A multicenter performance evaluation of the new Elecsys Vitamin D total III assay versus reference isotope dilution liquid chromatography tandem mass spectrometry and commercially available comparators
BACKGROUND: Vitamin D deficiency/insufficiency and toxicity are worldwide issues; thus, accurate diagnostic assays are required to measure vitamin D. We evaluated the performance of the new Elecsys(®) Vitamin D total III assay (Roche Diagnostics International Ltd). METHODS: Repeatability and interme...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9459274/ https://www.ncbi.nlm.nih.gov/pubmed/35852988 http://dx.doi.org/10.1002/jcla.24610 |
_version_ | 1784786473327788032 |
---|---|
author | Findeisen, Peter Leis, Marta Bendig, Garnet Grimme, Jill Moser, Elizabeth Vogl, Christian Christenson, Robert |
author_facet | Findeisen, Peter Leis, Marta Bendig, Garnet Grimme, Jill Moser, Elizabeth Vogl, Christian Christenson, Robert |
author_sort | Findeisen, Peter |
collection | PubMed |
description | BACKGROUND: Vitamin D deficiency/insufficiency and toxicity are worldwide issues; thus, accurate diagnostic assays are required to measure vitamin D. We evaluated the performance of the new Elecsys(®) Vitamin D total III assay (Roche Diagnostics International Ltd). METHODS: Repeatability and intermediate precision of the Elecsys Vitamin D total III assay (cobas e 601 analyzer) were evaluated at three sites using five human serum pools (HSPs) and two PreciControls (five‐day model, one reagent lot [CLSI‐EP05‐A3]) and compared against prespecified acceptance criteria. A serum verification panel, with reference isotope dilution liquid chromatography tandem mass spectrometry (ID‐LC–MS/MS) values, was used for comparator assay/concordance studies at two sites, assessed using unweighted Deming regression. Testing of serum vs. plasma on the Elecsys assay was conducted at one site using samples from healthy adults; assessed using Passing‐Bablok regression. RESULTS: Repeatability (HSP1 [16.8–18.4 ng/ml], SD 0.87–1.07; HSP5 [94.5–98.0 ng/ml], CV 1.58%–2.76%) and intermediate precision (HSP1, SD 1.14–1.77; HSP5, CV 2.00%–4.13%) met acceptance criteria across sites. Agreement was observed between the Elecsys assay and (i) the ID‐LC–MS/MS verification panel (slope, 0.936–1.01; Pearson's r, 0.960–0.986) and (ii) comparator assays (slope, 0.921–1.15; Pearson's r, 0.958–0.982). The Elecsys assay correctly assigned the highest combined percentage of samples to deficient (100%) and insufficient (89.5%) vitamin D categories vs. comparator assays and demonstrated comparable performance in serum and plasma (y = 0.103 + 0.984x). CONCLUSIONS: The Elecsys Vitamin D total III assay demonstrated good analytical performance and compared favorably with other assays, supporting its use in clinical practice. |
format | Online Article Text |
id | pubmed-9459274 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | John Wiley and Sons Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-94592742022-09-12 A multicenter performance evaluation of the new Elecsys Vitamin D total III assay versus reference isotope dilution liquid chromatography tandem mass spectrometry and commercially available comparators Findeisen, Peter Leis, Marta Bendig, Garnet Grimme, Jill Moser, Elizabeth Vogl, Christian Christenson, Robert J Clin Lab Anal Research Articles BACKGROUND: Vitamin D deficiency/insufficiency and toxicity are worldwide issues; thus, accurate diagnostic assays are required to measure vitamin D. We evaluated the performance of the new Elecsys(®) Vitamin D total III assay (Roche Diagnostics International Ltd). METHODS: Repeatability and intermediate precision of the Elecsys Vitamin D total III assay (cobas e 601 analyzer) were evaluated at three sites using five human serum pools (HSPs) and two PreciControls (five‐day model, one reagent lot [CLSI‐EP05‐A3]) and compared against prespecified acceptance criteria. A serum verification panel, with reference isotope dilution liquid chromatography tandem mass spectrometry (ID‐LC–MS/MS) values, was used for comparator assay/concordance studies at two sites, assessed using unweighted Deming regression. Testing of serum vs. plasma on the Elecsys assay was conducted at one site using samples from healthy adults; assessed using Passing‐Bablok regression. RESULTS: Repeatability (HSP1 [16.8–18.4 ng/ml], SD 0.87–1.07; HSP5 [94.5–98.0 ng/ml], CV 1.58%–2.76%) and intermediate precision (HSP1, SD 1.14–1.77; HSP5, CV 2.00%–4.13%) met acceptance criteria across sites. Agreement was observed between the Elecsys assay and (i) the ID‐LC–MS/MS verification panel (slope, 0.936–1.01; Pearson's r, 0.960–0.986) and (ii) comparator assays (slope, 0.921–1.15; Pearson's r, 0.958–0.982). The Elecsys assay correctly assigned the highest combined percentage of samples to deficient (100%) and insufficient (89.5%) vitamin D categories vs. comparator assays and demonstrated comparable performance in serum and plasma (y = 0.103 + 0.984x). CONCLUSIONS: The Elecsys Vitamin D total III assay demonstrated good analytical performance and compared favorably with other assays, supporting its use in clinical practice. John Wiley and Sons Inc. 2022-07-19 /pmc/articles/PMC9459274/ /pubmed/35852988 http://dx.doi.org/10.1002/jcla.24610 Text en © 2022 The Authors. Journal of Clinical Laboratory Analysis published by Wiley Periodicals LLC. https://creativecommons.org/licenses/by/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Research Articles Findeisen, Peter Leis, Marta Bendig, Garnet Grimme, Jill Moser, Elizabeth Vogl, Christian Christenson, Robert A multicenter performance evaluation of the new Elecsys Vitamin D total III assay versus reference isotope dilution liquid chromatography tandem mass spectrometry and commercially available comparators |
title | A multicenter performance evaluation of the new Elecsys Vitamin D total III assay versus reference isotope dilution liquid chromatography tandem mass spectrometry and commercially available comparators |
title_full | A multicenter performance evaluation of the new Elecsys Vitamin D total III assay versus reference isotope dilution liquid chromatography tandem mass spectrometry and commercially available comparators |
title_fullStr | A multicenter performance evaluation of the new Elecsys Vitamin D total III assay versus reference isotope dilution liquid chromatography tandem mass spectrometry and commercially available comparators |
title_full_unstemmed | A multicenter performance evaluation of the new Elecsys Vitamin D total III assay versus reference isotope dilution liquid chromatography tandem mass spectrometry and commercially available comparators |
title_short | A multicenter performance evaluation of the new Elecsys Vitamin D total III assay versus reference isotope dilution liquid chromatography tandem mass spectrometry and commercially available comparators |
title_sort | multicenter performance evaluation of the new elecsys vitamin d total iii assay versus reference isotope dilution liquid chromatography tandem mass spectrometry and commercially available comparators |
topic | Research Articles |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9459274/ https://www.ncbi.nlm.nih.gov/pubmed/35852988 http://dx.doi.org/10.1002/jcla.24610 |
work_keys_str_mv | AT findeisenpeter amulticenterperformanceevaluationofthenewelecsysvitamindtotaliiiassayversusreferenceisotopedilutionliquidchromatographytandemmassspectrometryandcommerciallyavailablecomparators AT leismarta amulticenterperformanceevaluationofthenewelecsysvitamindtotaliiiassayversusreferenceisotopedilutionliquidchromatographytandemmassspectrometryandcommerciallyavailablecomparators AT bendiggarnet amulticenterperformanceevaluationofthenewelecsysvitamindtotaliiiassayversusreferenceisotopedilutionliquidchromatographytandemmassspectrometryandcommerciallyavailablecomparators AT grimmejill amulticenterperformanceevaluationofthenewelecsysvitamindtotaliiiassayversusreferenceisotopedilutionliquidchromatographytandemmassspectrometryandcommerciallyavailablecomparators AT moserelizabeth amulticenterperformanceevaluationofthenewelecsysvitamindtotaliiiassayversusreferenceisotopedilutionliquidchromatographytandemmassspectrometryandcommerciallyavailablecomparators AT voglchristian amulticenterperformanceevaluationofthenewelecsysvitamindtotaliiiassayversusreferenceisotopedilutionliquidchromatographytandemmassspectrometryandcommerciallyavailablecomparators AT christensonrobert amulticenterperformanceevaluationofthenewelecsysvitamindtotaliiiassayversusreferenceisotopedilutionliquidchromatographytandemmassspectrometryandcommerciallyavailablecomparators AT findeisenpeter multicenterperformanceevaluationofthenewelecsysvitamindtotaliiiassayversusreferenceisotopedilutionliquidchromatographytandemmassspectrometryandcommerciallyavailablecomparators AT leismarta multicenterperformanceevaluationofthenewelecsysvitamindtotaliiiassayversusreferenceisotopedilutionliquidchromatographytandemmassspectrometryandcommerciallyavailablecomparators AT bendiggarnet multicenterperformanceevaluationofthenewelecsysvitamindtotaliiiassayversusreferenceisotopedilutionliquidchromatographytandemmassspectrometryandcommerciallyavailablecomparators AT grimmejill multicenterperformanceevaluationofthenewelecsysvitamindtotaliiiassayversusreferenceisotopedilutionliquidchromatographytandemmassspectrometryandcommerciallyavailablecomparators AT moserelizabeth multicenterperformanceevaluationofthenewelecsysvitamindtotaliiiassayversusreferenceisotopedilutionliquidchromatographytandemmassspectrometryandcommerciallyavailablecomparators AT voglchristian multicenterperformanceevaluationofthenewelecsysvitamindtotaliiiassayversusreferenceisotopedilutionliquidchromatographytandemmassspectrometryandcommerciallyavailablecomparators AT christensonrobert multicenterperformanceevaluationofthenewelecsysvitamindtotaliiiassayversusreferenceisotopedilutionliquidchromatographytandemmassspectrometryandcommerciallyavailablecomparators |