Dosimetric comparison of two afterloaders and treatment planning systems in vaginal high-dose-rate brachytherapy

PURPOSE: In treatment planning for high-dose-rate (HDR) single-channel vaginal cylinder brachytherapy, dose distribution along the cylinder is influenced by the anisotropy of the source. Differences in anisotropy are due to differences in source dimensions and characteristics. In this study, we comp...

Descripción completa

Detalles Bibliográficos
Autores principales: Marshall, Deborah C., Bloom, Julie R., Ghiassi-Nejad, Zahra, Gupta, Vishal, Saitta, Audrey, Powers, Allison, Sheu, Ren-Dih, Lo, Yeh-Chi, Dumane, Vishruta
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Termedia Publishing House 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9528830/
https://www.ncbi.nlm.nih.gov/pubmed/36199949
http://dx.doi.org/10.5114/jcb.2022.118633
_version_ 1784801372051341312
author Marshall, Deborah C.
Bloom, Julie R.
Ghiassi-Nejad, Zahra
Gupta, Vishal
Saitta, Audrey
Powers, Allison
Sheu, Ren-Dih
Lo, Yeh-Chi
Dumane, Vishruta
author_facet Marshall, Deborah C.
Bloom, Julie R.
Ghiassi-Nejad, Zahra
Gupta, Vishal
Saitta, Audrey
Powers, Allison
Sheu, Ren-Dih
Lo, Yeh-Chi
Dumane, Vishruta
author_sort Marshall, Deborah C.
collection PubMed
description PURPOSE: In treatment planning for high-dose-rate (HDR) single-channel vaginal cylinder brachytherapy, dose distribution along the cylinder is influenced by the anisotropy of the source. Differences in anisotropy are due to differences in source dimensions and characteristics. In this study, we compared HDR vaginal cylinder brachytherapy treatment plans from two afterloader/treatment planning systems. MATERIAL AND METHODS: Seventy-five plans with prescription to the surface were generated for cylinders in Varian BrachyVision and Elekta Oncentra. To understand the impact of source anisotropy on dose distribution to the surface of the cylinder, potential effect caused by differences in cylinder geometry between systems was eliminated by re-planning Varian cylinder using Elekta source model. Mean relative dose was calculated for each point as well as the dome and length of the cylinder. Related-samples Wilcoxon signed-rank tests were performed to compare the mean relative dose between systems. RESULTS: Treatment plans with VariSource iX source and cylinder demonstrated 16.2% lower (p < 0.001) dose at the tip compared to Elekta v.3. Average dose to the points along the dome of cylinder was 128.4% ±17.9% prescription dose with VariSource iX source and cylinder, and 99.9% ±4.3% with Elekta v.3 source and cylinder. For the same cylinder geometry, the effect of source characteristics produced up to 36.8% difference in dose homogeneity. When cylinder types were planned with the same source, there was no significant difference in dose distribution. CONCLUSIONS: This study demonstrates that the effect of source characteristics produced up to 37% difference in dose homogeneity when comparing two afterloader/treatment planning systems, independent of cylinder geometry. This insight on variation in dose surrounding source system is imperative for dosimetry considerations. Depending on the choice of afterloader, the extent of EQD(2) for tumor control versus normal tissue toxicity can vary.
format Online
Article
Text
id pubmed-9528830
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Termedia Publishing House
record_format MEDLINE/PubMed
spelling pubmed-95288302022-10-04 Dosimetric comparison of two afterloaders and treatment planning systems in vaginal high-dose-rate brachytherapy Marshall, Deborah C. Bloom, Julie R. Ghiassi-Nejad, Zahra Gupta, Vishal Saitta, Audrey Powers, Allison Sheu, Ren-Dih Lo, Yeh-Chi Dumane, Vishruta J Contemp Brachytherapy Original Paper PURPOSE: In treatment planning for high-dose-rate (HDR) single-channel vaginal cylinder brachytherapy, dose distribution along the cylinder is influenced by the anisotropy of the source. Differences in anisotropy are due to differences in source dimensions and characteristics. In this study, we compared HDR vaginal cylinder brachytherapy treatment plans from two afterloader/treatment planning systems. MATERIAL AND METHODS: Seventy-five plans with prescription to the surface were generated for cylinders in Varian BrachyVision and Elekta Oncentra. To understand the impact of source anisotropy on dose distribution to the surface of the cylinder, potential effect caused by differences in cylinder geometry between systems was eliminated by re-planning Varian cylinder using Elekta source model. Mean relative dose was calculated for each point as well as the dome and length of the cylinder. Related-samples Wilcoxon signed-rank tests were performed to compare the mean relative dose between systems. RESULTS: Treatment plans with VariSource iX source and cylinder demonstrated 16.2% lower (p < 0.001) dose at the tip compared to Elekta v.3. Average dose to the points along the dome of cylinder was 128.4% ±17.9% prescription dose with VariSource iX source and cylinder, and 99.9% ±4.3% with Elekta v.3 source and cylinder. For the same cylinder geometry, the effect of source characteristics produced up to 36.8% difference in dose homogeneity. When cylinder types were planned with the same source, there was no significant difference in dose distribution. CONCLUSIONS: This study demonstrates that the effect of source characteristics produced up to 37% difference in dose homogeneity when comparing two afterloader/treatment planning systems, independent of cylinder geometry. This insight on variation in dose surrounding source system is imperative for dosimetry considerations. Depending on the choice of afterloader, the extent of EQD(2) for tumor control versus normal tissue toxicity can vary. Termedia Publishing House 2022-08-08 2022-08 /pmc/articles/PMC9528830/ /pubmed/36199949 http://dx.doi.org/10.5114/jcb.2022.118633 Text en Copyright © 2022 Termedia https://creativecommons.org/licenses/by-nc-sa/4.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International (CC BY-NC-SA 4.0). License (http://creativecommons.org/licenses/by-nc-sa/4.0/ (https://creativecommons.org/licenses/by-nc-sa/4.0/) )
spellingShingle Original Paper
Marshall, Deborah C.
Bloom, Julie R.
Ghiassi-Nejad, Zahra
Gupta, Vishal
Saitta, Audrey
Powers, Allison
Sheu, Ren-Dih
Lo, Yeh-Chi
Dumane, Vishruta
Dosimetric comparison of two afterloaders and treatment planning systems in vaginal high-dose-rate brachytherapy
title Dosimetric comparison of two afterloaders and treatment planning systems in vaginal high-dose-rate brachytherapy
title_full Dosimetric comparison of two afterloaders and treatment planning systems in vaginal high-dose-rate brachytherapy
title_fullStr Dosimetric comparison of two afterloaders and treatment planning systems in vaginal high-dose-rate brachytherapy
title_full_unstemmed Dosimetric comparison of two afterloaders and treatment planning systems in vaginal high-dose-rate brachytherapy
title_short Dosimetric comparison of two afterloaders and treatment planning systems in vaginal high-dose-rate brachytherapy
title_sort dosimetric comparison of two afterloaders and treatment planning systems in vaginal high-dose-rate brachytherapy
topic Original Paper
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9528830/
https://www.ncbi.nlm.nih.gov/pubmed/36199949
http://dx.doi.org/10.5114/jcb.2022.118633
work_keys_str_mv AT marshalldeborahc dosimetriccomparisonoftwoafterloadersandtreatmentplanningsystemsinvaginalhighdoseratebrachytherapy
AT bloomjulier dosimetriccomparisonoftwoafterloadersandtreatmentplanningsystemsinvaginalhighdoseratebrachytherapy
AT ghiassinejadzahra dosimetriccomparisonoftwoafterloadersandtreatmentplanningsystemsinvaginalhighdoseratebrachytherapy
AT guptavishal dosimetriccomparisonoftwoafterloadersandtreatmentplanningsystemsinvaginalhighdoseratebrachytherapy
AT saittaaudrey dosimetriccomparisonoftwoafterloadersandtreatmentplanningsystemsinvaginalhighdoseratebrachytherapy
AT powersallison dosimetriccomparisonoftwoafterloadersandtreatmentplanningsystemsinvaginalhighdoseratebrachytherapy
AT sheurendih dosimetriccomparisonoftwoafterloadersandtreatmentplanningsystemsinvaginalhighdoseratebrachytherapy
AT loyehchi dosimetriccomparisonoftwoafterloadersandtreatmentplanningsystemsinvaginalhighdoseratebrachytherapy
AT dumanevishruta dosimetriccomparisonoftwoafterloadersandtreatmentplanningsystemsinvaginalhighdoseratebrachytherapy