Cargando…

Inappropriate shock delivery as a result of electromagnetic interference originating from the faulty electrical installation

We present a case report of a 74‐year‐old male patient with an implantable cardioverter defibrillator who suffered an inappropriate defibrillation shock while bathing in the tub. Insight in the ICD stored electrogram episodes revealed electromagnetic interferences, with a typical 50 Hz electrical ar...

Descripción completa

Detalles Bibliográficos
Autores principales: Babic, Milos D, Tomovic, Milosav, Milosevic, Maja, Djurdjevic, Branko, Zugic, Vasko, Nikolic, Aleksandra
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9484028/
https://www.ncbi.nlm.nih.gov/pubmed/35467789
http://dx.doi.org/10.1111/anec.12952
_version_ 1784791798852354048
author Babic, Milos D
Tomovic, Milosav
Milosevic, Maja
Djurdjevic, Branko
Zugic, Vasko
Nikolic, Aleksandra
author_facet Babic, Milos D
Tomovic, Milosav
Milosevic, Maja
Djurdjevic, Branko
Zugic, Vasko
Nikolic, Aleksandra
author_sort Babic, Milos D
collection PubMed
description We present a case report of a 74‐year‐old male patient with an implantable cardioverter defibrillator who suffered an inappropriate defibrillation shock while bathing in the tub. Insight in the ICD stored electrogram episodes revealed electromagnetic interferences, with a typical 50 Hz electrical artifact mimicking fast ventricular tachycardia as a device misinterpreted. After this event, the maintenance workers investigated the electrical installation in the bathroom and revealed that there was voltage leaking between electrical installation and metal pipes. After the repair was completed without any additional programming, the patient has had no subsequent shocks.
format Online
Article
Text
id pubmed-9484028
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-94840282022-09-29 Inappropriate shock delivery as a result of electromagnetic interference originating from the faulty electrical installation Babic, Milos D Tomovic, Milosav Milosevic, Maja Djurdjevic, Branko Zugic, Vasko Nikolic, Aleksandra Ann Noninvasive Electrocardiol Case Reports We present a case report of a 74‐year‐old male patient with an implantable cardioverter defibrillator who suffered an inappropriate defibrillation shock while bathing in the tub. Insight in the ICD stored electrogram episodes revealed electromagnetic interferences, with a typical 50 Hz electrical artifact mimicking fast ventricular tachycardia as a device misinterpreted. After this event, the maintenance workers investigated the electrical installation in the bathroom and revealed that there was voltage leaking between electrical installation and metal pipes. After the repair was completed without any additional programming, the patient has had no subsequent shocks. John Wiley and Sons Inc. 2022-04-25 /pmc/articles/PMC9484028/ /pubmed/35467789 http://dx.doi.org/10.1111/anec.12952 Text en © 2022 The Authors. Annals of Noninvasive Electrocardiology published by Wiley Periodicals LLC. https://creativecommons.org/licenses/by-nc-nd/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by-nc-nd/4.0/ (https://creativecommons.org/licenses/by-nc-nd/4.0/) License, which permits use and distribution in any medium, provided the original work is properly cited, the use is non‐commercial and no modifications or adaptations are made.
spellingShingle Case Reports
Babic, Milos D
Tomovic, Milosav
Milosevic, Maja
Djurdjevic, Branko
Zugic, Vasko
Nikolic, Aleksandra
Inappropriate shock delivery as a result of electromagnetic interference originating from the faulty electrical installation
title Inappropriate shock delivery as a result of electromagnetic interference originating from the faulty electrical installation
title_full Inappropriate shock delivery as a result of electromagnetic interference originating from the faulty electrical installation
title_fullStr Inappropriate shock delivery as a result of electromagnetic interference originating from the faulty electrical installation
title_full_unstemmed Inappropriate shock delivery as a result of electromagnetic interference originating from the faulty electrical installation
title_short Inappropriate shock delivery as a result of electromagnetic interference originating from the faulty electrical installation
title_sort inappropriate shock delivery as a result of electromagnetic interference originating from the faulty electrical installation
topic Case Reports
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9484028/
https://www.ncbi.nlm.nih.gov/pubmed/35467789
http://dx.doi.org/10.1111/anec.12952
work_keys_str_mv AT babicmilosd inappropriateshockdeliveryasaresultofelectromagneticinterferenceoriginatingfromthefaultyelectricalinstallation
AT tomovicmilosav inappropriateshockdeliveryasaresultofelectromagneticinterferenceoriginatingfromthefaultyelectricalinstallation
AT milosevicmaja inappropriateshockdeliveryasaresultofelectromagneticinterferenceoriginatingfromthefaultyelectricalinstallation
AT djurdjevicbranko inappropriateshockdeliveryasaresultofelectromagneticinterferenceoriginatingfromthefaultyelectricalinstallation
AT zugicvasko inappropriateshockdeliveryasaresultofelectromagneticinterferenceoriginatingfromthefaultyelectricalinstallation
AT nikolicaleksandra inappropriateshockdeliveryasaresultofelectromagneticinterferenceoriginatingfromthefaultyelectricalinstallation