Cargando…

Performance evaluation of the BC-720 auto hematology analyzer and establishment of the reference intervals of erythrocyte sedimentation rate in healthy adults

BACKGROUND: Erythrocyte sedimentation rate (ESR) is a new reporting parameter of the BC-720 auto hematology analyzer; however, no biological reference interval for healthy adults has been established for this parameter. METHODS: Outpatients or hospitalized patients with ESR test orders were selected...

Descripción completa

Detalles Bibliográficos
Autores principales: Shu, Guofang, Ding, Rui, Ding, Rong, He, Zhi, Shen, Yimin, Liu, Dongmei, Duan, Zhaohui
Formato: Online Artículo Texto
Lenguaje:English
Publicado: AME Publishing Company 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9511180/
https://www.ncbi.nlm.nih.gov/pubmed/36172102
http://dx.doi.org/10.21037/atm-22-3486
_version_ 1784797606888603648
author Shu, Guofang
Ding, Rui
Ding, Rong
He, Zhi
Shen, Yimin
Liu, Dongmei
Duan, Zhaohui
author_facet Shu, Guofang
Ding, Rui
Ding, Rong
He, Zhi
Shen, Yimin
Liu, Dongmei
Duan, Zhaohui
author_sort Shu, Guofang
collection PubMed
description BACKGROUND: Erythrocyte sedimentation rate (ESR) is a new reporting parameter of the BC-720 auto hematology analyzer; however, no biological reference interval for healthy adults has been established for this parameter. METHODS: Outpatients or hospitalized patients with ESR test orders were selected. The ESR was measured by the standard method of ESR (Westergren) recommended by the International Council for Standardization in Hematology (ICSH), the BC-720 hematology analyzer, and the LBY-XC40B auto ESR analyzer. The data were statistically analyzed and compared among different methods. The repeatability and carryover rate (CR) of the BC-720 were assessed in randomly selected samples for each range segment. Blood Samples from three hospitals in China were collected, and the reference interval of the BC-720 ESR was determined. RESULTS: The ESR results measured by the BC-720 correlated well with the Westergren method (r=0.957, y = 0.359 + 1.016x), and there was no significant difference between these two methods (P>0.05). The correlation between LBY-XC40B auto ESR analyzer and Westergren was y = 1 + 1.25x and r=0.856. The BC-720 ESR has good repeatability [standard deviation (SD) ≤1 mm/h, coefficient of variation (CV) ≤5%], and the CR was less than 1%. The 95th percentile of the biological reference interval for BC-720 ESR is 15 mm/h for men and 24 mm/h for women. CONCLUSIONS: The Mindray BC-720 ESR showed high accuracy and good repeatability, which provided a faster, safer, and more reliable method to measure ESR. The reference intervals for BC-720 ESR could guide better clinical decisions for the laboratories utilizing this new method.
format Online
Article
Text
id pubmed-9511180
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher AME Publishing Company
record_format MEDLINE/PubMed
spelling pubmed-95111802022-09-27 Performance evaluation of the BC-720 auto hematology analyzer and establishment of the reference intervals of erythrocyte sedimentation rate in healthy adults Shu, Guofang Ding, Rui Ding, Rong He, Zhi Shen, Yimin Liu, Dongmei Duan, Zhaohui Ann Transl Med Original Article BACKGROUND: Erythrocyte sedimentation rate (ESR) is a new reporting parameter of the BC-720 auto hematology analyzer; however, no biological reference interval for healthy adults has been established for this parameter. METHODS: Outpatients or hospitalized patients with ESR test orders were selected. The ESR was measured by the standard method of ESR (Westergren) recommended by the International Council for Standardization in Hematology (ICSH), the BC-720 hematology analyzer, and the LBY-XC40B auto ESR analyzer. The data were statistically analyzed and compared among different methods. The repeatability and carryover rate (CR) of the BC-720 were assessed in randomly selected samples for each range segment. Blood Samples from three hospitals in China were collected, and the reference interval of the BC-720 ESR was determined. RESULTS: The ESR results measured by the BC-720 correlated well with the Westergren method (r=0.957, y = 0.359 + 1.016x), and there was no significant difference between these two methods (P>0.05). The correlation between LBY-XC40B auto ESR analyzer and Westergren was y = 1 + 1.25x and r=0.856. The BC-720 ESR has good repeatability [standard deviation (SD) ≤1 mm/h, coefficient of variation (CV) ≤5%], and the CR was less than 1%. The 95th percentile of the biological reference interval for BC-720 ESR is 15 mm/h for men and 24 mm/h for women. CONCLUSIONS: The Mindray BC-720 ESR showed high accuracy and good repeatability, which provided a faster, safer, and more reliable method to measure ESR. The reference intervals for BC-720 ESR could guide better clinical decisions for the laboratories utilizing this new method. AME Publishing Company 2022-09 /pmc/articles/PMC9511180/ /pubmed/36172102 http://dx.doi.org/10.21037/atm-22-3486 Text en 2022 Annals of Translational Medicine. All rights reserved. https://creativecommons.org/licenses/by-nc-nd/4.0/Open Access Statement: This is an Open Access article distributed in accordance with the Creative Commons Attribution-NonCommercial-NoDerivs 4.0 International License (CC BY-NC-ND 4.0), which permits the non-commercial replication and distribution of the article with the strict proviso that no changes or edits are made and the original work is properly cited (including links to both the formal publication through the relevant DOI and the license). See: https://creativecommons.org/licenses/by-nc-nd/4.0 (https://creativecommons.org/licenses/by-nc-nd/4.0/) .
spellingShingle Original Article
Shu, Guofang
Ding, Rui
Ding, Rong
He, Zhi
Shen, Yimin
Liu, Dongmei
Duan, Zhaohui
Performance evaluation of the BC-720 auto hematology analyzer and establishment of the reference intervals of erythrocyte sedimentation rate in healthy adults
title Performance evaluation of the BC-720 auto hematology analyzer and establishment of the reference intervals of erythrocyte sedimentation rate in healthy adults
title_full Performance evaluation of the BC-720 auto hematology analyzer and establishment of the reference intervals of erythrocyte sedimentation rate in healthy adults
title_fullStr Performance evaluation of the BC-720 auto hematology analyzer and establishment of the reference intervals of erythrocyte sedimentation rate in healthy adults
title_full_unstemmed Performance evaluation of the BC-720 auto hematology analyzer and establishment of the reference intervals of erythrocyte sedimentation rate in healthy adults
title_short Performance evaluation of the BC-720 auto hematology analyzer and establishment of the reference intervals of erythrocyte sedimentation rate in healthy adults
title_sort performance evaluation of the bc-720 auto hematology analyzer and establishment of the reference intervals of erythrocyte sedimentation rate in healthy adults
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9511180/
https://www.ncbi.nlm.nih.gov/pubmed/36172102
http://dx.doi.org/10.21037/atm-22-3486
work_keys_str_mv AT shuguofang performanceevaluationofthebc720autohematologyanalyzerandestablishmentofthereferenceintervalsoferythrocytesedimentationrateinhealthyadults
AT dingrui performanceevaluationofthebc720autohematologyanalyzerandestablishmentofthereferenceintervalsoferythrocytesedimentationrateinhealthyadults
AT dingrong performanceevaluationofthebc720autohematologyanalyzerandestablishmentofthereferenceintervalsoferythrocytesedimentationrateinhealthyadults
AT hezhi performanceevaluationofthebc720autohematologyanalyzerandestablishmentofthereferenceintervalsoferythrocytesedimentationrateinhealthyadults
AT shenyimin performanceevaluationofthebc720autohematologyanalyzerandestablishmentofthereferenceintervalsoferythrocytesedimentationrateinhealthyadults
AT liudongmei performanceevaluationofthebc720autohematologyanalyzerandestablishmentofthereferenceintervalsoferythrocytesedimentationrateinhealthyadults
AT duanzhaohui performanceevaluationofthebc720autohematologyanalyzerandestablishmentofthereferenceintervalsoferythrocytesedimentationrateinhealthyadults