CyberKnife Xsight versus fiducial-based target-tracking: a novel 3D dosimetric comparison in a dynamic phantom
BACKGROUND: The CyberKnife Xsight lung-tracking system (XLTS) provides an alternative to fiducial-based target-tracking systems (FTTS) for non-small-cell lung cancer (NSCLC) patients without invasive fiducial insertion procedures. This study provides a method for 3D independent dosimetric verificati...
Autores principales: | , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9461108/ https://www.ncbi.nlm.nih.gov/pubmed/36076249 http://dx.doi.org/10.1186/s13014-022-02123-1 |
_version_ | 1784786905894748160 |
---|---|
author | Klein, Thomas J. Gill, Suki Ebert, Martin A. Grogan, Garry Smith, Warwick Alkhatib, Zaid Geraghty, John Scott, Alison J. D. Brown, Alan Rowshanfarzad, Pejman |
author_facet | Klein, Thomas J. Gill, Suki Ebert, Martin A. Grogan, Garry Smith, Warwick Alkhatib, Zaid Geraghty, John Scott, Alison J. D. Brown, Alan Rowshanfarzad, Pejman |
author_sort | Klein, Thomas J. |
collection | PubMed |
description | BACKGROUND: The CyberKnife Xsight lung-tracking system (XLTS) provides an alternative to fiducial-based target-tracking systems (FTTS) for non-small-cell lung cancer (NSCLC) patients without invasive fiducial insertion procedures. This study provides a method for 3D independent dosimetric verification of the accuracy of the FTTS compared to the XLTS without relying on log-files generated by the CyberKnife system. METHODS: A respiratory motion trace was taken from a 4D-CT of a real lung cancer patient and applied to a modified QUASAR™ respiratory motion phantom. A novel approach to 3D dosimetry was developed using Gafchromic EBT3 film, allowing the 3D dose distribution delivered to the moving phantom to be reconstructed. Treatments were planned using the recommended margins for one and three fiducial markers and XLTS 2-view, 1-view and 0-view target-tracking modalities. The dose delivery accuracy was analysed by comparing the reconstructed dose distributions to the planned dose distributions using gamma index analysis. RESULTS: For the 3%/2 mm gamma criterion, gamma passing rates up to 99.37% were observed for the static deliveries. The 3-fiducial and 1-fiducial-based deliveries exhibited passing rates of 93.74% and 97.82%, respectively, in the absence of target rotation. When target rotation was considered, the passing rate for 1-fiducial tracking degraded to 91.24%. The passing rates observed for XLTS 2-view, 1-view and 0-view target-tracking were 92.78%, 96.22% and 76.08%, respectively. CONCLUSIONS: Except for the XLTS 0-view, the dosimetric accuracy of the XLTS was comparable to the FTTS under equivalent treatment conditions. This study gives us further confidence in the CyberKnife XLTS and FTTS systems. |
format | Online Article Text |
id | pubmed-9461108 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-94611082022-09-10 CyberKnife Xsight versus fiducial-based target-tracking: a novel 3D dosimetric comparison in a dynamic phantom Klein, Thomas J. Gill, Suki Ebert, Martin A. Grogan, Garry Smith, Warwick Alkhatib, Zaid Geraghty, John Scott, Alison J. D. Brown, Alan Rowshanfarzad, Pejman Radiat Oncol Research BACKGROUND: The CyberKnife Xsight lung-tracking system (XLTS) provides an alternative to fiducial-based target-tracking systems (FTTS) for non-small-cell lung cancer (NSCLC) patients without invasive fiducial insertion procedures. This study provides a method for 3D independent dosimetric verification of the accuracy of the FTTS compared to the XLTS without relying on log-files generated by the CyberKnife system. METHODS: A respiratory motion trace was taken from a 4D-CT of a real lung cancer patient and applied to a modified QUASAR™ respiratory motion phantom. A novel approach to 3D dosimetry was developed using Gafchromic EBT3 film, allowing the 3D dose distribution delivered to the moving phantom to be reconstructed. Treatments were planned using the recommended margins for one and three fiducial markers and XLTS 2-view, 1-view and 0-view target-tracking modalities. The dose delivery accuracy was analysed by comparing the reconstructed dose distributions to the planned dose distributions using gamma index analysis. RESULTS: For the 3%/2 mm gamma criterion, gamma passing rates up to 99.37% were observed for the static deliveries. The 3-fiducial and 1-fiducial-based deliveries exhibited passing rates of 93.74% and 97.82%, respectively, in the absence of target rotation. When target rotation was considered, the passing rate for 1-fiducial tracking degraded to 91.24%. The passing rates observed for XLTS 2-view, 1-view and 0-view target-tracking were 92.78%, 96.22% and 76.08%, respectively. CONCLUSIONS: Except for the XLTS 0-view, the dosimetric accuracy of the XLTS was comparable to the FTTS under equivalent treatment conditions. This study gives us further confidence in the CyberKnife XLTS and FTTS systems. BioMed Central 2022-09-08 /pmc/articles/PMC9461108/ /pubmed/36076249 http://dx.doi.org/10.1186/s13014-022-02123-1 Text en © The Author(s) 2022 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) . The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/ (https://creativecommons.org/publicdomain/zero/1.0/) ) applies to the data made available in this article, unless otherwise stated in a credit line to the data. |
spellingShingle | Research Klein, Thomas J. Gill, Suki Ebert, Martin A. Grogan, Garry Smith, Warwick Alkhatib, Zaid Geraghty, John Scott, Alison J. D. Brown, Alan Rowshanfarzad, Pejman CyberKnife Xsight versus fiducial-based target-tracking: a novel 3D dosimetric comparison in a dynamic phantom |
title | CyberKnife Xsight versus fiducial-based target-tracking: a novel 3D dosimetric comparison in a dynamic phantom |
title_full | CyberKnife Xsight versus fiducial-based target-tracking: a novel 3D dosimetric comparison in a dynamic phantom |
title_fullStr | CyberKnife Xsight versus fiducial-based target-tracking: a novel 3D dosimetric comparison in a dynamic phantom |
title_full_unstemmed | CyberKnife Xsight versus fiducial-based target-tracking: a novel 3D dosimetric comparison in a dynamic phantom |
title_short | CyberKnife Xsight versus fiducial-based target-tracking: a novel 3D dosimetric comparison in a dynamic phantom |
title_sort | cyberknife xsight versus fiducial-based target-tracking: a novel 3d dosimetric comparison in a dynamic phantom |
topic | Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9461108/ https://www.ncbi.nlm.nih.gov/pubmed/36076249 http://dx.doi.org/10.1186/s13014-022-02123-1 |
work_keys_str_mv | AT kleinthomasj cyberknifexsightversusfiducialbasedtargettrackinganovel3ddosimetriccomparisoninadynamicphantom AT gillsuki cyberknifexsightversusfiducialbasedtargettrackinganovel3ddosimetriccomparisoninadynamicphantom AT ebertmartina cyberknifexsightversusfiducialbasedtargettrackinganovel3ddosimetriccomparisoninadynamicphantom AT grogangarry cyberknifexsightversusfiducialbasedtargettrackinganovel3ddosimetriccomparisoninadynamicphantom AT smithwarwick cyberknifexsightversusfiducialbasedtargettrackinganovel3ddosimetriccomparisoninadynamicphantom AT alkhatibzaid cyberknifexsightversusfiducialbasedtargettrackinganovel3ddosimetriccomparisoninadynamicphantom AT geraghtyjohn cyberknifexsightversusfiducialbasedtargettrackinganovel3ddosimetriccomparisoninadynamicphantom AT scottalisonjd cyberknifexsightversusfiducialbasedtargettrackinganovel3ddosimetriccomparisoninadynamicphantom AT brownalan cyberknifexsightversusfiducialbasedtargettrackinganovel3ddosimetriccomparisoninadynamicphantom AT rowshanfarzadpejman cyberknifexsightversusfiducialbasedtargettrackinganovel3ddosimetriccomparisoninadynamicphantom |