Intracochlear electrode array position and cochlear implant outcomes using the nucleus slim modiolar electrode and the extended round window approach: a follow-up study

PURPOSE: The aim of this study was to evaluate the intracochlear position of the Slim Modiolar Electrode (SME) after insertion via the extended Round Window (eRW) approach, and to correlate this with residual hearing preservation and speech perception outcomes. METHODS: Twenty-three adult participan...

Descripción completa

Detalles Bibliográficos
Autores principales: Klabbers, Tim M., Heutink, Floris, Huinck, Wendy J., van der Woude, Willem-Jan, Verbist, Berit M., Mylanus, Emmanuel A. M.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Berlin Heidelberg 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9474433/
https://www.ncbi.nlm.nih.gov/pubmed/35041067
http://dx.doi.org/10.1007/s00405-021-07247-w
_version_ 1784789715892830208
author Klabbers, Tim M.
Heutink, Floris
Huinck, Wendy J.
van der Woude, Willem-Jan
Verbist, Berit M.
Mylanus, Emmanuel A. M.
author_facet Klabbers, Tim M.
Heutink, Floris
Huinck, Wendy J.
van der Woude, Willem-Jan
Verbist, Berit M.
Mylanus, Emmanuel A. M.
author_sort Klabbers, Tim M.
collection PubMed
description PURPOSE: The aim of this study was to evaluate the intracochlear position of the Slim Modiolar Electrode (SME) after insertion via the extended Round Window (eRW) approach, and to correlate this with residual hearing preservation and speech perception outcomes. METHODS: Twenty-three adult participants, consecutively implanted with the SME via the eRW approach, were included in this prospective, single-center, observational study. Electrode position was evaluated intra-operatively using X-ray fluoroscopy and TIM measurement, and post-operatively using ultra-high resolution CT. Residual hearing [threshold shift in PTA between pre- and post-operative measurement, relative hearing preservation (RHP%)] and speech perception were evaluated at 2 and 12 months after surgery. RESULTS: In each of the 23 participants, complete scala tympani positioning of the electrode array could be achieved. In one participant, an initial tip fold-over was corrected during surgery. Average age at implantation was 63.3 years (SD 13.3, range 28–76) and mean preoperative residual hearing was 81.5 dB. The average post-operative PTA threshold shift was 16.2 dB (SD 10.8) at 2 months post-operatively, corresponding with a RHP% score of 44% (SD 34.9). At 12 months, the average RHP% score decreased to 37%. Postoperative phoneme scores improved from 27.1% preoperatively, to 72.1% and 82.1% at 2 and 12 months after surgery, respectively. CONCLUSION: Use of the eRW approach results in an increased likelihood of complete scala tympani insertion when inserting the SME, with subsequent excellent levels of speech perception. However, residual hearing preservation was found to be moderate, possibly as a result of the extended round window approach, emphasizing that it is not an all-purpose approach for inserting this particular electrode array.
format Online
Article
Text
id pubmed-9474433
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Springer Berlin Heidelberg
record_format MEDLINE/PubMed
spelling pubmed-94744332022-09-16 Intracochlear electrode array position and cochlear implant outcomes using the nucleus slim modiolar electrode and the extended round window approach: a follow-up study Klabbers, Tim M. Heutink, Floris Huinck, Wendy J. van der Woude, Willem-Jan Verbist, Berit M. Mylanus, Emmanuel A. M. Eur Arch Otorhinolaryngol Otology PURPOSE: The aim of this study was to evaluate the intracochlear position of the Slim Modiolar Electrode (SME) after insertion via the extended Round Window (eRW) approach, and to correlate this with residual hearing preservation and speech perception outcomes. METHODS: Twenty-three adult participants, consecutively implanted with the SME via the eRW approach, were included in this prospective, single-center, observational study. Electrode position was evaluated intra-operatively using X-ray fluoroscopy and TIM measurement, and post-operatively using ultra-high resolution CT. Residual hearing [threshold shift in PTA between pre- and post-operative measurement, relative hearing preservation (RHP%)] and speech perception were evaluated at 2 and 12 months after surgery. RESULTS: In each of the 23 participants, complete scala tympani positioning of the electrode array could be achieved. In one participant, an initial tip fold-over was corrected during surgery. Average age at implantation was 63.3 years (SD 13.3, range 28–76) and mean preoperative residual hearing was 81.5 dB. The average post-operative PTA threshold shift was 16.2 dB (SD 10.8) at 2 months post-operatively, corresponding with a RHP% score of 44% (SD 34.9). At 12 months, the average RHP% score decreased to 37%. Postoperative phoneme scores improved from 27.1% preoperatively, to 72.1% and 82.1% at 2 and 12 months after surgery, respectively. CONCLUSION: Use of the eRW approach results in an increased likelihood of complete scala tympani insertion when inserting the SME, with subsequent excellent levels of speech perception. However, residual hearing preservation was found to be moderate, possibly as a result of the extended round window approach, emphasizing that it is not an all-purpose approach for inserting this particular electrode array. Springer Berlin Heidelberg 2022-01-18 2022 /pmc/articles/PMC9474433/ /pubmed/35041067 http://dx.doi.org/10.1007/s00405-021-07247-w Text en © The Author(s) 2022 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) .
spellingShingle Otology
Klabbers, Tim M.
Heutink, Floris
Huinck, Wendy J.
van der Woude, Willem-Jan
Verbist, Berit M.
Mylanus, Emmanuel A. M.
Intracochlear electrode array position and cochlear implant outcomes using the nucleus slim modiolar electrode and the extended round window approach: a follow-up study
title Intracochlear electrode array position and cochlear implant outcomes using the nucleus slim modiolar electrode and the extended round window approach: a follow-up study
title_full Intracochlear electrode array position and cochlear implant outcomes using the nucleus slim modiolar electrode and the extended round window approach: a follow-up study
title_fullStr Intracochlear electrode array position and cochlear implant outcomes using the nucleus slim modiolar electrode and the extended round window approach: a follow-up study
title_full_unstemmed Intracochlear electrode array position and cochlear implant outcomes using the nucleus slim modiolar electrode and the extended round window approach: a follow-up study
title_short Intracochlear electrode array position and cochlear implant outcomes using the nucleus slim modiolar electrode and the extended round window approach: a follow-up study
title_sort intracochlear electrode array position and cochlear implant outcomes using the nucleus slim modiolar electrode and the extended round window approach: a follow-up study
topic Otology
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9474433/
https://www.ncbi.nlm.nih.gov/pubmed/35041067
http://dx.doi.org/10.1007/s00405-021-07247-w
work_keys_str_mv AT klabberstimm intracochlearelectrodearraypositionandcochlearimplantoutcomesusingthenucleusslimmodiolarelectrodeandtheextendedroundwindowapproachafollowupstudy
AT heutinkfloris intracochlearelectrodearraypositionandcochlearimplantoutcomesusingthenucleusslimmodiolarelectrodeandtheextendedroundwindowapproachafollowupstudy
AT huinckwendyj intracochlearelectrodearraypositionandcochlearimplantoutcomesusingthenucleusslimmodiolarelectrodeandtheextendedroundwindowapproachafollowupstudy
AT vanderwoudewillemjan intracochlearelectrodearraypositionandcochlearimplantoutcomesusingthenucleusslimmodiolarelectrodeandtheextendedroundwindowapproachafollowupstudy
AT verbistberitm intracochlearelectrodearraypositionandcochlearimplantoutcomesusingthenucleusslimmodiolarelectrodeandtheextendedroundwindowapproachafollowupstudy
AT mylanusemmanuelam intracochlearelectrodearraypositionandcochlearimplantoutcomesusingthenucleusslimmodiolarelectrodeandtheextendedroundwindowapproachafollowupstudy