Evaluation of the performance of two new generation pulse oximeters in cats at different probe positions and under the influence of vasoconstriction
OBJECTIVES: The aim of this study was to compare the failure rate of two new generation pulse oximeters at different probe positions, and with and without vasoconstriction, in anaesthetised cats. METHODS: This prospective clinical study included 103 cats in which the new generation pulse oximeters,...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
SAGE Publications
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9510937/ https://www.ncbi.nlm.nih.gov/pubmed/34904479 http://dx.doi.org/10.1177/1098612X211063768 |
_version_ | 1784797553208852480 |
---|---|
author | Dörfelt, René Diels, Julia Hartmann, Katrin |
author_facet | Dörfelt, René Diels, Julia Hartmann, Katrin |
author_sort | Dörfelt, René |
collection | PubMed |
description | OBJECTIVES: The aim of this study was to compare the failure rate of two new generation pulse oximeters at different probe positions, and with and without vasoconstriction, in anaesthetised cats. METHODS: This prospective clinical study included 103 cats in which the new generation pulse oximeters, the Rad-5 (Masimo) and EDAN H100N (EDAN), were evaluated. Premedication consisted of the vasoconstrictive drug combination butorphanol (0.2 mg/kg IV) and dexmedetomidine (5 µg/kg IV), or butorphanol only (0.2 mg/kg IV). Pulse oximeter failure rate at the tongue was compared between both groups. Pulse oximeter failure rate was also analysed at the alternative probe positions of the lip, pinna, knee fold and toe in the butorphanol group. Student’s t-test, Wilcoxon matched pairs signed rank test, Mann−Whitney U-test, Friedman test and χ(2) test were performed. A P value <0.05 was considered to be statistically significant. RESULTS: Overall failure to achieve an adequate signal was 37.6% with the Masimo and 48.0% with the EDAN pulse oximeter (P <0.0001). At the standard probe position on the tongue, the Masimo failed in 4.5%, while the EDAN failed in 35.3% (P <0.0001). Vasoactive premedication increased the failure rate for the Masimo from 3.8% to 5.2% (P = 0.3414) and for the EDAN from 22.4% to 49.0% (P <0.0001). At the alternative probe positions of the lip and knee fold, failure rates for the Masimo were lower (39.7% and 81.4%) than with the EDAN (52.6% and 94.4%; P = 0.0231 and P = 0.0005, respectively), while the Masimo failed more often at the pinna (63.5%) than the EDAN (47.4%; P = 0.0044). At the alternative probe position of the toe, the failure rate for the Masimo (32.7%) was not different from the EDAN (38.5%; P = 0.7547). CONCLUSIONS AND RELEVANCE: The Masimo pulse oximeter had lower signal failure rates at the standard probe position on the tongue and at 2/4 alternative probe positions. The standard probe position on the tongue had the lowest failure rate for both devices. Dexmedetomidine-induced vasoconstriction increased the failure rate for the EDAN but not for the Masimo pulse oximeter. |
format | Online Article Text |
id | pubmed-9510937 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | SAGE Publications |
record_format | MEDLINE/PubMed |
spelling | pubmed-95109372022-09-27 Evaluation of the performance of two new generation pulse oximeters in cats at different probe positions and under the influence of vasoconstriction Dörfelt, René Diels, Julia Hartmann, Katrin J Feline Med Surg Original Articles OBJECTIVES: The aim of this study was to compare the failure rate of two new generation pulse oximeters at different probe positions, and with and without vasoconstriction, in anaesthetised cats. METHODS: This prospective clinical study included 103 cats in which the new generation pulse oximeters, the Rad-5 (Masimo) and EDAN H100N (EDAN), were evaluated. Premedication consisted of the vasoconstrictive drug combination butorphanol (0.2 mg/kg IV) and dexmedetomidine (5 µg/kg IV), or butorphanol only (0.2 mg/kg IV). Pulse oximeter failure rate at the tongue was compared between both groups. Pulse oximeter failure rate was also analysed at the alternative probe positions of the lip, pinna, knee fold and toe in the butorphanol group. Student’s t-test, Wilcoxon matched pairs signed rank test, Mann−Whitney U-test, Friedman test and χ(2) test were performed. A P value <0.05 was considered to be statistically significant. RESULTS: Overall failure to achieve an adequate signal was 37.6% with the Masimo and 48.0% with the EDAN pulse oximeter (P <0.0001). At the standard probe position on the tongue, the Masimo failed in 4.5%, while the EDAN failed in 35.3% (P <0.0001). Vasoactive premedication increased the failure rate for the Masimo from 3.8% to 5.2% (P = 0.3414) and for the EDAN from 22.4% to 49.0% (P <0.0001). At the alternative probe positions of the lip and knee fold, failure rates for the Masimo were lower (39.7% and 81.4%) than with the EDAN (52.6% and 94.4%; P = 0.0231 and P = 0.0005, respectively), while the Masimo failed more often at the pinna (63.5%) than the EDAN (47.4%; P = 0.0044). At the alternative probe position of the toe, the failure rate for the Masimo (32.7%) was not different from the EDAN (38.5%; P = 0.7547). CONCLUSIONS AND RELEVANCE: The Masimo pulse oximeter had lower signal failure rates at the standard probe position on the tongue and at 2/4 alternative probe positions. The standard probe position on the tongue had the lowest failure rate for both devices. Dexmedetomidine-induced vasoconstriction increased the failure rate for the EDAN but not for the Masimo pulse oximeter. SAGE Publications 2021-12-14 2022-10 /pmc/articles/PMC9510937/ /pubmed/34904479 http://dx.doi.org/10.1177/1098612X211063768 Text en © The Author(s) 2021 https://creativecommons.org/licenses/by-nc/4.0/This article is distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 License (https://creativecommons.org/licenses/by-nc/4.0/) which permits non-commercial use, reproduction and distribution of the work without further permission provided the original work is attributed as specified on the SAGE and Open Access pages (https://us.sagepub.com/en-us/nam/open-access-at-sage). |
spellingShingle | Original Articles Dörfelt, René Diels, Julia Hartmann, Katrin Evaluation of the performance of two new generation pulse oximeters in cats at different probe positions and under the influence of vasoconstriction |
title | Evaluation of the performance of two new generation pulse oximeters
in cats at different probe positions and under the influence of
vasoconstriction |
title_full | Evaluation of the performance of two new generation pulse oximeters
in cats at different probe positions and under the influence of
vasoconstriction |
title_fullStr | Evaluation of the performance of two new generation pulse oximeters
in cats at different probe positions and under the influence of
vasoconstriction |
title_full_unstemmed | Evaluation of the performance of two new generation pulse oximeters
in cats at different probe positions and under the influence of
vasoconstriction |
title_short | Evaluation of the performance of two new generation pulse oximeters
in cats at different probe positions and under the influence of
vasoconstriction |
title_sort | evaluation of the performance of two new generation pulse oximeters
in cats at different probe positions and under the influence of
vasoconstriction |
topic | Original Articles |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9510937/ https://www.ncbi.nlm.nih.gov/pubmed/34904479 http://dx.doi.org/10.1177/1098612X211063768 |
work_keys_str_mv | AT dorfeltrene evaluationoftheperformanceoftwonewgenerationpulseoximetersincatsatdifferentprobepositionsandundertheinfluenceofvasoconstriction AT dielsjulia evaluationoftheperformanceoftwonewgenerationpulseoximetersincatsatdifferentprobepositionsandundertheinfluenceofvasoconstriction AT hartmannkatrin evaluationoftheperformanceoftwonewgenerationpulseoximetersincatsatdifferentprobepositionsandundertheinfluenceofvasoconstriction |