Clinical Usefulness of Ultraperformance Liquid Chromatography-Tandem Mass Spectrometry Method for Low Serum Testosterone Measurement

BACKGROUND: Mass spectrometry methods exhibit higher accuracy and lower variability than immunoassays at low testosterone concentrations. We developed and validated an ultraperformance liquid chromatography-tandem mass spectrometry (UPLC-MS/MS) assay for quantifying serum total testosterone. METHODS...

Descripción completa

Detalles Bibliográficos
Autores principales: Cho, Sung-Eun, Han, Jungsun, Park, Ju-Hee, Park, Euna, Kim, Geun Young, Lee, Jun Hyung, Yi, Ahram, Lee, Sang Gon, Lee, Eun Hee, Yun, Yeo-Min
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Korean Society for Laboratory Medicine 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9467846/
https://www.ncbi.nlm.nih.gov/pubmed/36045053
http://dx.doi.org/10.3343/alm.2023.43.1.19
_version_ 1784788281737609216
author Cho, Sung-Eun
Han, Jungsun
Park, Ju-Hee
Park, Euna
Kim, Geun Young
Lee, Jun Hyung
Yi, Ahram
Lee, Sang Gon
Lee, Eun Hee
Yun, Yeo-Min
author_facet Cho, Sung-Eun
Han, Jungsun
Park, Ju-Hee
Park, Euna
Kim, Geun Young
Lee, Jun Hyung
Yi, Ahram
Lee, Sang Gon
Lee, Eun Hee
Yun, Yeo-Min
author_sort Cho, Sung-Eun
collection PubMed
description BACKGROUND: Mass spectrometry methods exhibit higher accuracy and lower variability than immunoassays at low testosterone concentrations. We developed and validated an ultraperformance liquid chromatography-tandem mass spectrometry (UPLC-MS/MS) assay for quantifying serum total testosterone. METHODS: We used an ExionLC UPLC (Sciex, Framingham, MA, USA) system and a Sciex Triple Quad 6500+ (Sciex) MS/MS system in electrospray ionization and positive ion modes with multiple reaction monitoring transitions to evaluate precision, accuracy, linearity, lower limit of quantitation (LLOQ), carryover, ion suppression, stability, and reference intervals. For method comparison, we measured serum testosterone concentrations using this method in 40 subjects whose testosterone concentrations ranged from 0.14 to 55.48 nmol/L as determined using the Architect i2000 immunoassay (Abbott Diagnostics, Abbott Park, IL, USA) and in an additional 160 sera with testosterone concentrations <1.67 nmol/L. RESULTS: The intra- and inter-run precision CVs were <2.81%, and the accuracy bias values were <3.85%, which were all acceptable. The verified linear interval was 0.03–180.84 nmol/L; the LLOQ was 0.03 nmol/L. No significant carryover and ion suppression were observed. The testosterone in serum was stable at 4°C, at –20°C, and after three freeze-thaw cycles. The reference intervals were successfully verified. The correlation was good at testosterone concentrations of 0.14–55.48 nmol/L; however, the Architect assay showed positive percent bias at concentrations <1.67 nmol/L. CONCLUSIONS: The UPLC-MS/MS assay shows acceptable performance, with a lower LLOQ than the immunoassay. This method will enable the quantitation of low testosterone concentrations.
format Online
Article
Text
id pubmed-9467846
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher Korean Society for Laboratory Medicine
record_format MEDLINE/PubMed
spelling pubmed-94678462022-12-26 Clinical Usefulness of Ultraperformance Liquid Chromatography-Tandem Mass Spectrometry Method for Low Serum Testosterone Measurement Cho, Sung-Eun Han, Jungsun Park, Ju-Hee Park, Euna Kim, Geun Young Lee, Jun Hyung Yi, Ahram Lee, Sang Gon Lee, Eun Hee Yun, Yeo-Min Ann Lab Med Original Article BACKGROUND: Mass spectrometry methods exhibit higher accuracy and lower variability than immunoassays at low testosterone concentrations. We developed and validated an ultraperformance liquid chromatography-tandem mass spectrometry (UPLC-MS/MS) assay for quantifying serum total testosterone. METHODS: We used an ExionLC UPLC (Sciex, Framingham, MA, USA) system and a Sciex Triple Quad 6500+ (Sciex) MS/MS system in electrospray ionization and positive ion modes with multiple reaction monitoring transitions to evaluate precision, accuracy, linearity, lower limit of quantitation (LLOQ), carryover, ion suppression, stability, and reference intervals. For method comparison, we measured serum testosterone concentrations using this method in 40 subjects whose testosterone concentrations ranged from 0.14 to 55.48 nmol/L as determined using the Architect i2000 immunoassay (Abbott Diagnostics, Abbott Park, IL, USA) and in an additional 160 sera with testosterone concentrations <1.67 nmol/L. RESULTS: The intra- and inter-run precision CVs were <2.81%, and the accuracy bias values were <3.85%, which were all acceptable. The verified linear interval was 0.03–180.84 nmol/L; the LLOQ was 0.03 nmol/L. No significant carryover and ion suppression were observed. The testosterone in serum was stable at 4°C, at –20°C, and after three freeze-thaw cycles. The reference intervals were successfully verified. The correlation was good at testosterone concentrations of 0.14–55.48 nmol/L; however, the Architect assay showed positive percent bias at concentrations <1.67 nmol/L. CONCLUSIONS: The UPLC-MS/MS assay shows acceptable performance, with a lower LLOQ than the immunoassay. This method will enable the quantitation of low testosterone concentrations. Korean Society for Laboratory Medicine 2023-01-01 2022-09-01 /pmc/articles/PMC9467846/ /pubmed/36045053 http://dx.doi.org/10.3343/alm.2023.43.1.19 Text en © Korean Society for Laboratory Medicine https://creativecommons.org/licenses/by-nc/4.0/This is an open-access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0 (https://creativecommons.org/licenses/by-nc/4.0/) ) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
Cho, Sung-Eun
Han, Jungsun
Park, Ju-Hee
Park, Euna
Kim, Geun Young
Lee, Jun Hyung
Yi, Ahram
Lee, Sang Gon
Lee, Eun Hee
Yun, Yeo-Min
Clinical Usefulness of Ultraperformance Liquid Chromatography-Tandem Mass Spectrometry Method for Low Serum Testosterone Measurement
title Clinical Usefulness of Ultraperformance Liquid Chromatography-Tandem Mass Spectrometry Method for Low Serum Testosterone Measurement
title_full Clinical Usefulness of Ultraperformance Liquid Chromatography-Tandem Mass Spectrometry Method for Low Serum Testosterone Measurement
title_fullStr Clinical Usefulness of Ultraperformance Liquid Chromatography-Tandem Mass Spectrometry Method for Low Serum Testosterone Measurement
title_full_unstemmed Clinical Usefulness of Ultraperformance Liquid Chromatography-Tandem Mass Spectrometry Method for Low Serum Testosterone Measurement
title_short Clinical Usefulness of Ultraperformance Liquid Chromatography-Tandem Mass Spectrometry Method for Low Serum Testosterone Measurement
title_sort clinical usefulness of ultraperformance liquid chromatography-tandem mass spectrometry method for low serum testosterone measurement
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9467846/
https://www.ncbi.nlm.nih.gov/pubmed/36045053
http://dx.doi.org/10.3343/alm.2023.43.1.19
work_keys_str_mv AT chosungeun clinicalusefulnessofultraperformanceliquidchromatographytandemmassspectrometrymethodforlowserumtestosteronemeasurement
AT hanjungsun clinicalusefulnessofultraperformanceliquidchromatographytandemmassspectrometrymethodforlowserumtestosteronemeasurement
AT parkjuhee clinicalusefulnessofultraperformanceliquidchromatographytandemmassspectrometrymethodforlowserumtestosteronemeasurement
AT parkeuna clinicalusefulnessofultraperformanceliquidchromatographytandemmassspectrometrymethodforlowserumtestosteronemeasurement
AT kimgeunyoung clinicalusefulnessofultraperformanceliquidchromatographytandemmassspectrometrymethodforlowserumtestosteronemeasurement
AT leejunhyung clinicalusefulnessofultraperformanceliquidchromatographytandemmassspectrometrymethodforlowserumtestosteronemeasurement
AT yiahram clinicalusefulnessofultraperformanceliquidchromatographytandemmassspectrometrymethodforlowserumtestosteronemeasurement
AT leesanggon clinicalusefulnessofultraperformanceliquidchromatographytandemmassspectrometrymethodforlowserumtestosteronemeasurement
AT leeeunhee clinicalusefulnessofultraperformanceliquidchromatographytandemmassspectrometrymethodforlowserumtestosteronemeasurement
AT yunyeomin clinicalusefulnessofultraperformanceliquidchromatographytandemmassspectrometrymethodforlowserumtestosteronemeasurement