Commissioning of GPU-based multi-criteria optimizer combined with plan navigation tools for high-dose-rate brachytherapy

PURPOSE: Recently, our GPU-based multi-criteria optimization (gMCO) algorithm has been integrated in a graphical user interface (gMCO-GUI) that allows real-time plan navigation through a gMCO-generated set of Pareto-optimal plans for high-dose-rate (HDR) brachytherapy. This work reports on the commi...

Descripción completa

Detalles Bibliográficos
Autores principales: Bélanger, Cédric, Aubin, Sylviane, Beaulieu, Luc, Poulin, Éric
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Termedia Publishing House 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9528822/
https://www.ncbi.nlm.nih.gov/pubmed/36199940
http://dx.doi.org/10.5114/jcb.2022.118995
_version_ 1784801370044366848
author Bélanger, Cédric
Aubin, Sylviane
Beaulieu, Luc
Poulin, Éric
author_facet Bélanger, Cédric
Aubin, Sylviane
Beaulieu, Luc
Poulin, Éric
author_sort Bélanger, Cédric
collection PubMed
description PURPOSE: Recently, our GPU-based multi-criteria optimization (gMCO) algorithm has been integrated in a graphical user interface (gMCO-GUI) that allows real-time plan navigation through a gMCO-generated set of Pareto-optimal plans for high-dose-rate (HDR) brachytherapy. This work reports on the commissioning of the gMCO algorithm into clinical workflow. MATERIAL AND METHODS: Our MCO workflow was validated against Oncentra Prostate v. 4.2.2 (OcP) and Oncentra Brachy v. 4.6.0 (OcB). 40 HDR prostate brachytherapy patients (20 with OcP and 20 with OcB) were retrospectively re-planned with gMCO algorithm by generating 2,000 Pareto-optimal plans. A single gMCO treatment plan was exported using gMCO-GUI plan navigation tools. The optimized dwell positions and dwell times of gMCO plans were exported via DICOM RTPLAN files to OcP/OcB, where final dosimetry was calculated. TG43 implementation in gMCO was validated against the consensus data of flexisource. Five analytical shapes were used as the ground truth for volume calculations. Dose-volume histogram (DVH) curves generated by gMCO were compared with the ones generated by OcP/OcB. 3D dose distributions (and isodose lines) were validated against OcP/OcB using dice similarity coefficient (DSC), 95% undirected Hausdorff distance (95% HD), and γ analysis. RESULTS: Differences between –0.4% and 0.3% were observed between gMCO calculated dose rates and the flexisource consensus data. gMCO volumes were within ±2% agreement in 3/5 volumes (deviations within –2.9% and 0.1%). For 9 key DVH indices, the differences between gMCO and OcP/OcB were within ±1.2%. Regarding the accuracy of key isodose lines, the mean DSC was greater than 0.98, and the mean 95% HD was below 0.4 mm. The fraction of voxels with γ ≤ 1 was greater than 99% for all cases with 1%/1 mm threshold. CONCLUSIONS: The GPU-based MCO workflow was successfully integrated into the clinical workflow and validated against OcP and OcB.
format Online
Article
Text
id pubmed-9528822
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Termedia Publishing House
record_format MEDLINE/PubMed
spelling pubmed-95288222022-10-04 Commissioning of GPU-based multi-criteria optimizer combined with plan navigation tools for high-dose-rate brachytherapy Bélanger, Cédric Aubin, Sylviane Beaulieu, Luc Poulin, Éric J Contemp Brachytherapy Original Paper PURPOSE: Recently, our GPU-based multi-criteria optimization (gMCO) algorithm has been integrated in a graphical user interface (gMCO-GUI) that allows real-time plan navigation through a gMCO-generated set of Pareto-optimal plans for high-dose-rate (HDR) brachytherapy. This work reports on the commissioning of the gMCO algorithm into clinical workflow. MATERIAL AND METHODS: Our MCO workflow was validated against Oncentra Prostate v. 4.2.2 (OcP) and Oncentra Brachy v. 4.6.0 (OcB). 40 HDR prostate brachytherapy patients (20 with OcP and 20 with OcB) were retrospectively re-planned with gMCO algorithm by generating 2,000 Pareto-optimal plans. A single gMCO treatment plan was exported using gMCO-GUI plan navigation tools. The optimized dwell positions and dwell times of gMCO plans were exported via DICOM RTPLAN files to OcP/OcB, where final dosimetry was calculated. TG43 implementation in gMCO was validated against the consensus data of flexisource. Five analytical shapes were used as the ground truth for volume calculations. Dose-volume histogram (DVH) curves generated by gMCO were compared with the ones generated by OcP/OcB. 3D dose distributions (and isodose lines) were validated against OcP/OcB using dice similarity coefficient (DSC), 95% undirected Hausdorff distance (95% HD), and γ analysis. RESULTS: Differences between –0.4% and 0.3% were observed between gMCO calculated dose rates and the flexisource consensus data. gMCO volumes were within ±2% agreement in 3/5 volumes (deviations within –2.9% and 0.1%). For 9 key DVH indices, the differences between gMCO and OcP/OcB were within ±1.2%. Regarding the accuracy of key isodose lines, the mean DSC was greater than 0.98, and the mean 95% HD was below 0.4 mm. The fraction of voxels with γ ≤ 1 was greater than 99% for all cases with 1%/1 mm threshold. CONCLUSIONS: The GPU-based MCO workflow was successfully integrated into the clinical workflow and validated against OcP and OcB. Termedia Publishing House 2022-08-23 2022-08 /pmc/articles/PMC9528822/ /pubmed/36199940 http://dx.doi.org/10.5114/jcb.2022.118995 Text en Copyright © 2022 Termedia https://creativecommons.org/licenses/by-nc-sa/4.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International (CC BY-NC-SA 4.0). License (http://creativecommons.org/licenses/by-nc-sa/4.0/ (https://creativecommons.org/licenses/by-nc-sa/4.0/) )
spellingShingle Original Paper
Bélanger, Cédric
Aubin, Sylviane
Beaulieu, Luc
Poulin, Éric
Commissioning of GPU-based multi-criteria optimizer combined with plan navigation tools for high-dose-rate brachytherapy
title Commissioning of GPU-based multi-criteria optimizer combined with plan navigation tools for high-dose-rate brachytherapy
title_full Commissioning of GPU-based multi-criteria optimizer combined with plan navigation tools for high-dose-rate brachytherapy
title_fullStr Commissioning of GPU-based multi-criteria optimizer combined with plan navigation tools for high-dose-rate brachytherapy
title_full_unstemmed Commissioning of GPU-based multi-criteria optimizer combined with plan navigation tools for high-dose-rate brachytherapy
title_short Commissioning of GPU-based multi-criteria optimizer combined with plan navigation tools for high-dose-rate brachytherapy
title_sort commissioning of gpu-based multi-criteria optimizer combined with plan navigation tools for high-dose-rate brachytherapy
topic Original Paper
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9528822/
https://www.ncbi.nlm.nih.gov/pubmed/36199940
http://dx.doi.org/10.5114/jcb.2022.118995
work_keys_str_mv AT belangercedric commissioningofgpubasedmulticriteriaoptimizercombinedwithplannavigationtoolsforhighdoseratebrachytherapy
AT aubinsylviane commissioningofgpubasedmulticriteriaoptimizercombinedwithplannavigationtoolsforhighdoseratebrachytherapy
AT beaulieuluc commissioningofgpubasedmulticriteriaoptimizercombinedwithplannavigationtoolsforhighdoseratebrachytherapy
AT poulineric commissioningofgpubasedmulticriteriaoptimizercombinedwithplannavigationtoolsforhighdoseratebrachytherapy