Low-level (40)Ca determinations using nitrous oxide with reaction cell inductively coupled plasma–tandem mass spectrometry

In inductively coupled plasma mass spectrometry, the most abundant Ca isotope ((40)Ca) suffers from isobaric interference with argon, hindering the potential for low detection limits of Ca. A powerful approach is to remove the interference by using a reaction gas in a reaction cell. Ammonia (NH(3))...

Descripción completa

Detalles Bibliográficos
Autores principales: Lancaster, Shaun T., Prohaska, Thomas, Irrgeher, Johanna
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Berlin Heidelberg 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9482902/
https://www.ncbi.nlm.nih.gov/pubmed/35641642
http://dx.doi.org/10.1007/s00216-022-04146-9
Descripción
Sumario:In inductively coupled plasma mass spectrometry, the most abundant Ca isotope ((40)Ca) suffers from isobaric interference with argon, hindering the potential for low detection limits of Ca. A powerful approach is to remove the interference by using a reaction gas in a reaction cell. Ammonia (NH(3)) has proven to be an effective reaction gas by process of a charge transfer reaction. However, NH(3) is highly corrosive and toxic and cannot remove isobaric (40) K. Therefore, this work proposes the use of nitrous oxide (N(2)O) to mass shift the target analyte (40)Ca to (40)Ca(16)O(+) as a non-corrosive and non-toxic alternative. Instrument performance testing demonstrated that N(2)O was capable of reaching equivalent detection limits (0.015 ng g(−1)) and background equivalence concentrations (0.041 ng g(−1)) to that of NH(3) and limited by the blank only. Further investigation of matrix interferences with synthetic standards highlighted that the N(2)O approach supports the separation of potassium (K) and magnesium (Mg)–based interferences at tested concentrations of more than 600 times and almost 800 times higher than Ca respectively, whereas NH(3) was found to only support the removal of Mg. This work highlights a clear advantage of N(2)O for low-level Ca determinations with high matrix loads, as well as compatibility with other instrumentation sensitive to corrosion that supports reaction cell technology. GRAPHICAL ABSTRACT: [Image: see text]