Swept-source OCT for corneal graft quantitative evaluation in the eye bank and the correlation of the measurements to pre-excision values

Quantitative evaluation of the human corneal grafts stored in the tissue banks is usually limited to endothelial cell density and central thickness. Swept-source OCT (SS-OCT) is capable of measuring the central curvatures of the corneal tissue prepared for transplantation without loss of sterileness...

Descripción completa

Detalles Bibliográficos
Autores principales: Kaluzny, Bartlomiej J., Mlyniuk, Patryk, Liberski, Slawomir, Jimenez-Villar, Alfonso, Machalińska, Anna, Grulkowski, Ireneusz
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Nature Publishing Group UK 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9437096/
https://www.ncbi.nlm.nih.gov/pubmed/36050445
http://dx.doi.org/10.1038/s41598-022-19225-1
_version_ 1784781518148730880
author Kaluzny, Bartlomiej J.
Mlyniuk, Patryk
Liberski, Slawomir
Jimenez-Villar, Alfonso
Machalińska, Anna
Grulkowski, Ireneusz
author_facet Kaluzny, Bartlomiej J.
Mlyniuk, Patryk
Liberski, Slawomir
Jimenez-Villar, Alfonso
Machalińska, Anna
Grulkowski, Ireneusz
author_sort Kaluzny, Bartlomiej J.
collection PubMed
description Quantitative evaluation of the human corneal grafts stored in the tissue banks is usually limited to endothelial cell density and central thickness. Swept-source OCT (SS-OCT) is capable of measuring the central curvatures of the corneal tissue prepared for transplantation without loss of sterileness, providing insights on its refractive state. The aim of the paper is to compare in vitro SS-OCT measurements with pre-excision values. Hand-held keratometry and ultrasound pachymetry was performed on 22 corneas before excision of corneoscleral button and insertion in the vial with Eusol-C solution (AlchimiaS.r.l, Nicolò, Italy). After 12 to 36 h of hypothermic storage the corneas were examined within the vials with custom built SS-OCT system maintaining a sterile environment. The anterior and posterior central curvatures, and central corneal thickness (CCT) were measured. Rotation of the corneoscleral button was controlled by making a 6-o'clock mark during excision. Mean pre-excision CCT was 626.45 ± 28.71 µm and 468.05 ± 52.96 µm when measured with SS OCT (r = 0.55; p < 0.001). Respective values for average keratometry were 7.74 ± 0.39 mm and 7.92 ± 0.57 mm (r = 0.6; p = 0.22). Although high differences were observed in corneal thickness, keratometric radius of curvature at the flat (r = 0.42; p < 0.001) and steep (r = 0.62; p = 0.014) meridian of the anterior corneal surface, as well as corneal anterior astigmatism (r = 0.3; p < 0.001), showed good correlation with pre-excision values. SS-OCT is capable of providing quantitative evaluation of the human corneal grafts in hypothermic storage. Good correlation between curvature measurements before excision and during banking in the vial indicates its clinical utility.
format Online
Article
Text
id pubmed-9437096
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Nature Publishing Group UK
record_format MEDLINE/PubMed
spelling pubmed-94370962022-09-03 Swept-source OCT for corneal graft quantitative evaluation in the eye bank and the correlation of the measurements to pre-excision values Kaluzny, Bartlomiej J. Mlyniuk, Patryk Liberski, Slawomir Jimenez-Villar, Alfonso Machalińska, Anna Grulkowski, Ireneusz Sci Rep Article Quantitative evaluation of the human corneal grafts stored in the tissue banks is usually limited to endothelial cell density and central thickness. Swept-source OCT (SS-OCT) is capable of measuring the central curvatures of the corneal tissue prepared for transplantation without loss of sterileness, providing insights on its refractive state. The aim of the paper is to compare in vitro SS-OCT measurements with pre-excision values. Hand-held keratometry and ultrasound pachymetry was performed on 22 corneas before excision of corneoscleral button and insertion in the vial with Eusol-C solution (AlchimiaS.r.l, Nicolò, Italy). After 12 to 36 h of hypothermic storage the corneas were examined within the vials with custom built SS-OCT system maintaining a sterile environment. The anterior and posterior central curvatures, and central corneal thickness (CCT) were measured. Rotation of the corneoscleral button was controlled by making a 6-o'clock mark during excision. Mean pre-excision CCT was 626.45 ± 28.71 µm and 468.05 ± 52.96 µm when measured with SS OCT (r = 0.55; p < 0.001). Respective values for average keratometry were 7.74 ± 0.39 mm and 7.92 ± 0.57 mm (r = 0.6; p = 0.22). Although high differences were observed in corneal thickness, keratometric radius of curvature at the flat (r = 0.42; p < 0.001) and steep (r = 0.62; p = 0.014) meridian of the anterior corneal surface, as well as corneal anterior astigmatism (r = 0.3; p < 0.001), showed good correlation with pre-excision values. SS-OCT is capable of providing quantitative evaluation of the human corneal grafts in hypothermic storage. Good correlation between curvature measurements before excision and during banking in the vial indicates its clinical utility. Nature Publishing Group UK 2022-09-01 /pmc/articles/PMC9437096/ /pubmed/36050445 http://dx.doi.org/10.1038/s41598-022-19225-1 Text en © The Author(s) 2022 https://creativecommons.org/licenses/by/4.0/Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) .
spellingShingle Article
Kaluzny, Bartlomiej J.
Mlyniuk, Patryk
Liberski, Slawomir
Jimenez-Villar, Alfonso
Machalińska, Anna
Grulkowski, Ireneusz
Swept-source OCT for corneal graft quantitative evaluation in the eye bank and the correlation of the measurements to pre-excision values
title Swept-source OCT for corneal graft quantitative evaluation in the eye bank and the correlation of the measurements to pre-excision values
title_full Swept-source OCT for corneal graft quantitative evaluation in the eye bank and the correlation of the measurements to pre-excision values
title_fullStr Swept-source OCT for corneal graft quantitative evaluation in the eye bank and the correlation of the measurements to pre-excision values
title_full_unstemmed Swept-source OCT for corneal graft quantitative evaluation in the eye bank and the correlation of the measurements to pre-excision values
title_short Swept-source OCT for corneal graft quantitative evaluation in the eye bank and the correlation of the measurements to pre-excision values
title_sort swept-source oct for corneal graft quantitative evaluation in the eye bank and the correlation of the measurements to pre-excision values
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9437096/
https://www.ncbi.nlm.nih.gov/pubmed/36050445
http://dx.doi.org/10.1038/s41598-022-19225-1
work_keys_str_mv AT kaluznybartlomiejj sweptsourceoctforcornealgraftquantitativeevaluationintheeyebankandthecorrelationofthemeasurementstopreexcisionvalues
AT mlyniukpatryk sweptsourceoctforcornealgraftquantitativeevaluationintheeyebankandthecorrelationofthemeasurementstopreexcisionvalues
AT liberskislawomir sweptsourceoctforcornealgraftquantitativeevaluationintheeyebankandthecorrelationofthemeasurementstopreexcisionvalues
AT jimenezvillaralfonso sweptsourceoctforcornealgraftquantitativeevaluationintheeyebankandthecorrelationofthemeasurementstopreexcisionvalues
AT machalinskaanna sweptsourceoctforcornealgraftquantitativeevaluationintheeyebankandthecorrelationofthemeasurementstopreexcisionvalues
AT grulkowskiireneusz sweptsourceoctforcornealgraftquantitativeevaluationintheeyebankandthecorrelationofthemeasurementstopreexcisionvalues