Concordance of three point of care testing devices with clinical chemistry laboratory standard assays and patient-reported outcomes of blood sampling methods

BACKGROUND: Point of care testing (POCT) devices have been developed to facilitate immediate results with the potential to aid screening for new disease and enable patients to self-monitor their disease. Non-communicable diseases (NCDs) are the major cause of mortality globally and are increasing in...

Descripción completa

Detalles Bibliográficos
Autores principales: Yonel, Z., Kuningas, K., Sharma, P., Dutton, M., Jalal, Z., Cockwell, P., Webber, J., Narendran, P., Dietrich, T., Chapple, I. L. C.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9493167/
https://www.ncbi.nlm.nih.gov/pubmed/36138408
http://dx.doi.org/10.1186/s12911-022-01999-z
_version_ 1784793642008838144
author Yonel, Z.
Kuningas, K.
Sharma, P.
Dutton, M.
Jalal, Z.
Cockwell, P.
Webber, J.
Narendran, P.
Dietrich, T.
Chapple, I. L. C.
author_facet Yonel, Z.
Kuningas, K.
Sharma, P.
Dutton, M.
Jalal, Z.
Cockwell, P.
Webber, J.
Narendran, P.
Dietrich, T.
Chapple, I. L. C.
author_sort Yonel, Z.
collection PubMed
description BACKGROUND: Point of care testing (POCT) devices have been developed to facilitate immediate results with the potential to aid screening for new disease and enable patients to self-monitor their disease. Non-communicable diseases (NCDs) are the major cause of mortality globally and are increasing in prevalence as the population ages. Allied health care professionals (AHPs) are skilled in undertaking risk assessment and delivering preventative advice, providing opportunities to access large proportions of the population who may not visit their doctor, within non-traditional community settings. There is evidence of high levels of support from public, patients and health professionals for engaging AHPs in risk-targeted early case detection of certain NCDs. Thus, POCT devices offer a potential alternative to traditional venous blood collection, as novel care pathways for increasing early case detection and access to preventative care. The objectives of this study were to: (i) determine the concordance of the specific POCT devices with laboratory-based standard assays employed within clinical biochemistry laboratories. (ii) compare the sampling experience of both methods via patient-reported experiences. METHODS: A prospective, two-centre study was undertaken involving 158 participants who provided informed consent. Venous blood was collected for traditional assays of HbA1c, creatinine/ estimated Glomerular-Filtration-Rate (eGFR) and vitamin-D. Capillary blood was collected by finger prick test and also assayed for the same biochemical indices (Nova StatSensor (creatinine/eGFR); Siemens DCA-Vantage (HbA1C); CityAssays (vitamin-D)). All users were provided with device training. Participants reported any discomfort experienced by each simultaneously applied method (randomised in order) via a 100 mm Visual-Analogue-Scale. RESULTS: Results for each POCT device and the laboratory standard were analysed by Bland-Altman plots to determine assay concordance. POCT devices demonstrated good concordance with laboratory testing, with at least 95% of all samples being within two standard deviations, for each of the devices tested. The majority of participants reported less discomfort with POCT than venepuncture, with the average reported discomfort being 17/100 mm less for POCT compared to venous blood sample collection on the visual analogue scale. CONCLUSIONS: The POCT devices demonstrated acceptable concordance with laboratory-based assays, and patients reported lower levels of discomfort compared to traditional means of blood collection. This study demonstrates the potential of using these devices as acceptable methods for opportunistic testing of “at-risk” individuals within non-traditional community care settings.
format Online
Article
Text
id pubmed-9493167
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-94931672022-09-22 Concordance of three point of care testing devices with clinical chemistry laboratory standard assays and patient-reported outcomes of blood sampling methods Yonel, Z. Kuningas, K. Sharma, P. Dutton, M. Jalal, Z. Cockwell, P. Webber, J. Narendran, P. Dietrich, T. Chapple, I. L. C. BMC Med Inform Decis Mak Research BACKGROUND: Point of care testing (POCT) devices have been developed to facilitate immediate results with the potential to aid screening for new disease and enable patients to self-monitor their disease. Non-communicable diseases (NCDs) are the major cause of mortality globally and are increasing in prevalence as the population ages. Allied health care professionals (AHPs) are skilled in undertaking risk assessment and delivering preventative advice, providing opportunities to access large proportions of the population who may not visit their doctor, within non-traditional community settings. There is evidence of high levels of support from public, patients and health professionals for engaging AHPs in risk-targeted early case detection of certain NCDs. Thus, POCT devices offer a potential alternative to traditional venous blood collection, as novel care pathways for increasing early case detection and access to preventative care. The objectives of this study were to: (i) determine the concordance of the specific POCT devices with laboratory-based standard assays employed within clinical biochemistry laboratories. (ii) compare the sampling experience of both methods via patient-reported experiences. METHODS: A prospective, two-centre study was undertaken involving 158 participants who provided informed consent. Venous blood was collected for traditional assays of HbA1c, creatinine/ estimated Glomerular-Filtration-Rate (eGFR) and vitamin-D. Capillary blood was collected by finger prick test and also assayed for the same biochemical indices (Nova StatSensor (creatinine/eGFR); Siemens DCA-Vantage (HbA1C); CityAssays (vitamin-D)). All users were provided with device training. Participants reported any discomfort experienced by each simultaneously applied method (randomised in order) via a 100 mm Visual-Analogue-Scale. RESULTS: Results for each POCT device and the laboratory standard were analysed by Bland-Altman plots to determine assay concordance. POCT devices demonstrated good concordance with laboratory testing, with at least 95% of all samples being within two standard deviations, for each of the devices tested. The majority of participants reported less discomfort with POCT than venepuncture, with the average reported discomfort being 17/100 mm less for POCT compared to venous blood sample collection on the visual analogue scale. CONCLUSIONS: The POCT devices demonstrated acceptable concordance with laboratory-based assays, and patients reported lower levels of discomfort compared to traditional means of blood collection. This study demonstrates the potential of using these devices as acceptable methods for opportunistic testing of “at-risk” individuals within non-traditional community care settings. BioMed Central 2022-09-22 /pmc/articles/PMC9493167/ /pubmed/36138408 http://dx.doi.org/10.1186/s12911-022-01999-z Text en © The Author(s) 2022 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) . The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/ (https://creativecommons.org/publicdomain/zero/1.0/) ) applies to the data made available in this article, unless otherwise stated in a credit line to the data.
spellingShingle Research
Yonel, Z.
Kuningas, K.
Sharma, P.
Dutton, M.
Jalal, Z.
Cockwell, P.
Webber, J.
Narendran, P.
Dietrich, T.
Chapple, I. L. C.
Concordance of three point of care testing devices with clinical chemistry laboratory standard assays and patient-reported outcomes of blood sampling methods
title Concordance of three point of care testing devices with clinical chemistry laboratory standard assays and patient-reported outcomes of blood sampling methods
title_full Concordance of three point of care testing devices with clinical chemistry laboratory standard assays and patient-reported outcomes of blood sampling methods
title_fullStr Concordance of three point of care testing devices with clinical chemistry laboratory standard assays and patient-reported outcomes of blood sampling methods
title_full_unstemmed Concordance of three point of care testing devices with clinical chemistry laboratory standard assays and patient-reported outcomes of blood sampling methods
title_short Concordance of three point of care testing devices with clinical chemistry laboratory standard assays and patient-reported outcomes of blood sampling methods
title_sort concordance of three point of care testing devices with clinical chemistry laboratory standard assays and patient-reported outcomes of blood sampling methods
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9493167/
https://www.ncbi.nlm.nih.gov/pubmed/36138408
http://dx.doi.org/10.1186/s12911-022-01999-z
work_keys_str_mv AT yonelz concordanceofthreepointofcaretestingdeviceswithclinicalchemistrylaboratorystandardassaysandpatientreportedoutcomesofbloodsamplingmethods
AT kuningask concordanceofthreepointofcaretestingdeviceswithclinicalchemistrylaboratorystandardassaysandpatientreportedoutcomesofbloodsamplingmethods
AT sharmap concordanceofthreepointofcaretestingdeviceswithclinicalchemistrylaboratorystandardassaysandpatientreportedoutcomesofbloodsamplingmethods
AT duttonm concordanceofthreepointofcaretestingdeviceswithclinicalchemistrylaboratorystandardassaysandpatientreportedoutcomesofbloodsamplingmethods
AT jalalz concordanceofthreepointofcaretestingdeviceswithclinicalchemistrylaboratorystandardassaysandpatientreportedoutcomesofbloodsamplingmethods
AT cockwellp concordanceofthreepointofcaretestingdeviceswithclinicalchemistrylaboratorystandardassaysandpatientreportedoutcomesofbloodsamplingmethods
AT webberj concordanceofthreepointofcaretestingdeviceswithclinicalchemistrylaboratorystandardassaysandpatientreportedoutcomesofbloodsamplingmethods
AT narendranp concordanceofthreepointofcaretestingdeviceswithclinicalchemistrylaboratorystandardassaysandpatientreportedoutcomesofbloodsamplingmethods
AT dietricht concordanceofthreepointofcaretestingdeviceswithclinicalchemistrylaboratorystandardassaysandpatientreportedoutcomesofbloodsamplingmethods
AT chappleilc concordanceofthreepointofcaretestingdeviceswithclinicalchemistrylaboratorystandardassaysandpatientreportedoutcomesofbloodsamplingmethods