Cargando…
Inappropriate shock delivery as a result of electromagnetic interference originating from the faulty electrical installation
We present a case report of a 74‐year‐old male patient with an implantable cardioverter defibrillator who suffered an inappropriate defibrillation shock while bathing in the tub. Insight in the ICD stored electrogram episodes revealed electromagnetic interferences, with a typical 50 Hz electrical ar...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9484028/ https://www.ncbi.nlm.nih.gov/pubmed/35467789 http://dx.doi.org/10.1111/anec.12952 |
_version_ | 1784791798852354048 |
---|---|
author | Babic, Milos D Tomovic, Milosav Milosevic, Maja Djurdjevic, Branko Zugic, Vasko Nikolic, Aleksandra |
author_facet | Babic, Milos D Tomovic, Milosav Milosevic, Maja Djurdjevic, Branko Zugic, Vasko Nikolic, Aleksandra |
author_sort | Babic, Milos D |
collection | PubMed |
description | We present a case report of a 74‐year‐old male patient with an implantable cardioverter defibrillator who suffered an inappropriate defibrillation shock while bathing in the tub. Insight in the ICD stored electrogram episodes revealed electromagnetic interferences, with a typical 50 Hz electrical artifact mimicking fast ventricular tachycardia as a device misinterpreted. After this event, the maintenance workers investigated the electrical installation in the bathroom and revealed that there was voltage leaking between electrical installation and metal pipes. After the repair was completed without any additional programming, the patient has had no subsequent shocks. |
format | Online Article Text |
id | pubmed-9484028 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | John Wiley and Sons Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-94840282022-09-29 Inappropriate shock delivery as a result of electromagnetic interference originating from the faulty electrical installation Babic, Milos D Tomovic, Milosav Milosevic, Maja Djurdjevic, Branko Zugic, Vasko Nikolic, Aleksandra Ann Noninvasive Electrocardiol Case Reports We present a case report of a 74‐year‐old male patient with an implantable cardioverter defibrillator who suffered an inappropriate defibrillation shock while bathing in the tub. Insight in the ICD stored electrogram episodes revealed electromagnetic interferences, with a typical 50 Hz electrical artifact mimicking fast ventricular tachycardia as a device misinterpreted. After this event, the maintenance workers investigated the electrical installation in the bathroom and revealed that there was voltage leaking between electrical installation and metal pipes. After the repair was completed without any additional programming, the patient has had no subsequent shocks. John Wiley and Sons Inc. 2022-04-25 /pmc/articles/PMC9484028/ /pubmed/35467789 http://dx.doi.org/10.1111/anec.12952 Text en © 2022 The Authors. Annals of Noninvasive Electrocardiology published by Wiley Periodicals LLC. https://creativecommons.org/licenses/by-nc-nd/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by-nc-nd/4.0/ (https://creativecommons.org/licenses/by-nc-nd/4.0/) License, which permits use and distribution in any medium, provided the original work is properly cited, the use is non‐commercial and no modifications or adaptations are made. |
spellingShingle | Case Reports Babic, Milos D Tomovic, Milosav Milosevic, Maja Djurdjevic, Branko Zugic, Vasko Nikolic, Aleksandra Inappropriate shock delivery as a result of electromagnetic interference originating from the faulty electrical installation |
title | Inappropriate shock delivery as a result of electromagnetic interference originating from the faulty electrical installation |
title_full | Inappropriate shock delivery as a result of electromagnetic interference originating from the faulty electrical installation |
title_fullStr | Inappropriate shock delivery as a result of electromagnetic interference originating from the faulty electrical installation |
title_full_unstemmed | Inappropriate shock delivery as a result of electromagnetic interference originating from the faulty electrical installation |
title_short | Inappropriate shock delivery as a result of electromagnetic interference originating from the faulty electrical installation |
title_sort | inappropriate shock delivery as a result of electromagnetic interference originating from the faulty electrical installation |
topic | Case Reports |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9484028/ https://www.ncbi.nlm.nih.gov/pubmed/35467789 http://dx.doi.org/10.1111/anec.12952 |
work_keys_str_mv | AT babicmilosd inappropriateshockdeliveryasaresultofelectromagneticinterferenceoriginatingfromthefaultyelectricalinstallation AT tomovicmilosav inappropriateshockdeliveryasaresultofelectromagneticinterferenceoriginatingfromthefaultyelectricalinstallation AT milosevicmaja inappropriateshockdeliveryasaresultofelectromagneticinterferenceoriginatingfromthefaultyelectricalinstallation AT djurdjevicbranko inappropriateshockdeliveryasaresultofelectromagneticinterferenceoriginatingfromthefaultyelectricalinstallation AT zugicvasko inappropriateshockdeliveryasaresultofelectromagneticinterferenceoriginatingfromthefaultyelectricalinstallation AT nikolicaleksandra inappropriateshockdeliveryasaresultofelectromagneticinterferenceoriginatingfromthefaultyelectricalinstallation |