Practical method for RF pulse distortion compensation using multiple square pulses for low-field MRI
Since recovery time of the RF coil is long at low field MRI, the rising and the ring-down times of the square pulse are also long, which means the applied sinc pulse can easily be distorted from the changing amplitude. However, both the rising time and ring-down time can be calculated using Q-factor...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Public Library of Science
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9481036/ https://www.ncbi.nlm.nih.gov/pubmed/36112594 http://dx.doi.org/10.1371/journal.pone.0273432 |
Sumario: | Since recovery time of the RF coil is long at low field MRI, the rising and the ring-down times of the square pulse are also long, which means the applied sinc pulse can easily be distorted from the changing amplitude. However, both the rising time and ring-down time can be calculated using Q-factor. Using this information, an RF square pulse were compensated by appending two square pulses before and after the RF pulse. The durations of these RF square pulses were calculated using the Q-factor. Since the amplitude of the sinc pulse changes continuously, a series of square pulses were applied to apply sinc pulse to the coil. The minimum number of square pulses and the amplitude of the square pulses were calculated. It was successfully demonstrated that the sinc pulse can be compensated using a series of square pulses. The more number of square pulses were used, the smoother sinc pulse was applied to the RF coil. The Q-factor was experimentally calculated from the ring-down time of a signal induced in a sniffer loop which was connected to an oscilloscope. The resulting Q-factor was then used to calculate both the duration and amplitude of the square pulses for compensation. Echo trains were also acquired in an inhomogeneous B0 field using the compensated RF pulses. In order to enhance the SNR of the echo trains, a pre-polarization pulse was added to the CPMG spin echo sequence. The SNRs of the echo signal acquired using compensated pulses were compared with those of signal obtained with uncompensated pulses and showed significant improvements of 61.1% and 51.5% for the square and sinc shaped pulses respectively. |
---|