Cargando…

Evaluation of Saliva as a Matrix for RT-PCR Analysis and Two Rapid Antigen Tests for the Detection of SARS-CoV-2

The use of saliva for the detection of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) sparks debate due to presumed lower sensitivity and lack of standardization. Our aim was to evaluate the performance characteristics of (i) saliva collected by the ORAcollect(TM) device as a matrix fo...

Descripción completa

Detalles Bibliográficos
Autores principales: De Meyer, Julie, Goris, Hanne, Mortelé, Olivier, Spiessens, An, Hans, Guy, Jansens, Hilde, Goossens, Herman, Matheeussen, Veerle, Vandamme, Sarah
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9502549/
https://www.ncbi.nlm.nih.gov/pubmed/36146737
http://dx.doi.org/10.3390/v14091931
Descripción
Sumario:The use of saliva for the detection of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) sparks debate due to presumed lower sensitivity and lack of standardization. Our aim was to evaluate the performance characteristics of (i) saliva collected by the ORAcollect(TM) device as a matrix for SARS-CoV-2 reverse-transcriptase polymerase chain reaction (RT-PCR), and (ii) 2 saliva rapid antigen tests (AgRDT). From 342 ambulatory individuals, both a nasopharyngeal swab and saliva sample via ORAcollect(TM) were obtained for a SARS-CoV-2 RT-PCR test. Furthermore, 54 and 123 additionally performed the V-Chek(TM) or Whistling(TM) saliva AgRDT. In total, 35% of individuals screened positive for SARS-CoV-2 via nasopharyngeal swab. Saliva, as a matrix for the RT-PCR, had a specificity of 96.5% and a negative predictive value (NPV) of 91.3%. Interestingly, 6 out of 8 patients thought to be false positive in saliva re-tested positive by nasopharyngeal sampling after 2 to 9 days. Both V-Chek(TM) and Whistling(TM) AgRDT had a lack of sensitivity, resulting in an NPV of 66.9 and 67.3%, respectively. Saliva proved to be a sensitive and specific matrix for SARS-CoV-2 detection by the RT-PCR. In this setting, saliva might have an earlier window of detection than the nasopharyngeal swab. By contrast, both AgRDT showed an unacceptably low sensitivity and NPV.