Different glucose analyzers report different glucose concentration values in term newborns

BACKGROUND: The American Academy of Pediatrics and Pediatric Endocrine Society neonatal hypoglycemia guidelines based their glucose concentration treatment thresholds on studies that predominantly used Beckman and Yellow Springs Glucose Oxidase Analyzers. Currently, a majority (76%) of U.S. hospital...

Descripción completa

Detalles Bibliográficos
Autores principales: Duke, Rina P., Bai, Shasha, Bornhorst, Joshua A., ElHassan, Nahed O., Kaiser, Jeffrey R.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Frontiers Media S.A. 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9532546/
https://www.ncbi.nlm.nih.gov/pubmed/36210937
http://dx.doi.org/10.3389/fped.2022.933508
_version_ 1784802135596072960
author Duke, Rina P.
Bai, Shasha
Bornhorst, Joshua A.
ElHassan, Nahed O.
Kaiser, Jeffrey R.
author_facet Duke, Rina P.
Bai, Shasha
Bornhorst, Joshua A.
ElHassan, Nahed O.
Kaiser, Jeffrey R.
author_sort Duke, Rina P.
collection PubMed
description BACKGROUND: The American Academy of Pediatrics and Pediatric Endocrine Society neonatal hypoglycemia guidelines based their glucose concentration treatment thresholds on studies that predominantly used Beckman and Yellow Springs Glucose Oxidase Analyzers. Currently, a majority (76%) of U.S. hospital laboratories utilizing glucose oxidase methodology use Vitros(®) Glucose Analyzers. However, a bias of ~+5% between glucose concentrations from Beckman vs. Vitros Glucose Analyzers has been reported; this could have a clinically significant effect when using published guideline treatment thresholds. METHODS: To determine if there is similar instrument bias between Beckman and Vitros Analyzers in reported glucose concentrations from term newborns, we compared plasma glucose concentrations measured within the first 3 h after birth by Beckman vs. Vitros Analyzers in a total of 1,987 newborns (Beckman n = 904, Vitros n = 1,083). Data were fit using nonlinear cubic spline models between collection time and glucose concentration. RESULTS: The non-linear patterns of initial glucose concentrations (during the first 3 h after birth) as measured by Beckman and Vitros Analyzers paralleled each other with no overlap of the fit spline curve 95% confidence intervals, with an approximate +5 mg/dL constant bias. Additionally, in method comparison studies performed in the Chemistry Laboratory on adult samples, there was a +4.2-7.4 mg/dL measured glucose bias for the Beckman vs. Vitros Analyzer. CONCLUSION: Glucose concentrations from term, appropriate size for gestational age newborns were about 5 mg/dL higher when measured by Beckman vs. Vitros Analyzers. Perhaps, concentrations of 45 mg/dL reported from Beckman Analyzers may be equivalent to 40 mg/dL from Vitros Analyzers. When managing neonatal hypoglycemia, it is important to know which analyzer was used and whether adjusting for potential instrument bias is necessary when following published guidelines.
format Online
Article
Text
id pubmed-9532546
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Frontiers Media S.A.
record_format MEDLINE/PubMed
spelling pubmed-95325462022-10-06 Different glucose analyzers report different glucose concentration values in term newborns Duke, Rina P. Bai, Shasha Bornhorst, Joshua A. ElHassan, Nahed O. Kaiser, Jeffrey R. Front Pediatr Pediatrics BACKGROUND: The American Academy of Pediatrics and Pediatric Endocrine Society neonatal hypoglycemia guidelines based their glucose concentration treatment thresholds on studies that predominantly used Beckman and Yellow Springs Glucose Oxidase Analyzers. Currently, a majority (76%) of U.S. hospital laboratories utilizing glucose oxidase methodology use Vitros(®) Glucose Analyzers. However, a bias of ~+5% between glucose concentrations from Beckman vs. Vitros Glucose Analyzers has been reported; this could have a clinically significant effect when using published guideline treatment thresholds. METHODS: To determine if there is similar instrument bias between Beckman and Vitros Analyzers in reported glucose concentrations from term newborns, we compared plasma glucose concentrations measured within the first 3 h after birth by Beckman vs. Vitros Analyzers in a total of 1,987 newborns (Beckman n = 904, Vitros n = 1,083). Data were fit using nonlinear cubic spline models between collection time and glucose concentration. RESULTS: The non-linear patterns of initial glucose concentrations (during the first 3 h after birth) as measured by Beckman and Vitros Analyzers paralleled each other with no overlap of the fit spline curve 95% confidence intervals, with an approximate +5 mg/dL constant bias. Additionally, in method comparison studies performed in the Chemistry Laboratory on adult samples, there was a +4.2-7.4 mg/dL measured glucose bias for the Beckman vs. Vitros Analyzer. CONCLUSION: Glucose concentrations from term, appropriate size for gestational age newborns were about 5 mg/dL higher when measured by Beckman vs. Vitros Analyzers. Perhaps, concentrations of 45 mg/dL reported from Beckman Analyzers may be equivalent to 40 mg/dL from Vitros Analyzers. When managing neonatal hypoglycemia, it is important to know which analyzer was used and whether adjusting for potential instrument bias is necessary when following published guidelines. Frontiers Media S.A. 2022-09-21 /pmc/articles/PMC9532546/ /pubmed/36210937 http://dx.doi.org/10.3389/fped.2022.933508 Text en Copyright © 2022 Duke, Bai, Bornhorst, ElHassan and Kaiser. https://creativecommons.org/licenses/by/4.0/This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
spellingShingle Pediatrics
Duke, Rina P.
Bai, Shasha
Bornhorst, Joshua A.
ElHassan, Nahed O.
Kaiser, Jeffrey R.
Different glucose analyzers report different glucose concentration values in term newborns
title Different glucose analyzers report different glucose concentration values in term newborns
title_full Different glucose analyzers report different glucose concentration values in term newborns
title_fullStr Different glucose analyzers report different glucose concentration values in term newborns
title_full_unstemmed Different glucose analyzers report different glucose concentration values in term newborns
title_short Different glucose analyzers report different glucose concentration values in term newborns
title_sort different glucose analyzers report different glucose concentration values in term newborns
topic Pediatrics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9532546/
https://www.ncbi.nlm.nih.gov/pubmed/36210937
http://dx.doi.org/10.3389/fped.2022.933508
work_keys_str_mv AT dukerinap differentglucoseanalyzersreportdifferentglucoseconcentrationvaluesintermnewborns
AT baishasha differentglucoseanalyzersreportdifferentglucoseconcentrationvaluesintermnewborns
AT bornhorstjoshuaa differentglucoseanalyzersreportdifferentglucoseconcentrationvaluesintermnewborns
AT elhassannahedo differentglucoseanalyzersreportdifferentglucoseconcentrationvaluesintermnewborns
AT kaiserjeffreyr differentglucoseanalyzersreportdifferentglucoseconcentrationvaluesintermnewborns