Dosimetric comparison of two afterloaders and treatment planning systems in vaginal high-dose-rate brachytherapy
PURPOSE: In treatment planning for high-dose-rate (HDR) single-channel vaginal cylinder brachytherapy, dose distribution along the cylinder is influenced by the anisotropy of the source. Differences in anisotropy are due to differences in source dimensions and characteristics. In this study, we comp...
Autores principales: | , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Termedia Publishing House
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9528830/ https://www.ncbi.nlm.nih.gov/pubmed/36199949 http://dx.doi.org/10.5114/jcb.2022.118633 |
_version_ | 1784801372051341312 |
---|---|
author | Marshall, Deborah C. Bloom, Julie R. Ghiassi-Nejad, Zahra Gupta, Vishal Saitta, Audrey Powers, Allison Sheu, Ren-Dih Lo, Yeh-Chi Dumane, Vishruta |
author_facet | Marshall, Deborah C. Bloom, Julie R. Ghiassi-Nejad, Zahra Gupta, Vishal Saitta, Audrey Powers, Allison Sheu, Ren-Dih Lo, Yeh-Chi Dumane, Vishruta |
author_sort | Marshall, Deborah C. |
collection | PubMed |
description | PURPOSE: In treatment planning for high-dose-rate (HDR) single-channel vaginal cylinder brachytherapy, dose distribution along the cylinder is influenced by the anisotropy of the source. Differences in anisotropy are due to differences in source dimensions and characteristics. In this study, we compared HDR vaginal cylinder brachytherapy treatment plans from two afterloader/treatment planning systems. MATERIAL AND METHODS: Seventy-five plans with prescription to the surface were generated for cylinders in Varian BrachyVision and Elekta Oncentra. To understand the impact of source anisotropy on dose distribution to the surface of the cylinder, potential effect caused by differences in cylinder geometry between systems was eliminated by re-planning Varian cylinder using Elekta source model. Mean relative dose was calculated for each point as well as the dome and length of the cylinder. Related-samples Wilcoxon signed-rank tests were performed to compare the mean relative dose between systems. RESULTS: Treatment plans with VariSource iX source and cylinder demonstrated 16.2% lower (p < 0.001) dose at the tip compared to Elekta v.3. Average dose to the points along the dome of cylinder was 128.4% ±17.9% prescription dose with VariSource iX source and cylinder, and 99.9% ±4.3% with Elekta v.3 source and cylinder. For the same cylinder geometry, the effect of source characteristics produced up to 36.8% difference in dose homogeneity. When cylinder types were planned with the same source, there was no significant difference in dose distribution. CONCLUSIONS: This study demonstrates that the effect of source characteristics produced up to 37% difference in dose homogeneity when comparing two afterloader/treatment planning systems, independent of cylinder geometry. This insight on variation in dose surrounding source system is imperative for dosimetry considerations. Depending on the choice of afterloader, the extent of EQD(2) for tumor control versus normal tissue toxicity can vary. |
format | Online Article Text |
id | pubmed-9528830 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | Termedia Publishing House |
record_format | MEDLINE/PubMed |
spelling | pubmed-95288302022-10-04 Dosimetric comparison of two afterloaders and treatment planning systems in vaginal high-dose-rate brachytherapy Marshall, Deborah C. Bloom, Julie R. Ghiassi-Nejad, Zahra Gupta, Vishal Saitta, Audrey Powers, Allison Sheu, Ren-Dih Lo, Yeh-Chi Dumane, Vishruta J Contemp Brachytherapy Original Paper PURPOSE: In treatment planning for high-dose-rate (HDR) single-channel vaginal cylinder brachytherapy, dose distribution along the cylinder is influenced by the anisotropy of the source. Differences in anisotropy are due to differences in source dimensions and characteristics. In this study, we compared HDR vaginal cylinder brachytherapy treatment plans from two afterloader/treatment planning systems. MATERIAL AND METHODS: Seventy-five plans with prescription to the surface were generated for cylinders in Varian BrachyVision and Elekta Oncentra. To understand the impact of source anisotropy on dose distribution to the surface of the cylinder, potential effect caused by differences in cylinder geometry between systems was eliminated by re-planning Varian cylinder using Elekta source model. Mean relative dose was calculated for each point as well as the dome and length of the cylinder. Related-samples Wilcoxon signed-rank tests were performed to compare the mean relative dose between systems. RESULTS: Treatment plans with VariSource iX source and cylinder demonstrated 16.2% lower (p < 0.001) dose at the tip compared to Elekta v.3. Average dose to the points along the dome of cylinder was 128.4% ±17.9% prescription dose with VariSource iX source and cylinder, and 99.9% ±4.3% with Elekta v.3 source and cylinder. For the same cylinder geometry, the effect of source characteristics produced up to 36.8% difference in dose homogeneity. When cylinder types were planned with the same source, there was no significant difference in dose distribution. CONCLUSIONS: This study demonstrates that the effect of source characteristics produced up to 37% difference in dose homogeneity when comparing two afterloader/treatment planning systems, independent of cylinder geometry. This insight on variation in dose surrounding source system is imperative for dosimetry considerations. Depending on the choice of afterloader, the extent of EQD(2) for tumor control versus normal tissue toxicity can vary. Termedia Publishing House 2022-08-08 2022-08 /pmc/articles/PMC9528830/ /pubmed/36199949 http://dx.doi.org/10.5114/jcb.2022.118633 Text en Copyright © 2022 Termedia https://creativecommons.org/licenses/by-nc-sa/4.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International (CC BY-NC-SA 4.0). License (http://creativecommons.org/licenses/by-nc-sa/4.0/ (https://creativecommons.org/licenses/by-nc-sa/4.0/) ) |
spellingShingle | Original Paper Marshall, Deborah C. Bloom, Julie R. Ghiassi-Nejad, Zahra Gupta, Vishal Saitta, Audrey Powers, Allison Sheu, Ren-Dih Lo, Yeh-Chi Dumane, Vishruta Dosimetric comparison of two afterloaders and treatment planning systems in vaginal high-dose-rate brachytherapy |
title | Dosimetric comparison of two afterloaders and treatment planning systems in vaginal high-dose-rate brachytherapy |
title_full | Dosimetric comparison of two afterloaders and treatment planning systems in vaginal high-dose-rate brachytherapy |
title_fullStr | Dosimetric comparison of two afterloaders and treatment planning systems in vaginal high-dose-rate brachytherapy |
title_full_unstemmed | Dosimetric comparison of two afterloaders and treatment planning systems in vaginal high-dose-rate brachytherapy |
title_short | Dosimetric comparison of two afterloaders and treatment planning systems in vaginal high-dose-rate brachytherapy |
title_sort | dosimetric comparison of two afterloaders and treatment planning systems in vaginal high-dose-rate brachytherapy |
topic | Original Paper |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9528830/ https://www.ncbi.nlm.nih.gov/pubmed/36199949 http://dx.doi.org/10.5114/jcb.2022.118633 |
work_keys_str_mv | AT marshalldeborahc dosimetriccomparisonoftwoafterloadersandtreatmentplanningsystemsinvaginalhighdoseratebrachytherapy AT bloomjulier dosimetriccomparisonoftwoafterloadersandtreatmentplanningsystemsinvaginalhighdoseratebrachytherapy AT ghiassinejadzahra dosimetriccomparisonoftwoafterloadersandtreatmentplanningsystemsinvaginalhighdoseratebrachytherapy AT guptavishal dosimetriccomparisonoftwoafterloadersandtreatmentplanningsystemsinvaginalhighdoseratebrachytherapy AT saittaaudrey dosimetriccomparisonoftwoafterloadersandtreatmentplanningsystemsinvaginalhighdoseratebrachytherapy AT powersallison dosimetriccomparisonoftwoafterloadersandtreatmentplanningsystemsinvaginalhighdoseratebrachytherapy AT sheurendih dosimetriccomparisonoftwoafterloadersandtreatmentplanningsystemsinvaginalhighdoseratebrachytherapy AT loyehchi dosimetriccomparisonoftwoafterloadersandtreatmentplanningsystemsinvaginalhighdoseratebrachytherapy AT dumanevishruta dosimetriccomparisonoftwoafterloadersandtreatmentplanningsystemsinvaginalhighdoseratebrachytherapy |