Biomechanical behaviour of tension-band-reconstruction titanium plate in open-door laminoplasty: a study based on finite element analysis

OBJECTIVE: To investigate and evaluate the biomechanical behaviour of tension-band-reconstruction (TBR) and ordinary titanium plates in open-door laminoplasty by finite element (FE) analysis. METHODS: TBR titanium plate and ordinary titanium plate were implanted into a validated finite element model...

Descripción completa

Detalles Bibliográficos
Autores principales: Xu, Hanpeng, Wu, Jincheng, Xie, Hongru, Wen, Wangqiang, Xu, Haoxiang, Du, Juan, Miao, Jun
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9454233/
https://www.ncbi.nlm.nih.gov/pubmed/36076212
http://dx.doi.org/10.1186/s12891-022-05804-w
Descripción
Sumario:OBJECTIVE: To investigate and evaluate the biomechanical behaviour of tension-band-reconstruction (TBR) and ordinary titanium plates in open-door laminoplasty by finite element (FE) analysis. METHODS: TBR titanium plate and ordinary titanium plate were implanted into a validated finite element model of healthy adult cervical vertebrae. Among them, 5 ordinary titanium plate were used in model A, 2 TBR titanium plates and 3 ordinary titanium plates were used in model B, and 5 TBR titanium plates were used in model C. The same loading conditions was applied identically to all models. Range of motion (ROM) of the vertebral body, stress distribution of the titanium plate and intradiscal pressure (IDP) were compared in flexion, extension, lateral bending and rotation. RESULTS: The ROM of model B and C was similar in flexion and extension, and both were smaller than that of model A. The highest von Mises stress in the titanium plate appears is in model C. The IDP in C2/3 was significantly higher than that in other segments in flexion. There was no significant difference in IDP among three models in left lateral bending and left axial rotation. CONCLUSION: Application of TBR titanium plate in open-door laminoplasty can reduced ROM in flexion, extension and axial rotation of the cervical vertebrae. But the increase of stress in TBR titanium plate could lead to higher risk of adverse events such as titanium plate deformation. Moreover, compared with complete TBR titanium plate, the combination of TBR titanium plate for C3 and C7 with ordinary titanium plate for the other vertebrae largely reduce the stress of the titanium plates by ensuring stability. The proposed FE model (C2-T1) exhibits a great potential in evaluating biomechanical behaviour of TBR titanium plate for open-door laminoplasty. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1186/s12891-022-05804-w.