Cargando…

Manual versus Mechanical Delivery of High-Quality Cardiopulmonary Resuscitation on a River-Based Fire Rescue Boat

OBJECTIVES: Studies have demonstrated the efficacy of mechanical devices at delivering high-quality cardiopulmonary resuscitation (HQ-CPR) in various transport settings. Herein, this study investigates the efficacy of manual and mechanical HQ-CPR delivery on a fire rescue boat. METHODS: A total of 1...

Descripción completa

Detalles Bibliográficos
Autores principales: Manoukian, Martin A.C., Tancredi, Daniel J., Linvill, Matthew T., Wynia, Elisabeth H., Beaver, Brianna, Rose, John S., Mumma, Bryn E.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Cambridge University Press 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9470525/
https://www.ncbi.nlm.nih.gov/pubmed/35875994
http://dx.doi.org/10.1017/S1049023X22001042
Descripción
Sumario:OBJECTIVES: Studies have demonstrated the efficacy of mechanical devices at delivering high-quality cardiopulmonary resuscitation (HQ-CPR) in various transport settings. Herein, this study investigates the efficacy of manual and mechanical HQ-CPR delivery on a fire rescue boat. METHODS: A total of 15 active firefighter-paramedics were recruited for a prospective manikin-based trial. Each paramedic performed two minutes manual compression-only CPR while navigating on a river-based fire rescue boat. The boat was piloted in either a stable linear manner or dynamic S-turn manner to simulate obstacle avoidance. For each session of manual HQ-CPR, a session of mechanical HQ-CPR was also performed with a LUCAS 3 (Stryker; Kalamazoo, Michigan USA). A total of 60 sessions were completed. Parameters recorded included compression fraction (CF) and the percentage of compressions with correct depth >5cm (D%), correct rate 100-120 (R%), full release (FR%), and correct hand position (HP%). A composite HQ-CPR score was calculated as follows: ((D% + R% + FR% + HP%)/4) * CF%). Differences in magnitude of change seen in stable versus dynamic navigation within study conditions were evaluated with a Z-score calculation. Difficulty of HQ-CPR delivery was assessed utilizing the Borg Rating of Perceived Exertion Scale. RESULTS: Participants were mostly male and had a median experience of 20 years. Manual HQ-CPR delivered during stable navigation out-performed manual HQ-CPR delivered during dynamic navigation for composite score and trended towards superiority for FR% and R%. There was no difference seen for any measured variable when comparing mechanical HQ-CPR delivered during stable navigation versus dynamic navigation. Mechanical HQ-CPR out-performed manual HQ-CPR during both stable and dynamic navigation in terms of composite score, FR%, and R%. Z-score calculation demonstrated that manual HQ-CPR delivery was significantly more affected by drive style than mechanical HQ-CPR delivery in terms of composite HQ-CPR score and trended towards significance for FR% and R%. Borg Rating of Perceived Exertion was higher for manual CPR delivered during dynamic sessions than for stable sessions. CONCLUSION: Mechanical HQ-CPR delivery is superior to manual HQ-CPR delivery during both stable and dynamic riverine navigation. Whereas manual HQ-CPR delivery was worse during dynamic transportation conditions compared to stable transport conditions, mechanical HQ-CPR delivery was unaffected by drive style. This suggests the utility of routine use of mechanical HQ-CPR devices in the riverine patient transport setting.