Comparison between a dual-time-window protocol and other simplified protocols for dynamic total-body (18)F-FDG PET imaging

PURPOSE: Efforts have been made both to avoid invasive blood sampling and to shorten the scan duration for dynamic positron emission tomography (PET) imaging. A total-body scanner, such as the uEXPLORER PET/CT, can relieve these challenges through the following features: First, the whole-body covera...

Descripción completa

Detalles Bibliográficos
Autores principales: Wang, Zhenguo, Wu, Yaping, Li, Xiaochen, Bai, Yan, Chen, Hongzhao, Ding, Jie, Shen, Chushu, Hu, Zhanli, Liang, Dong, Liu, Xin, Zheng, Hairong, Yang, Yongfeng, Zhou, Yun, Wang, Meiyun, Sun, Tao
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer International Publishing 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9474964/
https://www.ncbi.nlm.nih.gov/pubmed/36104580
http://dx.doi.org/10.1186/s40658-022-00492-w
_version_ 1784789805411860480
author Wang, Zhenguo
Wu, Yaping
Li, Xiaochen
Bai, Yan
Chen, Hongzhao
Ding, Jie
Shen, Chushu
Hu, Zhanli
Liang, Dong
Liu, Xin
Zheng, Hairong
Yang, Yongfeng
Zhou, Yun
Wang, Meiyun
Sun, Tao
author_facet Wang, Zhenguo
Wu, Yaping
Li, Xiaochen
Bai, Yan
Chen, Hongzhao
Ding, Jie
Shen, Chushu
Hu, Zhanli
Liang, Dong
Liu, Xin
Zheng, Hairong
Yang, Yongfeng
Zhou, Yun
Wang, Meiyun
Sun, Tao
author_sort Wang, Zhenguo
collection PubMed
description PURPOSE: Efforts have been made both to avoid invasive blood sampling and to shorten the scan duration for dynamic positron emission tomography (PET) imaging. A total-body scanner, such as the uEXPLORER PET/CT, can relieve these challenges through the following features: First, the whole-body coverage allows for noninvasive input function from the aortic arteries; second, with a dramatic increase in sensitivity, image quality can still be maintained at a high level even with a shorter scan duration than usual. We implemented a dual-time-window (DTW) protocol for a dynamic total-body (18)F-FDG PET scan to obtain multiple kinetic parameters. The DTW protocol was then compared to several other simplified quantification methods for total-body FDG imaging that were proposed for conventional setup. METHODS: The research included 28 patient scans performed on an uEXPLORER PET/CT. By discarding the corresponding data in the middle of the existing full 60-min dynamic scan, the DTW protocol was simulated. Nonlinear fitting was used to estimate the missing data in the interval. The full input function was obtained from 15 subjects using a hybrid approach with a population-based image-derived input function. Quantification was carried out in three areas: the cerebral cortex, muscle, and tumor lesion. Micro- and macro-kinetic parameters for different scan durations were estimated by assuming an irreversible two-tissue compartment model. The visual performance of parametric images and region of interest-based quantification in several parameters were evaluated. Furthermore, simplified quantification methods (DTW, Patlak, fractional uptake ratio [FUR], and standardized uptake value [SUV]) were compared for similarity to the reference net influx rate K(i). RESULTS: K(i) and K(1) derived from the DTW protocol showed overall good consistency (P < 0.01) with the reference from the 60-min dynamic scan with 10-min early scan and 5-min late scan (K(i) correlation: 0.971, 0.990, and 0.990; K(1) correlation: 0.820, 0.940, and 0.975 in the cerebral cortex, muscle, and tumor lesion, respectively). Similar correlationss were found for other micro-parameters. The DTW protocol had the lowest bias relative to standard K(i) than any of the quantification methods, followed by FUR and Patlak. SUV had the weakest correlation with K(i). The whole-body K(i) and K(1) images generated by the DTW protocol were consistent with the reference parametric images. CONCLUSIONS: Using the DTW protocol, the dynamic total-body FDG scan time can be reduced to 15 min while obtaining accurate K(i) and K(1) quantification and acceptable visual performance in parametric images. However, the trade-off between quantification accuracy and protocol implementation feasibility must be considered in practice. We recommend that the DTW protocol be used when the clinical task requires reliable visual assessment or quantifying multiple micro-parameters; FUR with a hybrid input function may be a more feasible approach to quantifying regional metabolic rate with a known lesion position or organs of interest. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1186/s40658-022-00492-w.
format Online
Article
Text
id pubmed-9474964
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Springer International Publishing
record_format MEDLINE/PubMed
spelling pubmed-94749642022-09-16 Comparison between a dual-time-window protocol and other simplified protocols for dynamic total-body (18)F-FDG PET imaging Wang, Zhenguo Wu, Yaping Li, Xiaochen Bai, Yan Chen, Hongzhao Ding, Jie Shen, Chushu Hu, Zhanli Liang, Dong Liu, Xin Zheng, Hairong Yang, Yongfeng Zhou, Yun Wang, Meiyun Sun, Tao EJNMMI Phys Original Research PURPOSE: Efforts have been made both to avoid invasive blood sampling and to shorten the scan duration for dynamic positron emission tomography (PET) imaging. A total-body scanner, such as the uEXPLORER PET/CT, can relieve these challenges through the following features: First, the whole-body coverage allows for noninvasive input function from the aortic arteries; second, with a dramatic increase in sensitivity, image quality can still be maintained at a high level even with a shorter scan duration than usual. We implemented a dual-time-window (DTW) protocol for a dynamic total-body (18)F-FDG PET scan to obtain multiple kinetic parameters. The DTW protocol was then compared to several other simplified quantification methods for total-body FDG imaging that were proposed for conventional setup. METHODS: The research included 28 patient scans performed on an uEXPLORER PET/CT. By discarding the corresponding data in the middle of the existing full 60-min dynamic scan, the DTW protocol was simulated. Nonlinear fitting was used to estimate the missing data in the interval. The full input function was obtained from 15 subjects using a hybrid approach with a population-based image-derived input function. Quantification was carried out in three areas: the cerebral cortex, muscle, and tumor lesion. Micro- and macro-kinetic parameters for different scan durations were estimated by assuming an irreversible two-tissue compartment model. The visual performance of parametric images and region of interest-based quantification in several parameters were evaluated. Furthermore, simplified quantification methods (DTW, Patlak, fractional uptake ratio [FUR], and standardized uptake value [SUV]) were compared for similarity to the reference net influx rate K(i). RESULTS: K(i) and K(1) derived from the DTW protocol showed overall good consistency (P < 0.01) with the reference from the 60-min dynamic scan with 10-min early scan and 5-min late scan (K(i) correlation: 0.971, 0.990, and 0.990; K(1) correlation: 0.820, 0.940, and 0.975 in the cerebral cortex, muscle, and tumor lesion, respectively). Similar correlationss were found for other micro-parameters. The DTW protocol had the lowest bias relative to standard K(i) than any of the quantification methods, followed by FUR and Patlak. SUV had the weakest correlation with K(i). The whole-body K(i) and K(1) images generated by the DTW protocol were consistent with the reference parametric images. CONCLUSIONS: Using the DTW protocol, the dynamic total-body FDG scan time can be reduced to 15 min while obtaining accurate K(i) and K(1) quantification and acceptable visual performance in parametric images. However, the trade-off between quantification accuracy and protocol implementation feasibility must be considered in practice. We recommend that the DTW protocol be used when the clinical task requires reliable visual assessment or quantifying multiple micro-parameters; FUR with a hybrid input function may be a more feasible approach to quantifying regional metabolic rate with a known lesion position or organs of interest. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1186/s40658-022-00492-w. Springer International Publishing 2022-09-14 /pmc/articles/PMC9474964/ /pubmed/36104580 http://dx.doi.org/10.1186/s40658-022-00492-w Text en © The Author(s) 2022 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) .
spellingShingle Original Research
Wang, Zhenguo
Wu, Yaping
Li, Xiaochen
Bai, Yan
Chen, Hongzhao
Ding, Jie
Shen, Chushu
Hu, Zhanli
Liang, Dong
Liu, Xin
Zheng, Hairong
Yang, Yongfeng
Zhou, Yun
Wang, Meiyun
Sun, Tao
Comparison between a dual-time-window protocol and other simplified protocols for dynamic total-body (18)F-FDG PET imaging
title Comparison between a dual-time-window protocol and other simplified protocols for dynamic total-body (18)F-FDG PET imaging
title_full Comparison between a dual-time-window protocol and other simplified protocols for dynamic total-body (18)F-FDG PET imaging
title_fullStr Comparison between a dual-time-window protocol and other simplified protocols for dynamic total-body (18)F-FDG PET imaging
title_full_unstemmed Comparison between a dual-time-window protocol and other simplified protocols for dynamic total-body (18)F-FDG PET imaging
title_short Comparison between a dual-time-window protocol and other simplified protocols for dynamic total-body (18)F-FDG PET imaging
title_sort comparison between a dual-time-window protocol and other simplified protocols for dynamic total-body (18)f-fdg pet imaging
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9474964/
https://www.ncbi.nlm.nih.gov/pubmed/36104580
http://dx.doi.org/10.1186/s40658-022-00492-w
work_keys_str_mv AT wangzhenguo comparisonbetweenadualtimewindowprotocolandothersimplifiedprotocolsfordynamictotalbody18ffdgpetimaging
AT wuyaping comparisonbetweenadualtimewindowprotocolandothersimplifiedprotocolsfordynamictotalbody18ffdgpetimaging
AT lixiaochen comparisonbetweenadualtimewindowprotocolandothersimplifiedprotocolsfordynamictotalbody18ffdgpetimaging
AT baiyan comparisonbetweenadualtimewindowprotocolandothersimplifiedprotocolsfordynamictotalbody18ffdgpetimaging
AT chenhongzhao comparisonbetweenadualtimewindowprotocolandothersimplifiedprotocolsfordynamictotalbody18ffdgpetimaging
AT dingjie comparisonbetweenadualtimewindowprotocolandothersimplifiedprotocolsfordynamictotalbody18ffdgpetimaging
AT shenchushu comparisonbetweenadualtimewindowprotocolandothersimplifiedprotocolsfordynamictotalbody18ffdgpetimaging
AT huzhanli comparisonbetweenadualtimewindowprotocolandothersimplifiedprotocolsfordynamictotalbody18ffdgpetimaging
AT liangdong comparisonbetweenadualtimewindowprotocolandothersimplifiedprotocolsfordynamictotalbody18ffdgpetimaging
AT liuxin comparisonbetweenadualtimewindowprotocolandothersimplifiedprotocolsfordynamictotalbody18ffdgpetimaging
AT zhenghairong comparisonbetweenadualtimewindowprotocolandothersimplifiedprotocolsfordynamictotalbody18ffdgpetimaging
AT yangyongfeng comparisonbetweenadualtimewindowprotocolandothersimplifiedprotocolsfordynamictotalbody18ffdgpetimaging
AT zhouyun comparisonbetweenadualtimewindowprotocolandothersimplifiedprotocolsfordynamictotalbody18ffdgpetimaging
AT wangmeiyun comparisonbetweenadualtimewindowprotocolandothersimplifiedprotocolsfordynamictotalbody18ffdgpetimaging
AT suntao comparisonbetweenadualtimewindowprotocolandothersimplifiedprotocolsfordynamictotalbody18ffdgpetimaging