Comparison of efficacy of filter paper cyanmethemoglobin method with automated hematology analyzer for estimation of hemoglobin

BACKGROUND: Screening of hemoglobin (Hb) before blood donation is one among the vital tests. It is performed to select a blood donor to prevent the collection of blood from an anemic person. However, no accurate, cost-effective, reliable, and standardized method is available to estimate Hb. OBJECTIV...

Descripción completa

Detalles Bibliográficos
Autores principales: Kumar, Lalit, Kangle, Ranjit
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Wolters Kluwer - Medknow 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9528559/
https://www.ncbi.nlm.nih.gov/pubmed/36199413
http://dx.doi.org/10.4103/ajts.AJTS_135_16
_version_ 1784801325027950592
author Kumar, Lalit
Kangle, Ranjit
author_facet Kumar, Lalit
Kangle, Ranjit
author_sort Kumar, Lalit
collection PubMed
description BACKGROUND: Screening of hemoglobin (Hb) before blood donation is one among the vital tests. It is performed to select a blood donor to prevent the collection of blood from an anemic person. However, no accurate, cost-effective, reliable, and standardized method is available to estimate Hb. OBJECTIVE: The aim is to evaluate the efficacy of filter paper cyanmethemoglobin (FPCH) method with the automated hematology analyzer in the estimation of Hb concentration for screening of a suitable donor. METHODOLOGY: This was a cross-sectional study in which the blood samples of 2000 patients visiting KLE's Dr. Prabhakar Kore Charitable Hospital, Belagavi, were collected in vials and directly estimated for Hb using automated hematology analyzer. To evaluate the efficacy of FPCH, 20 μL of blood sample was transferred onto Whatman filter paper and dried at room temperature. After drying, it was placed in 5 mL of Drabkin's solution for 30 min. Optical density was estimated by measuring the absorbance. Data were analyzed using SPSS version 20. The correlation coefficient, paired t-test, and difference between the means of both the methods were calculated. RESULTS: The mean Hb estimated by FPCH was 11.25 g/dL and automated hematology analyzer gave 11.35 g/dL. The difference in the means of both the methods was 0.1 g/dL. Paired t-test was done to test the level of significance and the result was 8.151 (95% confidence interval: 0.08–0.13 g/dL, P < 0.001). The correlation coefficient was found to be 0.976 (P < 0.001). CONCLUSION: FPCH is an efficient method, which is comparable to the automated hematology analyzers for Hb estimation. It could be used as an alternative screening tool for detection of Hb in a blood donation camp.
format Online
Article
Text
id pubmed-9528559
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Wolters Kluwer - Medknow
record_format MEDLINE/PubMed
spelling pubmed-95285592022-10-04 Comparison of efficacy of filter paper cyanmethemoglobin method with automated hematology analyzer for estimation of hemoglobin Kumar, Lalit Kangle, Ranjit Asian J Transfus Sci Original Article BACKGROUND: Screening of hemoglobin (Hb) before blood donation is one among the vital tests. It is performed to select a blood donor to prevent the collection of blood from an anemic person. However, no accurate, cost-effective, reliable, and standardized method is available to estimate Hb. OBJECTIVE: The aim is to evaluate the efficacy of filter paper cyanmethemoglobin (FPCH) method with the automated hematology analyzer in the estimation of Hb concentration for screening of a suitable donor. METHODOLOGY: This was a cross-sectional study in which the blood samples of 2000 patients visiting KLE's Dr. Prabhakar Kore Charitable Hospital, Belagavi, were collected in vials and directly estimated for Hb using automated hematology analyzer. To evaluate the efficacy of FPCH, 20 μL of blood sample was transferred onto Whatman filter paper and dried at room temperature. After drying, it was placed in 5 mL of Drabkin's solution for 30 min. Optical density was estimated by measuring the absorbance. Data were analyzed using SPSS version 20. The correlation coefficient, paired t-test, and difference between the means of both the methods were calculated. RESULTS: The mean Hb estimated by FPCH was 11.25 g/dL and automated hematology analyzer gave 11.35 g/dL. The difference in the means of both the methods was 0.1 g/dL. Paired t-test was done to test the level of significance and the result was 8.151 (95% confidence interval: 0.08–0.13 g/dL, P < 0.001). The correlation coefficient was found to be 0.976 (P < 0.001). CONCLUSION: FPCH is an efficient method, which is comparable to the automated hematology analyzers for Hb estimation. It could be used as an alternative screening tool for detection of Hb in a blood donation camp. Wolters Kluwer - Medknow 2022 2022-05-26 /pmc/articles/PMC9528559/ /pubmed/36199413 http://dx.doi.org/10.4103/ajts.AJTS_135_16 Text en Copyright: © 2022 Asian Journal of Transfusion Science https://creativecommons.org/licenses/by-nc-sa/4.0/This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.
spellingShingle Original Article
Kumar, Lalit
Kangle, Ranjit
Comparison of efficacy of filter paper cyanmethemoglobin method with automated hematology analyzer for estimation of hemoglobin
title Comparison of efficacy of filter paper cyanmethemoglobin method with automated hematology analyzer for estimation of hemoglobin
title_full Comparison of efficacy of filter paper cyanmethemoglobin method with automated hematology analyzer for estimation of hemoglobin
title_fullStr Comparison of efficacy of filter paper cyanmethemoglobin method with automated hematology analyzer for estimation of hemoglobin
title_full_unstemmed Comparison of efficacy of filter paper cyanmethemoglobin method with automated hematology analyzer for estimation of hemoglobin
title_short Comparison of efficacy of filter paper cyanmethemoglobin method with automated hematology analyzer for estimation of hemoglobin
title_sort comparison of efficacy of filter paper cyanmethemoglobin method with automated hematology analyzer for estimation of hemoglobin
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9528559/
https://www.ncbi.nlm.nih.gov/pubmed/36199413
http://dx.doi.org/10.4103/ajts.AJTS_135_16
work_keys_str_mv AT kumarlalit comparisonofefficacyoffilterpapercyanmethemoglobinmethodwithautomatedhematologyanalyzerforestimationofhemoglobin
AT kangleranjit comparisonofefficacyoffilterpapercyanmethemoglobinmethodwithautomatedhematologyanalyzerforestimationofhemoglobin