Gamma ray detection performance of newly developed MAPD-3NM-II photosensor with LaBr(3)(Ce) crystal

This paper presents the gamma-ray detection performance of the newly developed MAPD-3NM-II type SiPM sensor array (4 [Formula: see text] 4) with [Formula: see text] (Ce) scintillator. The gamma-ray spectra of various sources have been measured in the energy range from 26 keV up to 1332 keV. The newl...

Descripción completa

Detalles Bibliográficos
Autores principales: Holik, M., Ahmadov, F., Sadigov, A., Ahmadov, G., Berikov, D., Mamedov, F., Naghiyev, J., Stekl, I., Sadygov, Z., Mammadli, A., Garibli, A., Urban, O., Zich, J.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Nature Publishing Group UK 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9508114/
https://www.ncbi.nlm.nih.gov/pubmed/36151262
http://dx.doi.org/10.1038/s41598-022-20006-z
Descripción
Sumario:This paper presents the gamma-ray detection performance of the newly developed MAPD-3NM-II type SiPM sensor array (4 [Formula: see text] 4) with [Formula: see text] (Ce) scintillator. The gamma-ray spectra of various sources have been measured in the energy range from 26 keV up to 1332 keV. The newly developed array based on MAPD-3NM-II sensors proved [Formula: see text]  22% enhancement in energy resolution in comparison to the former MAPD-3NM-I based array. The energy resolution of 662 keV gamma-rays measured by MAPD-3NM-II was 3.3% while clearly surpassing 4.25% resolution of MAPD-3NM-I predecessor. The enhancement is related to the high PDE of the new MAPD-3NM-II. Obtained results show that the new MAPD-3NM-II demonstrated good energy resolution and linearity in the studied energy region. The energy resolution of the new detector developed based on MAPD-3NM-II was better than all previous products of MAPD.