Circumferential and Depth Variations in Tissue Dielectric Constant Values as Indices of Lower Leg Localized Skin Water

The primary goal of this study was to quantify circumferential (medial, anterior, lateral) and measurement depth variations (0.5 mm, 1.5 mm, 2.5 mm) in tissue dielectric constant (TDC) values as an aid to their use to assess the presence and progression of lower extremity edema and lymphedema. Measu...

Descripción completa

Detalles Bibliográficos
Autor principal: Mayrovitz, Harvey N
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Cureus 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9433792/
https://www.ncbi.nlm.nih.gov/pubmed/36059342
http://dx.doi.org/10.7759/cureus.27617
Descripción
Sumario:The primary goal of this study was to quantify circumferential (medial, anterior, lateral) and measurement depth variations (0.5 mm, 1.5 mm, 2.5 mm) in tissue dielectric constant (TDC) values as an aid to their use to assess the presence and progression of lower extremity edema and lymphedema. Measurements were done in 30 healthy non-edematous women to provide reference data to estimate expected values and thresholds when evaluating clinical edematous or lymphedematous conditions. A second goal was to determine the extent to which TDC values evaluated at lower leg sites depend on body mass index (BMI). The study protocol (#12180901) was approved by the university’s institutional review board and subjects were evaluated after signing an approved informed consent. The study group had an age range of 19-54 years with a mean age and SD of 30.6 ± 10.1 years and had a BMI between 18.1-44.1 Kg/m(2) with a mean BMI and SD of 24.5 ± 5.4 Kg/m(2). The main findings show that at the three circumferential sites (medial, anterior, and lateral) located eight cm from the mid-malleolus, there are small but statistically significant differences in TDC values at every measurement depth (0.5 mm, 1.5 mm and 2.5 mm). For each depth, the maximum difference occurs between the medial and lateral locations with lateral locations having a greater TDC value at all depths. Despite the wide range of BMI values of the subjects evaluated, no significant relationship between TDC value and BMI was detected. It is concluded that TDC measurements in the lower leg reveal statistically significant differences among circumferential sites and measurement depths that should be considered when evaluating or tracking lower extremity tissue water changes associated with edema, lymphedema or other conditions related to skin water. The absolute values of these non-edematous TDC values herein determined may provide a basis for calculating TDC thresholds applicable to edematous or lymphedematous lower leg conditions.