Computational wear of knee implant polyethylene insert surface under continuous dynamic loading and posterior tibial slope variation based on cadaver experiments with comparative verification
BACKGROUND: The effect of posterior tibial slope on the maximum contact pressure and wear volume of polyethylene (PE) insert were not given special attention. The effects of flexion angle, Anterior-Posterior (AP) Translation, and Tibial slope on the max contact pressure and wear of PE insert of TKR...
Autor principal: | |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9484235/ https://www.ncbi.nlm.nih.gov/pubmed/36123647 http://dx.doi.org/10.1186/s12891-022-05828-2 |
_version_ | 1784791838292443136 |
---|---|
author | Ozer, Alaettin |
author_facet | Ozer, Alaettin |
author_sort | Ozer, Alaettin |
collection | PubMed |
description | BACKGROUND: The effect of posterior tibial slope on the maximum contact pressure and wear volume of polyethylene (PE) insert were not given special attention. The effects of flexion angle, Anterior-Posterior (AP) Translation, and Tibial slope on the max contact pressure and wear of PE insert of TKR were investigated under loadings which were obtained in cadaver experiments by using Archard’s wear law. This study uses not only loads obtained from cadaver experiments but also dynamic flexion starting from 0 to 90 degrees. METHOD: Wear on knee implant PE insert was investigated using a 2.5 size 3 dimensional (3D) cruciate sacrificing total knee replacement model and Finite Element Method (FEM) under loadings and AP Translation data ranging from 0 to 90 flexion angles validated by cadaver experiments. Two types of analyses were done to measure the wear effect on knee implant PE insert. The first set of analyses included the flexion angles dynamically changing with the knee rotating from 0 to 90 angles according to the femur axis and the transient analyses for loadings changing with a certain angle and duration. RESULTS: It is seen that the contact pressure on the PE insert decreases as the cycle increases for both Flexion and Flexion+AP Translation. It is clear that as the cycle increases, the wear obtained for both cases increases. The loadings acting on the PE insert cannot create sufficient pressure due to the AP Translation effect at low speeds and have an effect to reduce the wear, while the effect increases with the wear as the cycle increases, and the AP Translation now contributes to the wear at high speeds. It is seen that as the posterior tibial slope angle increases, the maximum contact pressure values slightly decrease for the same cycle. CONCLUSIONS: This study indicated that AP Translation, which changes direction during flexion, had a significant effect on both contact pressure and wear. Unlike previous similar studies, it was seen that the amount of wear continues to increase as the cycle increases. This situation strengthens the argument that loading and AP Translation values that change with flexion shape the wear effects on PE Insert. |
format | Online Article Text |
id | pubmed-9484235 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-94842352022-09-20 Computational wear of knee implant polyethylene insert surface under continuous dynamic loading and posterior tibial slope variation based on cadaver experiments with comparative verification Ozer, Alaettin BMC Musculoskelet Disord Research BACKGROUND: The effect of posterior tibial slope on the maximum contact pressure and wear volume of polyethylene (PE) insert were not given special attention. The effects of flexion angle, Anterior-Posterior (AP) Translation, and Tibial slope on the max contact pressure and wear of PE insert of TKR were investigated under loadings which were obtained in cadaver experiments by using Archard’s wear law. This study uses not only loads obtained from cadaver experiments but also dynamic flexion starting from 0 to 90 degrees. METHOD: Wear on knee implant PE insert was investigated using a 2.5 size 3 dimensional (3D) cruciate sacrificing total knee replacement model and Finite Element Method (FEM) under loadings and AP Translation data ranging from 0 to 90 flexion angles validated by cadaver experiments. Two types of analyses were done to measure the wear effect on knee implant PE insert. The first set of analyses included the flexion angles dynamically changing with the knee rotating from 0 to 90 angles according to the femur axis and the transient analyses for loadings changing with a certain angle and duration. RESULTS: It is seen that the contact pressure on the PE insert decreases as the cycle increases for both Flexion and Flexion+AP Translation. It is clear that as the cycle increases, the wear obtained for both cases increases. The loadings acting on the PE insert cannot create sufficient pressure due to the AP Translation effect at low speeds and have an effect to reduce the wear, while the effect increases with the wear as the cycle increases, and the AP Translation now contributes to the wear at high speeds. It is seen that as the posterior tibial slope angle increases, the maximum contact pressure values slightly decrease for the same cycle. CONCLUSIONS: This study indicated that AP Translation, which changes direction during flexion, had a significant effect on both contact pressure and wear. Unlike previous similar studies, it was seen that the amount of wear continues to increase as the cycle increases. This situation strengthens the argument that loading and AP Translation values that change with flexion shape the wear effects on PE Insert. BioMed Central 2022-09-19 /pmc/articles/PMC9484235/ /pubmed/36123647 http://dx.doi.org/10.1186/s12891-022-05828-2 Text en © The Author(s) 2022 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) . The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/ (https://creativecommons.org/publicdomain/zero/1.0/) ) applies to the data made available in this article, unless otherwise stated in a credit line to the data. |
spellingShingle | Research Ozer, Alaettin Computational wear of knee implant polyethylene insert surface under continuous dynamic loading and posterior tibial slope variation based on cadaver experiments with comparative verification |
title | Computational wear of knee implant polyethylene insert surface under continuous dynamic loading and posterior tibial slope variation based on cadaver experiments with comparative verification |
title_full | Computational wear of knee implant polyethylene insert surface under continuous dynamic loading and posterior tibial slope variation based on cadaver experiments with comparative verification |
title_fullStr | Computational wear of knee implant polyethylene insert surface under continuous dynamic loading and posterior tibial slope variation based on cadaver experiments with comparative verification |
title_full_unstemmed | Computational wear of knee implant polyethylene insert surface under continuous dynamic loading and posterior tibial slope variation based on cadaver experiments with comparative verification |
title_short | Computational wear of knee implant polyethylene insert surface under continuous dynamic loading and posterior tibial slope variation based on cadaver experiments with comparative verification |
title_sort | computational wear of knee implant polyethylene insert surface under continuous dynamic loading and posterior tibial slope variation based on cadaver experiments with comparative verification |
topic | Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9484235/ https://www.ncbi.nlm.nih.gov/pubmed/36123647 http://dx.doi.org/10.1186/s12891-022-05828-2 |
work_keys_str_mv | AT ozeralaettin computationalwearofkneeimplantpolyethyleneinsertsurfaceundercontinuousdynamicloadingandposteriortibialslopevariationbasedoncadaverexperimentswithcomparativeverification |