Quality assurance and long‐term stability of a novel 3‐in‐1 X‐ray system for brachytherapy

PURPOSE: A novel, mobile 3‐in‐1 X‐ray system featuring radiography, fluoroscopy, and cone‐beam computed tomography (CBCT) has been launched for brachytherapy recently. Currently, there is no quality assurance (QA) procedure explicitly applicable to this system equipped with innovative technologies s...

Descripción completa

Detalles Bibliográficos
Autores principales: Karius, Andre, Szkitsak, Juliane, Boronikolas, Vasilios, Fietkau, Rainer, Bert, Christoph
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9512339/
https://www.ncbi.nlm.nih.gov/pubmed/35848090
http://dx.doi.org/10.1002/acm2.13727
_version_ 1784797828529258496
author Karius, Andre
Szkitsak, Juliane
Boronikolas, Vasilios
Fietkau, Rainer
Bert, Christoph
author_facet Karius, Andre
Szkitsak, Juliane
Boronikolas, Vasilios
Fietkau, Rainer
Bert, Christoph
author_sort Karius, Andre
collection PubMed
description PURPOSE: A novel, mobile 3‐in‐1 X‐ray system featuring radiography, fluoroscopy, and cone‐beam computed tomography (CBCT) has been launched for brachytherapy recently. Currently, there is no quality assurance (QA) procedure explicitly applicable to this system equipped with innovative technologies such as dynamic jaws and motorized lasers. We developed a dedicated QA procedure and, based on its performance for a duration of 6 months, provide an assessment of the device's stability over time. METHODS: With the developed QA procedure, we assessed the system's planar and CBCT‐imaging performance by investigating geometric accuracy, CT‐number stability, contrast‐noise‐ratio, uniformity, spatial resolution, low‐contrast detectability, dynamic range, and X‐ray exposure using dedicated phantoms. Furthermore, we evaluated geometric stability by using the flexmap‐approach and investigated the device's laser‐ and jaw‐positioning accuracy with an in‐house test phantom. CBCT‐ and planar‐imaging protocols for pelvis, breast, and abdomen imaging were examined. RESULTS: Planar‐ and CBCT‐imaging performances were widely stable with a geometric accuracy ≤1 mm, CT‐number stability of up to 46 HU, and uniformity variations of up to 48 HU over time. For planar imaging, low‐contrast detectability and dynamic range exceeded current recommendations. Although geometric stability was considered tolerable, partly substantial positioning inaccuracies of up to more than 120 mm and −13 mm were obtained for lasers and jaws, respectively. X‐ray exposure showed small variations of ≤0.56 μGy and ≤0.76 mGy for planar‐ and CBCT‐imaging, respectively. The conductance of the QA procedure allowed a smooth evaluation of the system's overall performance. CONCLUSION: We developed a QA workflow for a novel 3‐in‐1 X‐ray system allowing to assess the device's imaging and hardware performance. The system showed in general a reasonable imaging performance and stability over time, whereas improvements regarding laser and jaw accuracy are strictly required.
format Online
Article
Text
id pubmed-9512339
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-95123392022-09-30 Quality assurance and long‐term stability of a novel 3‐in‐1 X‐ray system for brachytherapy Karius, Andre Szkitsak, Juliane Boronikolas, Vasilios Fietkau, Rainer Bert, Christoph J Appl Clin Med Phys Technical Notes PURPOSE: A novel, mobile 3‐in‐1 X‐ray system featuring radiography, fluoroscopy, and cone‐beam computed tomography (CBCT) has been launched for brachytherapy recently. Currently, there is no quality assurance (QA) procedure explicitly applicable to this system equipped with innovative technologies such as dynamic jaws and motorized lasers. We developed a dedicated QA procedure and, based on its performance for a duration of 6 months, provide an assessment of the device's stability over time. METHODS: With the developed QA procedure, we assessed the system's planar and CBCT‐imaging performance by investigating geometric accuracy, CT‐number stability, contrast‐noise‐ratio, uniformity, spatial resolution, low‐contrast detectability, dynamic range, and X‐ray exposure using dedicated phantoms. Furthermore, we evaluated geometric stability by using the flexmap‐approach and investigated the device's laser‐ and jaw‐positioning accuracy with an in‐house test phantom. CBCT‐ and planar‐imaging protocols for pelvis, breast, and abdomen imaging were examined. RESULTS: Planar‐ and CBCT‐imaging performances were widely stable with a geometric accuracy ≤1 mm, CT‐number stability of up to 46 HU, and uniformity variations of up to 48 HU over time. For planar imaging, low‐contrast detectability and dynamic range exceeded current recommendations. Although geometric stability was considered tolerable, partly substantial positioning inaccuracies of up to more than 120 mm and −13 mm were obtained for lasers and jaws, respectively. X‐ray exposure showed small variations of ≤0.56 μGy and ≤0.76 mGy for planar‐ and CBCT‐imaging, respectively. The conductance of the QA procedure allowed a smooth evaluation of the system's overall performance. CONCLUSION: We developed a QA workflow for a novel 3‐in‐1 X‐ray system allowing to assess the device's imaging and hardware performance. The system showed in general a reasonable imaging performance and stability over time, whereas improvements regarding laser and jaw accuracy are strictly required. John Wiley and Sons Inc. 2022-07-18 /pmc/articles/PMC9512339/ /pubmed/35848090 http://dx.doi.org/10.1002/acm2.13727 Text en © 2022 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, LLC on behalf of The American Association of Physicists in Medicine. https://creativecommons.org/licenses/by/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Technical Notes
Karius, Andre
Szkitsak, Juliane
Boronikolas, Vasilios
Fietkau, Rainer
Bert, Christoph
Quality assurance and long‐term stability of a novel 3‐in‐1 X‐ray system for brachytherapy
title Quality assurance and long‐term stability of a novel 3‐in‐1 X‐ray system for brachytherapy
title_full Quality assurance and long‐term stability of a novel 3‐in‐1 X‐ray system for brachytherapy
title_fullStr Quality assurance and long‐term stability of a novel 3‐in‐1 X‐ray system for brachytherapy
title_full_unstemmed Quality assurance and long‐term stability of a novel 3‐in‐1 X‐ray system for brachytherapy
title_short Quality assurance and long‐term stability of a novel 3‐in‐1 X‐ray system for brachytherapy
title_sort quality assurance and long‐term stability of a novel 3‐in‐1 x‐ray system for brachytherapy
topic Technical Notes
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9512339/
https://www.ncbi.nlm.nih.gov/pubmed/35848090
http://dx.doi.org/10.1002/acm2.13727
work_keys_str_mv AT kariusandre qualityassuranceandlongtermstabilityofanovel3in1xraysystemforbrachytherapy
AT szkitsakjuliane qualityassuranceandlongtermstabilityofanovel3in1xraysystemforbrachytherapy
AT boronikolasvasilios qualityassuranceandlongtermstabilityofanovel3in1xraysystemforbrachytherapy
AT fietkaurainer qualityassuranceandlongtermstabilityofanovel3in1xraysystemforbrachytherapy
AT bertchristoph qualityassuranceandlongtermstabilityofanovel3in1xraysystemforbrachytherapy