Practical method for RF pulse distortion compensation using multiple square pulses for low-field MRI

Since recovery time of the RF coil is long at low field MRI, the rising and the ring-down times of the square pulse are also long, which means the applied sinc pulse can easily be distorted from the changing amplitude. However, both the rising time and ring-down time can be calculated using Q-factor...

Descripción completa

Detalles Bibliográficos
Autores principales: Ha, Yonghyun, Selvaganesan, Kartiga, Wu, Baosong, Hancock, Kasey, Rogers, Charles, Hosseinnezhadian, Sajad, Galiana, Gigi, Constable, R. Todd
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9481036/
https://www.ncbi.nlm.nih.gov/pubmed/36112594
http://dx.doi.org/10.1371/journal.pone.0273432
_version_ 1784791173318049792
author Ha, Yonghyun
Selvaganesan, Kartiga
Wu, Baosong
Hancock, Kasey
Rogers, Charles
Hosseinnezhadian, Sajad
Galiana, Gigi
Constable, R. Todd
author_facet Ha, Yonghyun
Selvaganesan, Kartiga
Wu, Baosong
Hancock, Kasey
Rogers, Charles
Hosseinnezhadian, Sajad
Galiana, Gigi
Constable, R. Todd
author_sort Ha, Yonghyun
collection PubMed
description Since recovery time of the RF coil is long at low field MRI, the rising and the ring-down times of the square pulse are also long, which means the applied sinc pulse can easily be distorted from the changing amplitude. However, both the rising time and ring-down time can be calculated using Q-factor. Using this information, an RF square pulse were compensated by appending two square pulses before and after the RF pulse. The durations of these RF square pulses were calculated using the Q-factor. Since the amplitude of the sinc pulse changes continuously, a series of square pulses were applied to apply sinc pulse to the coil. The minimum number of square pulses and the amplitude of the square pulses were calculated. It was successfully demonstrated that the sinc pulse can be compensated using a series of square pulses. The more number of square pulses were used, the smoother sinc pulse was applied to the RF coil. The Q-factor was experimentally calculated from the ring-down time of a signal induced in a sniffer loop which was connected to an oscilloscope. The resulting Q-factor was then used to calculate both the duration and amplitude of the square pulses for compensation. Echo trains were also acquired in an inhomogeneous B0 field using the compensated RF pulses. In order to enhance the SNR of the echo trains, a pre-polarization pulse was added to the CPMG spin echo sequence. The SNRs of the echo signal acquired using compensated pulses were compared with those of signal obtained with uncompensated pulses and showed significant improvements of 61.1% and 51.5% for the square and sinc shaped pulses respectively.
format Online
Article
Text
id pubmed-9481036
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Public Library of Science
record_format MEDLINE/PubMed
spelling pubmed-94810362022-09-17 Practical method for RF pulse distortion compensation using multiple square pulses for low-field MRI Ha, Yonghyun Selvaganesan, Kartiga Wu, Baosong Hancock, Kasey Rogers, Charles Hosseinnezhadian, Sajad Galiana, Gigi Constable, R. Todd PLoS One Research Article Since recovery time of the RF coil is long at low field MRI, the rising and the ring-down times of the square pulse are also long, which means the applied sinc pulse can easily be distorted from the changing amplitude. However, both the rising time and ring-down time can be calculated using Q-factor. Using this information, an RF square pulse were compensated by appending two square pulses before and after the RF pulse. The durations of these RF square pulses were calculated using the Q-factor. Since the amplitude of the sinc pulse changes continuously, a series of square pulses were applied to apply sinc pulse to the coil. The minimum number of square pulses and the amplitude of the square pulses were calculated. It was successfully demonstrated that the sinc pulse can be compensated using a series of square pulses. The more number of square pulses were used, the smoother sinc pulse was applied to the RF coil. The Q-factor was experimentally calculated from the ring-down time of a signal induced in a sniffer loop which was connected to an oscilloscope. The resulting Q-factor was then used to calculate both the duration and amplitude of the square pulses for compensation. Echo trains were also acquired in an inhomogeneous B0 field using the compensated RF pulses. In order to enhance the SNR of the echo trains, a pre-polarization pulse was added to the CPMG spin echo sequence. The SNRs of the echo signal acquired using compensated pulses were compared with those of signal obtained with uncompensated pulses and showed significant improvements of 61.1% and 51.5% for the square and sinc shaped pulses respectively. Public Library of Science 2022-09-16 /pmc/articles/PMC9481036/ /pubmed/36112594 http://dx.doi.org/10.1371/journal.pone.0273432 Text en © 2022 Ha et al https://creativecommons.org/licenses/by/4.0/This is an open access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/) , which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
spellingShingle Research Article
Ha, Yonghyun
Selvaganesan, Kartiga
Wu, Baosong
Hancock, Kasey
Rogers, Charles
Hosseinnezhadian, Sajad
Galiana, Gigi
Constable, R. Todd
Practical method for RF pulse distortion compensation using multiple square pulses for low-field MRI
title Practical method for RF pulse distortion compensation using multiple square pulses for low-field MRI
title_full Practical method for RF pulse distortion compensation using multiple square pulses for low-field MRI
title_fullStr Practical method for RF pulse distortion compensation using multiple square pulses for low-field MRI
title_full_unstemmed Practical method for RF pulse distortion compensation using multiple square pulses for low-field MRI
title_short Practical method for RF pulse distortion compensation using multiple square pulses for low-field MRI
title_sort practical method for rf pulse distortion compensation using multiple square pulses for low-field mri
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9481036/
https://www.ncbi.nlm.nih.gov/pubmed/36112594
http://dx.doi.org/10.1371/journal.pone.0273432
work_keys_str_mv AT hayonghyun practicalmethodforrfpulsedistortioncompensationusingmultiplesquarepulsesforlowfieldmri
AT selvaganesankartiga practicalmethodforrfpulsedistortioncompensationusingmultiplesquarepulsesforlowfieldmri
AT wubaosong practicalmethodforrfpulsedistortioncompensationusingmultiplesquarepulsesforlowfieldmri
AT hancockkasey practicalmethodforrfpulsedistortioncompensationusingmultiplesquarepulsesforlowfieldmri
AT rogerscharles practicalmethodforrfpulsedistortioncompensationusingmultiplesquarepulsesforlowfieldmri
AT hosseinnezhadiansajad practicalmethodforrfpulsedistortioncompensationusingmultiplesquarepulsesforlowfieldmri
AT galianagigi practicalmethodforrfpulsedistortioncompensationusingmultiplesquarepulsesforlowfieldmri
AT constablertodd practicalmethodforrfpulsedistortioncompensationusingmultiplesquarepulsesforlowfieldmri