Performance evaluation of presepsin using a Sysmex HISCL‐5000 analyzer and determination of reference interval

BACKGROUND: Analytical evaluation of newly developed presepsin by a Sysmex HISCL‐5000 (Sysmex, Japan) automated immune analyzer was performed. METHODS: For evaluation, sepsis patient samples were collected before treatment in an emergency department. Precision, linearity, limit of blank/limit of det...

Descripción completa

Detalles Bibliográficos
Autores principales: Kang, Taewon, Yoo, Jeaeun, Choi, Hyunyu, Lee, Seungok, Jekarl, Dong Wook, Kim, Yonggoo
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9459287/
https://www.ncbi.nlm.nih.gov/pubmed/35870180
http://dx.doi.org/10.1002/jcla.24618
_version_ 1784786476526993408
author Kang, Taewon
Yoo, Jeaeun
Choi, Hyunyu
Lee, Seungok
Jekarl, Dong Wook
Kim, Yonggoo
author_facet Kang, Taewon
Yoo, Jeaeun
Choi, Hyunyu
Lee, Seungok
Jekarl, Dong Wook
Kim, Yonggoo
author_sort Kang, Taewon
collection PubMed
description BACKGROUND: Analytical evaluation of newly developed presepsin by a Sysmex HISCL‐5000 (Sysmex, Japan) automated immune analyzer was performed. METHODS: For evaluation, sepsis patient samples were collected before treatment in an emergency department. Precision, linearity, limit of blank/limit of detection, method comparisons, and reference intervals were evaluated. Method comparisons were performed using a PATHFAST immune analyzer (LSI Medience Corporation, Japan). RESULTS: Precision using a 20x2x2 protocol for low (306 pg/mL) and high (1031 pg/mL) levels resulted in within‐laboratory standard deviation (95% confidence interval [CI]) and coefficient of variation (CV) %, which were as follows: 15.3 (13.1–18.7), 5.5% and 47.7, (40.5–58.1), 6.4%, respectively. Linearity using patient samples and calibrators were measured from 201 to 16,177 and 188 to 30,000 pg/mL, respectively. The regression equation was y = −23.2 + 1.008x (SE = 162.4) for low levels and y = 779.9 + 1.006x (SE = 668) for high levels. Method comparison by Passing–Bablock analysis was as follows: y = −209.77 + 1.047x (S(yx) = 335.3). The correlation coefficient (95% CI) was 0.869 (0.772–0.927) with statistical significance (p < 0.001). Reference intervals from 120 normal healthy subjects showed that 300 pg/mL was the cut off. Presepsin tended to show a higher value at higher ages and in males. Presepsin showed correlation with some parameters, and the correlation coefficient (p value) were as follows: hematocrit, 0.198 (0.03); eGFR (CKD‐EPI), −0.240 (0.0129); MDRD‐eGFR, −0.194 (0.048), respectively. CONCLUSION: Presepsin measurement by HISCL‐5000 showed reliable performance. Further clinical studies are required for the diagnosis and prognosis of sepsis.
format Online
Article
Text
id pubmed-9459287
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-94592872022-09-12 Performance evaluation of presepsin using a Sysmex HISCL‐5000 analyzer and determination of reference interval Kang, Taewon Yoo, Jeaeun Choi, Hyunyu Lee, Seungok Jekarl, Dong Wook Kim, Yonggoo J Clin Lab Anal Research Articles BACKGROUND: Analytical evaluation of newly developed presepsin by a Sysmex HISCL‐5000 (Sysmex, Japan) automated immune analyzer was performed. METHODS: For evaluation, sepsis patient samples were collected before treatment in an emergency department. Precision, linearity, limit of blank/limit of detection, method comparisons, and reference intervals were evaluated. Method comparisons were performed using a PATHFAST immune analyzer (LSI Medience Corporation, Japan). RESULTS: Precision using a 20x2x2 protocol for low (306 pg/mL) and high (1031 pg/mL) levels resulted in within‐laboratory standard deviation (95% confidence interval [CI]) and coefficient of variation (CV) %, which were as follows: 15.3 (13.1–18.7), 5.5% and 47.7, (40.5–58.1), 6.4%, respectively. Linearity using patient samples and calibrators were measured from 201 to 16,177 and 188 to 30,000 pg/mL, respectively. The regression equation was y = −23.2 + 1.008x (SE = 162.4) for low levels and y = 779.9 + 1.006x (SE = 668) for high levels. Method comparison by Passing–Bablock analysis was as follows: y = −209.77 + 1.047x (S(yx) = 335.3). The correlation coefficient (95% CI) was 0.869 (0.772–0.927) with statistical significance (p < 0.001). Reference intervals from 120 normal healthy subjects showed that 300 pg/mL was the cut off. Presepsin tended to show a higher value at higher ages and in males. Presepsin showed correlation with some parameters, and the correlation coefficient (p value) were as follows: hematocrit, 0.198 (0.03); eGFR (CKD‐EPI), −0.240 (0.0129); MDRD‐eGFR, −0.194 (0.048), respectively. CONCLUSION: Presepsin measurement by HISCL‐5000 showed reliable performance. Further clinical studies are required for the diagnosis and prognosis of sepsis. John Wiley and Sons Inc. 2022-07-23 /pmc/articles/PMC9459287/ /pubmed/35870180 http://dx.doi.org/10.1002/jcla.24618 Text en © 2022 The Authors. Journal of Clinical Laboratory Analysis published by Wiley Periodicals LLC. https://creativecommons.org/licenses/by/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research Articles
Kang, Taewon
Yoo, Jeaeun
Choi, Hyunyu
Lee, Seungok
Jekarl, Dong Wook
Kim, Yonggoo
Performance evaluation of presepsin using a Sysmex HISCL‐5000 analyzer and determination of reference interval
title Performance evaluation of presepsin using a Sysmex HISCL‐5000 analyzer and determination of reference interval
title_full Performance evaluation of presepsin using a Sysmex HISCL‐5000 analyzer and determination of reference interval
title_fullStr Performance evaluation of presepsin using a Sysmex HISCL‐5000 analyzer and determination of reference interval
title_full_unstemmed Performance evaluation of presepsin using a Sysmex HISCL‐5000 analyzer and determination of reference interval
title_short Performance evaluation of presepsin using a Sysmex HISCL‐5000 analyzer and determination of reference interval
title_sort performance evaluation of presepsin using a sysmex hiscl‐5000 analyzer and determination of reference interval
topic Research Articles
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9459287/
https://www.ncbi.nlm.nih.gov/pubmed/35870180
http://dx.doi.org/10.1002/jcla.24618
work_keys_str_mv AT kangtaewon performanceevaluationofpresepsinusingasysmexhiscl5000analyzeranddeterminationofreferenceinterval
AT yoojeaeun performanceevaluationofpresepsinusingasysmexhiscl5000analyzeranddeterminationofreferenceinterval
AT choihyunyu performanceevaluationofpresepsinusingasysmexhiscl5000analyzeranddeterminationofreferenceinterval
AT leeseungok performanceevaluationofpresepsinusingasysmexhiscl5000analyzeranddeterminationofreferenceinterval
AT jekarldongwook performanceevaluationofpresepsinusingasysmexhiscl5000analyzeranddeterminationofreferenceinterval
AT kimyonggoo performanceevaluationofpresepsinusingasysmexhiscl5000analyzeranddeterminationofreferenceinterval