Verification of stereotactic radiosurgery plans for multiple brain metastases using a virtual phantom-based procedure
BACKGROUND: The purpose of this study was to describe the use of the VIPER software for patient-specific quality assurance (PSQA) of single-isocenter multitarget (SIMT) stereotactic radiosurgery (SRS) plans. MATERIALS AND METHODS: Twenty clinical of intensity-modulated (IMRT) SIMT SRS plans were rev...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Via Medica
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9518761/ https://www.ncbi.nlm.nih.gov/pubmed/36186702 http://dx.doi.org/10.5603/RPOR.a2022.0042 |
Sumario: | BACKGROUND: The purpose of this study was to describe the use of the VIPER software for patient-specific quality assurance (PSQA) of single-isocenter multitarget (SIMT) stereotactic radiosurgery (SRS) plans. MATERIALS AND METHODS: Twenty clinical of intensity-modulated (IMRT) SIMT SRS plans were reviewed. A total of 88 brain metastases were included. Number of lesions per plan and their individual volumes ranged from 2 to 35 and from 0.03 to 32.8 cm(3), respectively. Plans were designed with the Eclipse system, and delivered using a Varian CLINAC linac. SRS technique consisted of non-coplanar static-field sliding-window IMRT. Each plan was mapped onto a virtual cylindrical water phantom (VCP) in the Eclipse to calculate a 3D dose distribution (verification plan). The VIPER software reconstructed the 3D dose distribution inside the VCP from the acquired in-air electronic portal image device (EPID) images of the treatment fields. A 3D gamma analysis was used to compare the reconstructed doses to the Eclipse planned doses on the VCP. Gamma passing rates (GPRs) were calculated using 3% global/2 mm criteria and dose thresholds ranged from 10% to 90% of the maximum dose. RESULTS: The averages (± 1 SD) of the 3D GPRs over the 20 SRS plans were: 99.9 ± 0.2%, 99.7 ± 0.3%, 99.6 ± 0.5%, 99.3 ± 0.9%,99.1 ± 1.6%, 99.0 ± 1.6%, and 98.5 ± 3.3%, for dose thresholds of 10%, 20%, 30%, 50%, 70%, 80% and 90% respectively. CONCLUSIONS: This work shows the feasibility of the VIPER software for PSQA of SIMT SRS plans, being a reliable alternative to commercially available 2D detector arrays. |
---|