an r of >0.9 using Spearman analysis, while the quantitative results of Aptima were slightly lower than those of Abbott. Usual mutations in the HBV S region had no impact on Aptima assay. This study showed that Aptima is a dual-targeted transcription-mediated amplification (TMA) assay suitable for HBV DNA detection in clinical practice, with quantitative performance comparable to that of the Abbott RealTime M2000 HBV Quant assay and qualitative performance comparable to that of the Procleix Ultrio Plus dHBV assay. IMPORTANCE The Aptima HBV Quant assay (Hologic Inc., San Diego, CA, USA) is a dual-target real-time transcription-mediated amplification (RT-TMA) assay. This study aims to evaluate whether this assay is suitable for HBV DNA detection. As a result, the assay showed high sensitivity with LoDs below 3.5 IU/mL. The amplification efficiency of Aptima for samples below 2,000 IU/mL is adequate for clinical practice, with an R(2) of >0.9 fitted by linear regression. Usual mutations in the HBV S region did not affect the performance of Aptima. Moreover, its performance was comparable to the widely used Abbott RealTime M2000 HBV Quant assay for detecting HBV DNA in plasma specimens. Although not indicated for use as a diagnostic or blood screening assay, the Aptima HBV Quant assay demonstrated comparable qualitative performance to the Procleix Ultrio Plus dHBV system. American Society for Microbiology 2022-07-26 /pmc/articles/PMC9431630/ /pubmed/35880868 http://dx.doi.org/10.1128/spectrum.01761-22 Text en Copyright © 2022 Yan et al. https://creativecommons.org/licenses/by/4.0/This is an open-access article distributed under the terms of the Creative Commons Attribution 4.0 International license (https://creativecommons.org/licenses/by/4.0/) . |
spellingShingle | Research Article Yan, Ying Chang, Le Ji, Huimin Han, Yang Wang, Lunan Evaluation of the Aptima HBV Quant Assay Compared to Abbott RealTime M2000 HBV Quant Assay and Procleix Ultrio Plus dHBV Assay in Plasma Samples |
title | Evaluation of the Aptima HBV Quant Assay Compared to Abbott RealTime M2000 HBV Quant Assay and Procleix Ultrio Plus dHBV Assay in Plasma Samples |
title_full | Evaluation of the Aptima HBV Quant Assay Compared to Abbott RealTime M2000 HBV Quant Assay and Procleix Ultrio Plus dHBV Assay in Plasma Samples |
title_fullStr | Evaluation of the Aptima HBV Quant Assay Compared to Abbott RealTime M2000 HBV Quant Assay and Procleix Ultrio Plus dHBV Assay in Plasma Samples |
title_full_unstemmed | Evaluation of the Aptima HBV Quant Assay Compared to Abbott RealTime M2000 HBV Quant Assay and Procleix Ultrio Plus dHBV Assay in Plasma Samples |
title_short | Evaluation of the Aptima HBV Quant Assay Compared to Abbott RealTime M2000 HBV Quant Assay and Procleix Ultrio Plus dHBV Assay in Plasma Samples |
title_sort | evaluation of the aptima hbv quant assay compared to abbott realtime m2000 hbv quant assay and procleix ultrio plus dhbv assay in plasma samples |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9431630/ https://www.ncbi.nlm.nih.gov/pubmed/35880868 http://dx.doi.org/10.1128/spectrum.01761-22 |
work_keys_str_mv | AT yanying evaluationoftheaptimahbvquantassaycomparedtoabbottrealtimem2000hbvquantassayandprocleixultrioplusdhbvassayinplasmasamples AT changle evaluationoftheaptimahbvquantassaycomparedtoabbottrealtimem2000hbvquantassayandprocleixultrioplusdhbvassayinplasmasamples AT jihuimin evaluationoftheaptimahbvquantassaycomparedtoabbottrealtimem2000hbvquantassayandprocleixultrioplusdhbvassayinplasmasamples AT hanyang evaluationoftheaptimahbvquantassaycomparedtoabbottrealtimem2000hbvquantassayandprocleixultrioplusdhbvassayinplasmasamples AT wanglunan evaluationoftheaptimahbvquantassaycomparedtoabbottrealtimem2000hbvquantassayandprocleixultrioplusdhbvassayinplasmasamples |