Feasibility of Blood Pressure Measurement With a Wearable (Watch‐Type) Monitor During Impending Syncopal Episodes
BACKGROUND: We assessed the reliability and feasibility of blood pressure (BP) measurements by means of a new wearable watch‐type BP monitor (HeartGuide) in detecting episodes of hypotensive (pre)syncope induced by tilt table test. METHODS AND RESULTS: An intrapatient comparison between systolic BP...
Autores principales: | , , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9496316/ https://www.ncbi.nlm.nih.gov/pubmed/35929469 http://dx.doi.org/10.1161/JAHA.122.026420 |
_version_ | 1784794238839422976 |
---|---|
author | Groppelli, Antonella Rafanelli, Martina Testa, Giuseppe Dario Agusto, Samuele Rivasi, Giulia Ungar, Andrea Carbone, Erika Soranna, Davide Zambon, Antonella Brignole, Michele Parati, Gianfranco |
author_facet | Groppelli, Antonella Rafanelli, Martina Testa, Giuseppe Dario Agusto, Samuele Rivasi, Giulia Ungar, Andrea Carbone, Erika Soranna, Davide Zambon, Antonella Brignole, Michele Parati, Gianfranco |
author_sort | Groppelli, Antonella |
collection | PubMed |
description | BACKGROUND: We assessed the reliability and feasibility of blood pressure (BP) measurements by means of a new wearable watch‐type BP monitor (HeartGuide) in detecting episodes of hypotensive (pre)syncope induced by tilt table test. METHODS AND RESULTS: An intrapatient comparison between systolic BP (SBP) measured by means of the HeartGuide device and noninvasive finger beat‐to‐beat BP monitoring was undertaken both at baseline in supine position and repeatedly during tilt table test in patients evaluated for reflex syncope. Intrapatient fall of systolic BP from baseline was measured. Eighty‐one patients (mean age, 61±19 years; 46 women) were included. Overall, HeartGuide was able to yield BP values at the time of BP nadir in 58 (72%) patients (average HeartGuide SBP 102±18 mm Hg, versus finger SBP 101±19 mm Hg). Compared with baseline, the maximum SBP decrease was on average −28.5±27.8 and −30.3±33.9 mm Hg respectively (Lin's concordance correlation coefficient=0.78, r=0.79, P=0.001). In the subgroup of 38 patients with tilt table test induced (pre)syncope, the average HeartGuide SBP during symptoms was 97±16 mm Hg, and the finger SBP was 94±18 mm Hg. Compared with baseline, the maximum SBP decrease was on average −35.2±29.3 and −43.3±31.8 mm Hg, respectively (Lin's concordance correlation coefficient=0.83, r=0.87, P=0.001). CONCLUSIONS: Our data indicate that the HeartGuide BP monitor can detect low BP during presyncope and that its measure of SBP change is consistent with that simultaneously obtained through continuous BP monitoring, despite some intrapatient variability. Thus, this device might be useful in determining the hypotensive nature of spontaneous (pre)syncopal symptoms, a possibility that should be verified by field studies. |
format | Online Article Text |
id | pubmed-9496316 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | John Wiley and Sons Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-94963162022-09-30 Feasibility of Blood Pressure Measurement With a Wearable (Watch‐Type) Monitor During Impending Syncopal Episodes Groppelli, Antonella Rafanelli, Martina Testa, Giuseppe Dario Agusto, Samuele Rivasi, Giulia Ungar, Andrea Carbone, Erika Soranna, Davide Zambon, Antonella Brignole, Michele Parati, Gianfranco J Am Heart Assoc Original Research BACKGROUND: We assessed the reliability and feasibility of blood pressure (BP) measurements by means of a new wearable watch‐type BP monitor (HeartGuide) in detecting episodes of hypotensive (pre)syncope induced by tilt table test. METHODS AND RESULTS: An intrapatient comparison between systolic BP (SBP) measured by means of the HeartGuide device and noninvasive finger beat‐to‐beat BP monitoring was undertaken both at baseline in supine position and repeatedly during tilt table test in patients evaluated for reflex syncope. Intrapatient fall of systolic BP from baseline was measured. Eighty‐one patients (mean age, 61±19 years; 46 women) were included. Overall, HeartGuide was able to yield BP values at the time of BP nadir in 58 (72%) patients (average HeartGuide SBP 102±18 mm Hg, versus finger SBP 101±19 mm Hg). Compared with baseline, the maximum SBP decrease was on average −28.5±27.8 and −30.3±33.9 mm Hg respectively (Lin's concordance correlation coefficient=0.78, r=0.79, P=0.001). In the subgroup of 38 patients with tilt table test induced (pre)syncope, the average HeartGuide SBP during symptoms was 97±16 mm Hg, and the finger SBP was 94±18 mm Hg. Compared with baseline, the maximum SBP decrease was on average −35.2±29.3 and −43.3±31.8 mm Hg, respectively (Lin's concordance correlation coefficient=0.83, r=0.87, P=0.001). CONCLUSIONS: Our data indicate that the HeartGuide BP monitor can detect low BP during presyncope and that its measure of SBP change is consistent with that simultaneously obtained through continuous BP monitoring, despite some intrapatient variability. Thus, this device might be useful in determining the hypotensive nature of spontaneous (pre)syncopal symptoms, a possibility that should be verified by field studies. John Wiley and Sons Inc. 2022-08-05 /pmc/articles/PMC9496316/ /pubmed/35929469 http://dx.doi.org/10.1161/JAHA.122.026420 Text en © 2022 The Authors. Published on behalf of the American Heart Association, Inc., by Wiley. https://creativecommons.org/licenses/by-nc-nd/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by-nc-nd/4.0/ (https://creativecommons.org/licenses/by-nc-nd/4.0/) License, which permits use and distribution in any medium, provided the original work is properly cited, the use is non‐commercial and no modifications or adaptations are made. |
spellingShingle | Original Research Groppelli, Antonella Rafanelli, Martina Testa, Giuseppe Dario Agusto, Samuele Rivasi, Giulia Ungar, Andrea Carbone, Erika Soranna, Davide Zambon, Antonella Brignole, Michele Parati, Gianfranco Feasibility of Blood Pressure Measurement With a Wearable (Watch‐Type) Monitor During Impending Syncopal Episodes |
title | Feasibility of Blood Pressure Measurement With a Wearable (Watch‐Type) Monitor During Impending Syncopal Episodes |
title_full | Feasibility of Blood Pressure Measurement With a Wearable (Watch‐Type) Monitor During Impending Syncopal Episodes |
title_fullStr | Feasibility of Blood Pressure Measurement With a Wearable (Watch‐Type) Monitor During Impending Syncopal Episodes |
title_full_unstemmed | Feasibility of Blood Pressure Measurement With a Wearable (Watch‐Type) Monitor During Impending Syncopal Episodes |
title_short | Feasibility of Blood Pressure Measurement With a Wearable (Watch‐Type) Monitor During Impending Syncopal Episodes |
title_sort | feasibility of blood pressure measurement with a wearable (watch‐type) monitor during impending syncopal episodes |
topic | Original Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9496316/ https://www.ncbi.nlm.nih.gov/pubmed/35929469 http://dx.doi.org/10.1161/JAHA.122.026420 |
work_keys_str_mv | AT groppelliantonella feasibilityofbloodpressuremeasurementwithawearablewatchtypemonitorduringimpendingsyncopalepisodes AT rafanellimartina feasibilityofbloodpressuremeasurementwithawearablewatchtypemonitorduringimpendingsyncopalepisodes AT testagiuseppedario feasibilityofbloodpressuremeasurementwithawearablewatchtypemonitorduringimpendingsyncopalepisodes AT agustosamuele feasibilityofbloodpressuremeasurementwithawearablewatchtypemonitorduringimpendingsyncopalepisodes AT rivasigiulia feasibilityofbloodpressuremeasurementwithawearablewatchtypemonitorduringimpendingsyncopalepisodes AT ungarandrea feasibilityofbloodpressuremeasurementwithawearablewatchtypemonitorduringimpendingsyncopalepisodes AT carboneerika feasibilityofbloodpressuremeasurementwithawearablewatchtypemonitorduringimpendingsyncopalepisodes AT sorannadavide feasibilityofbloodpressuremeasurementwithawearablewatchtypemonitorduringimpendingsyncopalepisodes AT zambonantonella feasibilityofbloodpressuremeasurementwithawearablewatchtypemonitorduringimpendingsyncopalepisodes AT brignolemichele feasibilityofbloodpressuremeasurementwithawearablewatchtypemonitorduringimpendingsyncopalepisodes AT paratigianfranco feasibilityofbloodpressuremeasurementwithawearablewatchtypemonitorduringimpendingsyncopalepisodes |