Cargando…

Inappropriate shock delivery as a result of electromagnetic interference originating from the faulty electrical installation

We present a case report of a 74‐year‐old male patient with an implantable cardioverter defibrillator who suffered an inappropriate defibrillation shock while bathing in the tub. Insight in the ICD stored electrogram episodes revealed electromagnetic interferences, with a typical 50 Hz electrical ar...

Descripción completa

Detalles Bibliográficos
Autores principales: Babic, Milos D, Tomovic, Milosav, Milosevic, Maja, Djurdjevic, Branko, Zugic, Vasko, Nikolic, Aleksandra
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9484028/
https://www.ncbi.nlm.nih.gov/pubmed/35467789
http://dx.doi.org/10.1111/anec.12952