CyberKnife Xsight versus fiducial-based target-tracking: a novel 3D dosimetric comparison in a dynamic phantom

BACKGROUND: The CyberKnife Xsight lung-tracking system (XLTS) provides an alternative to fiducial-based target-tracking systems (FTTS) for non-small-cell lung cancer (NSCLC) patients without invasive fiducial insertion procedures. This study provides a method for 3D independent dosimetric verificati...

Descripción completa

Detalles Bibliográficos
Autores principales: Klein, Thomas J., Gill, Suki, Ebert, Martin A., Grogan, Garry, Smith, Warwick, Alkhatib, Zaid, Geraghty, John, Scott, Alison J. D., Brown, Alan, Rowshanfarzad, Pejman
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9461108/
https://www.ncbi.nlm.nih.gov/pubmed/36076249
http://dx.doi.org/10.1186/s13014-022-02123-1
_version_ 1784786905894748160
author Klein, Thomas J.
Gill, Suki
Ebert, Martin A.
Grogan, Garry
Smith, Warwick
Alkhatib, Zaid
Geraghty, John
Scott, Alison J. D.
Brown, Alan
Rowshanfarzad, Pejman
author_facet Klein, Thomas J.
Gill, Suki
Ebert, Martin A.
Grogan, Garry
Smith, Warwick
Alkhatib, Zaid
Geraghty, John
Scott, Alison J. D.
Brown, Alan
Rowshanfarzad, Pejman
author_sort Klein, Thomas J.
collection PubMed
description BACKGROUND: The CyberKnife Xsight lung-tracking system (XLTS) provides an alternative to fiducial-based target-tracking systems (FTTS) for non-small-cell lung cancer (NSCLC) patients without invasive fiducial insertion procedures. This study provides a method for 3D independent dosimetric verification of the accuracy of the FTTS compared to the XLTS without relying on log-files generated by the CyberKnife system. METHODS: A respiratory motion trace was taken from a 4D-CT of a real lung cancer patient and applied to a modified QUASAR™ respiratory motion phantom. A novel approach to 3D dosimetry was developed using Gafchromic EBT3 film, allowing the 3D dose distribution delivered to the moving phantom to be reconstructed. Treatments were planned using the recommended margins for one and three fiducial markers and XLTS 2-view, 1-view and 0-view target-tracking modalities. The dose delivery accuracy was analysed by comparing the reconstructed dose distributions to the planned dose distributions using gamma index analysis. RESULTS: For the 3%/2 mm gamma criterion, gamma passing rates up to 99.37% were observed for the static deliveries. The 3-fiducial and 1-fiducial-based deliveries exhibited passing rates of 93.74% and 97.82%, respectively, in the absence of target rotation. When target rotation was considered, the passing rate for 1-fiducial tracking degraded to 91.24%. The passing rates observed for XLTS 2-view, 1-view and 0-view target-tracking were 92.78%, 96.22% and 76.08%, respectively. CONCLUSIONS: Except for the XLTS 0-view, the dosimetric accuracy of the XLTS was comparable to the FTTS under equivalent treatment conditions. This study gives us further confidence in the CyberKnife XLTS and FTTS systems.
format Online
Article
Text
id pubmed-9461108
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-94611082022-09-10 CyberKnife Xsight versus fiducial-based target-tracking: a novel 3D dosimetric comparison in a dynamic phantom Klein, Thomas J. Gill, Suki Ebert, Martin A. Grogan, Garry Smith, Warwick Alkhatib, Zaid Geraghty, John Scott, Alison J. D. Brown, Alan Rowshanfarzad, Pejman Radiat Oncol Research BACKGROUND: The CyberKnife Xsight lung-tracking system (XLTS) provides an alternative to fiducial-based target-tracking systems (FTTS) for non-small-cell lung cancer (NSCLC) patients without invasive fiducial insertion procedures. This study provides a method for 3D independent dosimetric verification of the accuracy of the FTTS compared to the XLTS without relying on log-files generated by the CyberKnife system. METHODS: A respiratory motion trace was taken from a 4D-CT of a real lung cancer patient and applied to a modified QUASAR™ respiratory motion phantom. A novel approach to 3D dosimetry was developed using Gafchromic EBT3 film, allowing the 3D dose distribution delivered to the moving phantom to be reconstructed. Treatments were planned using the recommended margins for one and three fiducial markers and XLTS 2-view, 1-view and 0-view target-tracking modalities. The dose delivery accuracy was analysed by comparing the reconstructed dose distributions to the planned dose distributions using gamma index analysis. RESULTS: For the 3%/2 mm gamma criterion, gamma passing rates up to 99.37% were observed for the static deliveries. The 3-fiducial and 1-fiducial-based deliveries exhibited passing rates of 93.74% and 97.82%, respectively, in the absence of target rotation. When target rotation was considered, the passing rate for 1-fiducial tracking degraded to 91.24%. The passing rates observed for XLTS 2-view, 1-view and 0-view target-tracking were 92.78%, 96.22% and 76.08%, respectively. CONCLUSIONS: Except for the XLTS 0-view, the dosimetric accuracy of the XLTS was comparable to the FTTS under equivalent treatment conditions. This study gives us further confidence in the CyberKnife XLTS and FTTS systems. BioMed Central 2022-09-08 /pmc/articles/PMC9461108/ /pubmed/36076249 http://dx.doi.org/10.1186/s13014-022-02123-1 Text en © The Author(s) 2022 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) . The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/ (https://creativecommons.org/publicdomain/zero/1.0/) ) applies to the data made available in this article, unless otherwise stated in a credit line to the data.
spellingShingle Research
Klein, Thomas J.
Gill, Suki
Ebert, Martin A.
Grogan, Garry
Smith, Warwick
Alkhatib, Zaid
Geraghty, John
Scott, Alison J. D.
Brown, Alan
Rowshanfarzad, Pejman
CyberKnife Xsight versus fiducial-based target-tracking: a novel 3D dosimetric comparison in a dynamic phantom
title CyberKnife Xsight versus fiducial-based target-tracking: a novel 3D dosimetric comparison in a dynamic phantom
title_full CyberKnife Xsight versus fiducial-based target-tracking: a novel 3D dosimetric comparison in a dynamic phantom
title_fullStr CyberKnife Xsight versus fiducial-based target-tracking: a novel 3D dosimetric comparison in a dynamic phantom
title_full_unstemmed CyberKnife Xsight versus fiducial-based target-tracking: a novel 3D dosimetric comparison in a dynamic phantom
title_short CyberKnife Xsight versus fiducial-based target-tracking: a novel 3D dosimetric comparison in a dynamic phantom
title_sort cyberknife xsight versus fiducial-based target-tracking: a novel 3d dosimetric comparison in a dynamic phantom
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9461108/
https://www.ncbi.nlm.nih.gov/pubmed/36076249
http://dx.doi.org/10.1186/s13014-022-02123-1
work_keys_str_mv AT kleinthomasj cyberknifexsightversusfiducialbasedtargettrackinganovel3ddosimetriccomparisoninadynamicphantom
AT gillsuki cyberknifexsightversusfiducialbasedtargettrackinganovel3ddosimetriccomparisoninadynamicphantom
AT ebertmartina cyberknifexsightversusfiducialbasedtargettrackinganovel3ddosimetriccomparisoninadynamicphantom
AT grogangarry cyberknifexsightversusfiducialbasedtargettrackinganovel3ddosimetriccomparisoninadynamicphantom
AT smithwarwick cyberknifexsightversusfiducialbasedtargettrackinganovel3ddosimetriccomparisoninadynamicphantom
AT alkhatibzaid cyberknifexsightversusfiducialbasedtargettrackinganovel3ddosimetriccomparisoninadynamicphantom
AT geraghtyjohn cyberknifexsightversusfiducialbasedtargettrackinganovel3ddosimetriccomparisoninadynamicphantom
AT scottalisonjd cyberknifexsightversusfiducialbasedtargettrackinganovel3ddosimetriccomparisoninadynamicphantom
AT brownalan cyberknifexsightversusfiducialbasedtargettrackinganovel3ddosimetriccomparisoninadynamicphantom
AT rowshanfarzadpejman cyberknifexsightversusfiducialbasedtargettrackinganovel3ddosimetriccomparisoninadynamicphantom